← Product Code [DCN](/productcode/DCN) · K083040

# S-TEST C-REACTIVE PROTEIN (CRP) (K083040)

_Alfa Wassermann Diagnostic Technologies, Inc. · DCN · Jan 30, 2009 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K083040

## Device Facts

- **Applicant:** Alfa Wassermann Diagnostic Technologies, Inc.
- **Product Code:** [DCN](/productcode/DCN.md)
- **Decision Date:** Jan 30, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5270
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The S-Test C-Reactive Protein Reagent is intended for the quantitative determination of C-reactive protein concentration in serum or heparin plasma using the S40 Clinical Analyzer. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only

## Device Story

S-Test CRP reagent cartridge is an in vitro diagnostic tool used with the S40 Clinical Analyzer. It measures CRP concentration in serum or heparin plasma samples via a photometric test. The principle of operation involves an antigen-antibody reaction between CRP in the patient sample and anti-CRP mouse monoclonal antibody-coated latex particles, resulting in agglutination. The analyzer measures the resulting turbidity photometrically. Used in clinical laboratories or physician office laboratories by trained personnel. The quantitative output assists healthcare providers in evaluating the extent of tissue injury. The device provides a standardized, automated method for CRP assessment, facilitating clinical decision-making regarding inflammatory or injury-related conditions.

## Clinical Evidence

Bench testing only. Precision studies (n=24 days) showed within-run CV 1.1-3.0% and total CV 2.3-4.3%. POL site precision showed within-run CV 1.0-7.5% and total CV 1.5-7.5%. Accuracy correlation study (n=95 samples, 5-91 mg/L) yielded r=0.995. Matrix comparison (serum vs. heparin plasma) using paired t-test showed no statistically significant difference (t-statistic 0.13 vs. t-critical 2.08). Detection limit is 2 mg/L.

## Technological Characteristics

Single-use plastic cartridge with two reagent wells. Reagents: 2-amino-2-hydroxymethyl-1,3-propanediol buffer and anti-human CRP mouse monoclonal antibody-coated latex particles. Sensing: latex immunoturbidimetry; bi-chromatic absorbance measurement (570 nm/800 nm). Connectivity: 2-D barcode for lot-specific calibration and parameters. Instrument: S40 Clinical Analyzer. Traceable to IRMM CRM470.

## Regulatory Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

## Predicate Devices

- Olympus AU640 Clinical Chemistry Analyzer, CRP Latex Test (k961274)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k083040

B. Purpose for Submission:
New device

C. Measurand:
C Reactive Protein (CRP)

D. Type of Test:
Quantitative – latex immunoturbimetric

E. Applicant:
Alfa Wassermann Diagnostic Technology, Inc.

F. Proprietary and Established Names:
S-Test C Reactive Protein (CRP) Reagent cartridge

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DCN – C Reactive Protein | Class II | 21 CFR§ 866.5270 | 75 Chemistry  |

H. Intended Use:

1. Intended use(s):
See indications for use below

2. Indication(s) for use:
The S-Test C-Reactive Protein Reagent is intended for the quantitative determination of C-reactive protein concentration in serum or heparin plasma using the S40 Clinical Analyzer. Measurement of C-reactive protein aids in

{1}

evaluation of the amount of injury to body tissues. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only

3. Special conditions for use statement(s):

For prescription use only

This test is not intended for high sensitivity CRP measurements.

4. Special instrument requirements:

S40 Clinical Analyzer

I. Device Description:

The S-Test C-Reactive Protein (CRP) device is a single use plastic cartridge composed of two individual reagents in separate wells, a reaction well with a photometric cuvette, and a bar code label which provides all the chemistry parameters, calibration factors, and product information for the instrument. The two reagent wells contain the following materials: (1) 2 amino-2 hydroxymethyl-1,3 propanediol in buffer and (2) anti-human c-reactive protein mouse monoclonal antibody coated latex particles.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Olympus AU640 Clinical Chemistry Analyzer, CRP Latex Test

2. Predicate 510(k) number(s):

k961274

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | S-Test CRP Reagent | CRP Latex Test  |
|  Sample type | Serum or lithium heparin plasma | Serum or lithium heparin plasma  |
|  Principle | Latex immunoturbidimetry | Latex immunoturbidimetry  |
|  Reaction type | Endpoint | Endpoint  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | S-Test CRP Reagent | CRP Latex Test  |
|  Indications for Use | Intended for the quantitative determination of C-reactive protein concentration in serum or heparin plasma using the S40 Clinical Analyzer. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues. This test is not intended for high sensitivity CRP measurement | Intended for the quantitative determination of C-reactive protein in serum or plasma on Olympus analyzers. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. High sensitivity CRP measurements may be useful as an independent marker for recurrent events in patients with stable coronary events or acute coronary syndromes  |
|  Instrument | S40 Clinical Analyzer | Olympus analyzers  |
|  Measuring Range | 5-98 mg/L | 1 to 480 mg/L  |
|  Detection Limit | 2 mg/L | ≤0.15 mg/L  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (2004)

