POWDER-FREE VINYL PATIENT EXAMINATION GLOVES

K083047 · Zibo Wanci Plastic & Rubber Co., Ltd. · LYZ · Feb 12, 2009 · General Hospital

Device Facts

Record IDK083047
Device NamePOWDER-FREE VINYL PATIENT EXAMINATION GLOVES
ApplicantZibo Wanci Plastic & Rubber Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateFeb 12, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Story

Disposable synthetic vinyl patient examination gloves; non-sterile; powder-free. Worn by healthcare personnel during patient examinations to provide a protective barrier against fluids, waste, and environmental contaminants. Device functions as a physical barrier; no electronic or mechanical components. Used in clinical or medical settings. Benefits include reduction of cross-contamination risk between patient and examiner.

Clinical Evidence

No clinical data. Bench testing only: physical and dimensions testing per ASTM D5250-00, FDA 1000 ml water fill test for pinholes, and biocompatibility testing (primary skin irritation and sensitization).

Technological Characteristics

Synthetic vinyl material; powder-free (<= 2 mg powder/glove); non-sterile. Conforms to ASTM D5250-00. Physical barrier sensing/actuation. No software, electronics, or energy source.

Indications for Use

Indicated for use by healthcare and similar personnel as a disposable barrier on hands or fingers to prevent cross-contamination between the examiner and the patient.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KCS 3047 FEB 1 2 2009 ## 510 (K) SUMMARY This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92. #### 1. Submitter's Identification: Geng, Xiao Cong Zibo Wanci Plastic & Rubber Co., Ltd. Hi-tech Industry Area, Cross of Jiging and Zhangtian freeway Zibo, Shandong, China Date summary prepared: October 9, 2008 #### 2. Name of the Device: Zibo Wanci Plastic & Rubber Co., Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free #### 3. Predicate Device Informaton: Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821) #### 4. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-00. #### ਨ Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. #### Comparison to Predicate Devices: 6. Zibo Wanci Plastic & Rubber Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. and Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Powderfree. {1}------------------------------------------------ ### .7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows: The standards used for Zibo Wanci Plastic & Rubber Co., Ltd. glove production are based on ASTM-D-5250-00. All testing meets requirements for Physical and Dimensions Testing conducted on gloves. Inspection Level S-2. AOL 2.5. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove). #### Discussion of Clinical Tests Performed: 8. Not Applicable -- There is no hypoallergenic claim. #### 9. Conclusions: Zibo Wanci Plastic & Rubber Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free conform fully to ASTM-D-5250-00 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services. The seal is circular with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" around the top half of the circle. The bottom half of the circle contains a stylized image of a human figure. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Zibo Wanci Plastic & Rubber Company, Limited C/o Mr. John Zhao Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710 : FEB 1 2 2009 K083047 Re: > Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile Regulation Number: 21 CFR 880.6250 | | | | | Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 3, 2009 Received: February 4, 2009 Dear Mr. Zhao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2 - Mr. Zhao Enclosure Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Anthony D. Nastair for Civil Judicial MD Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Zibo Wanci Plastic & Rubber Co., Ltd. Hi-tech Industry Area, Cross of Jiqing and Zhangtian freeway, Zibo, Shandong, China # INDICATIONS FOR USE Zibo Wanci Plastic & Rubber Co., Ltd. Applicant: 510(k) Number: Patient Vinyl Examination Gloves, Powderfree, Non-Sterile Device Name: Indications of Use: A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250) Prescription Use Over the Counter Use Factory Initials GCJ (fa 48m) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: KO83047
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...