K083081 · Interrad Medical, Inc. · OKC · Mar 4, 2009 · General Hospital
Device Facts
Record ID
K083081
Device Name
MODIFICATION TO SECURACATH, MODEL: SPK01
Applicant
Interrad Medical, Inc.
Product Code
OKC · General Hospital
Decision Date
Mar 4, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SecurAcath PICC is indicated for short or long term peripheral access to the central venous system. The catheter may be used for intravenous therapy, power injection of contrast media, and blood sampling and/or infusion therapy.. The maximum recommended infusion rate is 5 ml/sec and the maximum pressure of power injectors may not exceed 300 psi. The Secur Acath PICC includes means to secure the catheter via a subcutaneous anchor below the insertion site.
Device Story
SecurAcath is a single-use, sterile, flexible dual-lumen PICC catheter (up to 55 cm length) featuring an integrated subcutaneous securement mechanism. Designed for peripheral access to the central venous system; used for IV therapy, blood sampling, and power injection of contrast media (max 5 ml/sec, 300 psi). Clinicians insert the catheter and deploy the subcutaneous anchor below the insertion site to stabilize the device. Compatible with 0.018" guidewires; MR conditional. Provides securement to prevent catheter migration, potentially reducing complications and improving patient outcomes during long-term venous access.
Clinical Evidence
Bench testing only. Performance data included biocompatibility testing, dimensional verification, securement reliability, catheter tensile strength, and power injection capabilities. Results demonstrate the device meets all product specifications and is safe and effective for its intended use.
Technological Characteristics
Dual-lumen flexible catheter; length up to 55 cm with 1 cm markings; integrated subcutaneous securement mechanism in Y-body; MR conditional; compatible with 0.018" guidewires; sterile, single-use.
Indications for Use
Indicated for patients requiring short or long-term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, blood sampling, or infusion therapy.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
K083081
#### SECTION 3. 510(K) SUMMARY
MAR 4 2009
#### 3.1 ADMINISTRATIVE INFORMATION
# 3.1.1 Name and address
Sponsor: 181 Cheshire Lane, Suite 100 Plymouth, MN 55441 Tel: (763) 225-6699 Fax: (763) 225-6694
Contact Person Sew-Wah Tay, PhD Regulatory Consultant 18555 37th Ave N, Plymouth, MN 55446
Tel: 612-801-6782 Fax: 763-208-4465 Email: swtay@libramed.com
Date Prepared: December 24, 2008
## 3.1.2 Device Name
Trade Name Common Name Classification Name Classification SecurAcath Catheter Periphery Inserted Central Catheter (PICC) OKC - implanted subcutaneous securement catheter 21 CFR 880.5970 Class II SPK01
Model
## 3.1.3 Applicant
Applicant's Name: 181 Cheshire Lane, Suite 100 Plymouth, MN 55441 Tel: (763) 225-6699 Fax: (763) 225-6694
### 3.2 INDICATION FOR USE
The SecurAcath PICC is indicated for short or long term peripheral access to the central venous system. The catheter may be used for intravenous therapy, power injection of contrast media, blood sampling and/or infusion therapy. The maximum recommended infusion rate is 5 m/sec and the maximum pressure of power injectors may not exceed 300 psi. The SecurAcath PICC includes means to secure the catheter via a subcutaneous anchor below the insertion site.
{1}------------------------------------------------
### 3.3 DEVICE DESCRIPTION
The SecurAcath is a single use, sterile, flexible PICC catheter with a useful lengths of up to 55 cm. The length of the catheter has markings at 1 cm interval to allow the user to trim the catheter to the required length. The catheter has a built in subcutaneous securement mechanism to stabilize and keep the catheter in place. The catheter has two lumens which are identical in properties and characteristics.
The device is compatible with all 0.018" and smaller guidewires. The proximal end of the catheter has the standard radiology PICC configuration plus the activation mechanism of the securement system built into the Y-body. The catheter has been tested to be compatible with MRI (MR conditional) and use with power contrast injectors.
#### 3.4 SUBSTANTIAL EQUIVALENCE
The SecurAcath Catheter device covered by this submission is substantially equivalent to the Interrad Medical SecurAcath K082047 and Bard PowerPICC K051672
The SecurAcath™ has the same indication for use, same principles of operation, and similar technological characteristics as the previously cleared predicate devices. The differences between this device and its predicate devices do not raise new questions of safety or efficacy.
## 3.5 PERFORMANCE DATA
The performance test data is provided in the 510(k) submission. The performance data demonstrates that the device meets all the product specifications. Performance testing included, biocompatibility testing, dimensional verification; securement reliability, catheter tensile strength and power injection capabilities. Test results demonstrate that the device is safe and effective for its intended use.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird.
#### Public Health Service
MAR 4 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Interrad Medical, Incorporated C/o Sew-Wah Tay, Ph.D Regulatory Consultant Libra Medical LLC 18555 37th Avenue North Plymouth, Minnesota 55446
Re: K083081
Trade/Device Name: SecurAcath 5F Dual Lumen PICC 65cm with Subcutaneous Securement System Regulation Number: 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: OKC Dated: February 5, 2009 Received: February 6, 2009
Dear Dr. Tay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony D. anston for.
Ginette V. Michaud, M.D.
Ginette Y. Michaud. M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATION FOR USE STATEMENT SECTION 2.
510(k) Number (if known): K083081
Device Name: SecurAcath PICC
Indications for Use:
The SecurAcath PICC is indicated for short or long term peripheral access to the central venous system. The catheter may be used for intravenous therapy, power injection of contrast media, and blood sampling and/or infusion therapy.. The maximum recommended infusion rate is 5 ml/sec and the maximum pressure of power injectors may not exceed 300 psi. The Secur Acath PICC includes means to secure the catheter via a subcutaneous anchor below the insertion site.
Prescription Use X Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K083081
· 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.