CLSI EP10-A: Preliminary Evaluation of Quantitative Clinical Laboratory Methods; Approved Guideline -Second Edition (2002)

CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach: Approved Guideline (2003)

CLSI EP7-A: Interference Testing in Clinical Chemistry; Approved Guideline (2002)

CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline (2004)

CLSI EP9-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (2002)

{3}

L. Test Principle:

S-Test CRP – CRP in the sample causes an antigen-antibody reaction with antihuman C-reactive protein mouse monoclonal antibody-coated latex particles, resulting in agglutination. The C-reactive protein concentration is determined by measuring this agglutination by measuring the absorbance change. The rate of increase in absorbance, monitored bi-chromatically at 570 nm/800 nm, is directly proportional to the C-Reactive Protein concentration in the sample.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Three serum samples (with normal, intermediate and elevated levels) were tested for CRP on two S40 Clinical Analyzers two times per run, two runs per day, for a total of 24 days. The mean, standard deviations, and % coefficients of variation (CV) were calculated for each sample.

|  Sample 1
Mean = 13 mg/L CRP | Within
Run | Between
Run | Between
Day | Total  |
| --- | --- | --- | --- | --- |
|  Coefficient of
Variation | 3.0% | 2.5% | 1.7% | 4.3%  |
|  Sample 2
Mean = 29 mg/L CRP | Within
Run | Between
Run | Between
Day | Total  |
| --- | --- | --- | --- | --- |
|  Coefficient of
Variation | 1.8% | 1.8% | 1.6% | 3.0%  |
|  Sample 3
Mean = 54 mg/L CRP | Within
Run | Between
Run | Between
Day | Total  |
| --- | --- | --- | --- | --- |
|  Coefficient of
Variation | 1.1% | 0.9% | 1.8% | 2.3%  |

In-house precision studies were conducted by testing human serum pools at three levels (normal, intermediate and elevated). The samples were assayed three times a day for five days using one instrument. Precision studies were also conducted at three Physician Office Laboratories (POL) with four trained

{4}

operators typically found in these settings. Human serum pools at three concentrations were tested three times a day for five days on four instruments (one at each lab). The results are presented in the table below:

|   |   |   | %CV or SD (unit)  |   |
| --- | --- | --- | --- | --- |
|  Lab | Sample | Mean | Within-Run | Total  |
|  In-house | 1 | 9 | SD 0.4 | SD 0.4  |
|   |   |   |  4.8% | 4.8%  |
|  POL 1 | 1 | 9 | SD 0.5 | SD 0.5  |
|   |   |   |  5.8% | 5.8%  |
|  POL 2 | 1 | 8 | SD 0.5 | SD 0.5  |
|   |   |   |  6.2% | 6.2%  |
|  POL 3 | 1 | 9 | SD 0.6 | SD 0.6  |
|   |   |   |  7.5% | 7.5%  |
|  |   |   |   |   |
|  In-house | 2 | 28 | SD 0.8 | SD 0.8  |
|   |   |   |  3.0% | 3.0%  |
|  POL 1 | 2 | 29 | SD 0.4 | SD 0.5  |
|   |   |   |  1.4% | 1.8%  |
|  POL 2 | 2 | 28 | SD 0.3 | SD 0.5  |
|   |   |   |  1.1% | 1.7%  |
|  POL 3 | 2 | 28 | SD 1.4 | SD 1.7  |
|   |   |   |  5.1% | 6.0%  |
|  |   |   |   |   |
|  In-house | 3 | 71 | SD 1.4 | SD 1.4  |
|   |   |   |  2.0% | 2.0%  |
|  POL 1 | 3 | 75 | SD 1.3 | SD 1.3  |
|   |   |   |  1.7% | 1.7%  |
|  POL 2 | 3 | 71 | SD 0.7 | SD 1.1  |
|   |   |   |  1.0% | 1.5%  |
|  POL 3 | 3 | 73 | SD 2.2 | SD 3.3  |
|   |   |   |  3.0% | 4.4%  |

# b. Linearity/assay reportable range:

The reportable range is 5 to  $98\mathrm{mg / L}$ . The assay is linear from 5 to  $98\mathrm{mg / L}$  on the S40 Clinical Analyzer. If a sample result is less than  $5\mathrm{mg / L}$ , the result is flagged by the analyzer as  $&lt; 5\mathrm{mg / L}$ . If a sample result exceeds  $98\mathrm{mg / L}$ , the result is flagged by the analyzer as  $&gt;98\mathrm{mg / L}$ . This range is supported by the limit of detection study (section M.1.d below), the method comparison (section M.2.a below), and the linearity study shown below.

{5}

Linearity across the assay range was confirmed by testing commercial linearity standards at 12 levels ranging from 4.9 mg/L to 98.5 mg/L. Each level was tested in replicates of four. The percent recoveries ranged from 95.7 to 106.1%. The linear regression equation obtained for the study was y = 1.035x - 0.160, r² = 0.9982

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The S Test-CRP cartridges are factory calibrated and traceable to the Institute for Reference Material and Measurement (IRMM) Standard Reference Material CRM470. The 2-D barcode printed on each cartridge provides the analyzer with lot-specific calibration data.

d. Detection limit:

The Limit of Blank and Limit of Detection were determined by assaying low samples (5 serum samples were used) and blank samples (bovine serum albumin in saline). Both the low samples (60 reps, 20 reps per day) and the blanks were assayed each day (5 different samples, 4 reps on 3 days, total n = 60 low samples and 60 blank samples). The testing was split between two instruments. The limit of detection was determined to be 2 mg/L. The assay reportable range is from 5 to 98 mg/L on the S40 Clinical Analyzer. If a sample result is less than 5 mg/L, the result is flagged by the analyzer as &lt;5 mg/L. If a sample result exceeds 98 mg/L, the result is flagged by the analyzer as &gt;98 mg/L.

e. Analytical specificity:

Interference studies to determine the effects of Unconjugated Bilirubin, Hemolysis and Lipemia were performed. Serum pools containing 8 mg/L CRP and 21 mg/L CRP were spiked with various concentrations of unconjugated bilirubin (1.6 - 50 mg/dL), hemoglobin (31-1000 mg/dL) and Intralipid (63 - 2000 mg/dL). The sponsor states that interference is considered to be significant if the analyte recovery changes by more than ±10%. There was no significant interference observed from bilirubin, lipemia, or hemolysis at the tested concentrations.

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

{6}

A series of 95 serum specimens with CRP values ranging from 5 to 91 mg/L were assayed in singlicate on the S40 Clinical Analyzer using S-Test CRP Reagent and Olympus AU640 Analyzer as the reference method. Least-squares regression analysis (Deming) yielded the following results:

|  Regression Equation | y = 0.982x + 1.0  |
| --- | --- |
|  Correlation Coefficient | 0.9945  |
|  Std. Error Est. | 2.1  |
|  Confidence Interval Slope | 0.961 to 1.012  |
|  Confidence Interval Intercept | 0.4 to 1.6  |

## Point of Care in Physician's Office Laboratories

Performance for the S-Test CRP was evaluated at three Physician Office Laboratories and with a total of three operators. Operators assayed unaltered clinical serum samples with values ranging from 5 - 88 mg/L. The correlation study between the device and the predicate for serum yielded the following results.

|  Lab | 1 | 2 | 3  |
| --- | --- | --- | --- |
|  N | 43 | 42 | 43  |
|  Range (mg/L) | 6 -87 | 7 - 88 | 5 - 84  |
|  Regression equation | y = 0.959x +0.7 | y = 0.966x + 1.4 | Y = 0.958x + 0.6  |
|  Correlation coefficient | 0.9979 | 0.9978 | 0.9971  |
|  Standard error | 1.5 | 1.5 | 1.6  |
|  Confidence interval slope | 0.939 to 0.979 | 0.945 to 0.986 | 0.934 to 0.981  |
|  Confidence interval intercept | 0.0 to 1.3 | 0.7 to 2.1 | -0.1 to 1.3  |

## b. Matrix comparison:

A study was performed on the S40 by running CRP determinations on 22 paired samples drawn from the same patients in serum and heparin plasma tubes at four POL labs and at a clinical laboratory. The serum results ranged from 7 to 96 mg/L. Six samples were spiked. Least-squares regression analysis (Deming) yielded the following results (serum - x, plasma - y):

{7}

|  Regression Equation | y = 1.019x - 0.5  |
| --- | --- |
|  Correlation Coefficient | 0.9985  |
|  Std. Error Est. | 1.6  |
|  Confidence Interval Slope | 0.994 to 1.045  |
|  Confidence Interval Intercept | -1.6 to 0.5  |

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

CRP:  $&lt; 10 \mathrm{mg} / \mathrm{L}$

Referenced from: Tietz, N.W. (Ed.), Clinical Guide to Laboratory Tests, 3rd Edition, W.B. Saunders Co., Philadelphia, PA (1995).

The S-Test CRP assay is not intended for high sensitivity CRP (hsCRP) measurements.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K083040](https://fda-staging.innolitics.com/device/K083040)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
