The Breathe Technologies™ Ventilator, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician.
Device Story
Breathe Technologies Ventilator (BT-V) is a battery-powered, wearable volume assist ventilator; augments spontaneous breathing in tracheostomy patients. Device inputs: patient's spontaneous inspiratory effort. Device outputs: physician-prescribed volume delivered via Transtracheal Tube (BT-TT). Used in institutional settings; operated by trained personnel under physician direction. Small, lightweight (3.1 lb) form factor allows patient mobility, ambulation, and participation in physical/occupational therapy. Benefits include increased patient movement and rehabilitation potential compared to bulkier, full-featured ventilators. Healthcare providers use device to supplement inspiratory efforts, facilitating patient activity while maintaining respiratory support.
Clinical Evidence
Bench testing only. Biocompatibility testing performed in compliance with ISO 10993. Bench testing demonstrated compliance with medical community expectations, product labeling, and relevant guidance and standards. No clinical data provided.
Technological Characteristics
Battery-powered, wearable volume assist ventilator. Dimensions/weight: 3.1 lbs. Connectivity: None stated. Sterilization: Not specified. Software: Embedded firmware for volume assist. Biocompatibility: ISO 10993 compliant. Sensing/Actuation: Volume-based ventilation triggered by patient spontaneous breathing.
Indications for Use
Indicated for adult patients with respiratory insufficiency and a tracheostomy capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. Intended for continuous use in institutional settings to support ambulation and rehabilitation.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K0829982
# JUL 2009
| Category | Comments |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Sponsor: | Breathe Technologies<br>3089 Skyway Court<br>Fremont, CA 94539-5909<br>Tel: 510.360.9966<br>Fax: 510.360.9967 |
| Correspondent Contact<br>Information: | Suzon Lommel<br>Breathe Technologies<br>4000 Executive Parkway Suite 190<br>San Ramon, CA 94583 |
| Device Common Name: | Continuous Ventilator, Facility Use |
| Device Classification & Code: | Class II, CBK |
| Device Classification Name: | 21 CFR 868.5895 |
| Device Proprietary Name: | Ventilator |
## Section 5: 510(k) Summary
#### Predicate Device Information:
| Predicate Device: | LTV 1000 & 800 Models |
|-----------------------------------------|-------------------------------------|
| Predicate Device Manufacturer: | Pulmonetic Systems |
| Predicate Device Common Name: | Continuous Ventilator, Facility Use |
| Predicate Device Classification: | 21 CFR 868.5895 |
| Predicate Device Classification & Code: | Class II, CBK |
| Premarket Notification Number: | K981371 |
| Predicate Device: | Legendair XL2 |
|-----------------------------------------|-------------------------------------|
| Predicate Device Manufacturer: | Puritan Bennett |
| Predicate Device Common Name: | Continuous Ventilator, Facility Use |
| Predicate Device Classification: | 21 CFR 868.5895 |
| Predicate Device Classification & Code: | Class II, CBK |
| Premarket Notification Number: | K070899 |
## b. Date Summary Prepared
10 June 2009
#### c. Description of Device
## c.1. Intended Patient Populations and Medical Conditions
Breathe Technologies has identified a group of ventilator dependent, or highly oxygen dependent patients, in institutional settings that are being poorly served by current ventilators. This group needs ventilation support but only at low levels. They need mobility within the facility. "Mobility" includes modest movement such as free motion in
{1}------------------------------------------------
bed, standing and sitting to use a commode, and movement to the bathroom. It may also, at the physicians' discretion, allow a level of ambulation to support participation in physical therapy, but the weight and bulk of current ventilators greatly inhibit their potential for physical movement. The ease of carrying the 3.1 lb ventilator, while their inspiratory efforts are properly supplemented, may allow the potential for ambulation for these patients. It is intended only for institutional use, not for home use.
### c.2. General Description of the Breathe Technologies Ventilator, Transtracheal Tube and accessories
The Breathe Technologies™ Ventilator (BT-V) is a battery powered, wearable, volume ventilator that augments the patient's spontaneous breathing.
The BT-V administers this physician-prescribed volume to the patient via the attached Breathe Technologies Transtracheal Tube (BT-TT). The end of the BT-TT is inserted into the patient's tracheostomy tube.
The ventilator is small and light enough to be worn on a patient's belt, or slung over their shoulder, because it has only a subset of the features of a full-featured ventilator.
#### d. Intended Use
The Breathe Technologies™ Ventilator (BT-V), with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The BT-V is designed for continuous applications in an institutional environment. The BT-V is a restricted medical device intended for operation by trained personnel under the direction of a physician.
#### e. Comparison to Predicate Device
The Breathe Technologies Ventilation System (Ventilator and Accessories) is substantially equivalent to the predicate Pulmonetic Systems' LTV 1000 & 800 Models (K981371) and the Puritan Bennett Legendair XL2 (K070899) in intended use, classification code, technology design and physician's use in that it conforms to a subset of the intended use, technology, physician use, features, performance and alarms of the predicate devices.
By conforming to a subset of design and use of the predicate devices, the Breathe Technologies Ventilation System can be used by some spontaneously breathing, ventilator dependent patients, to ambulate in an institutional setting.
Breath Technologies concludes that the predicate devices and the Breathe Technologies Ventilator System (Ventilator and all accessories) are substantially equivalent.
{2}------------------------------------------------
## f. Summary of Supporting Data
Biocompatibility data demonstrates that the device is in compliance with ISO 10993.
Bench testing has demonstrated that the device is in compliance with the medical community's expectations, the product labeling and the pertinent sections of the guidance's and standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines that curve at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus symbol. The text is in all capital letters and is evenly spaced around the circle.
#### SEP 9 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Suzon Lommel Vice President, Regulatory and Quality Affairs Breathe Technologies, Incorporated 4000 Executive Parkway, Suite 190 San Ramon, California 94583
Re: K082982
Trade/Device Name: Breathe Technologies Ventilator & Accessories Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: ONZ Dated: June 25, 2009 Received: June 30, 2009
Dear Ms. Lommel:
This letter corrects our substantially equivalent letter of July 20, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Lommel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Runser
Susan Runner, DDS, MA Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Breathe Technologies. The word "Breathe" is in large, bold letters on the top line. Below that, in smaller letters, is the word "TECHNOLOGIES". Underneath the words is a curved line.
# Section 4: Indications for Use Statement
510(k) Number K082982:
Breathe Technologies Ventilator & Accessories Device Name:
## Indications For Use:
The Breathe Technologies™ Ventilator, with accessories, is a volume assist ventilator intended to aid adult patients with respiratory insufficiency. It is designed for patients with a tracheostomy that are capable of spontaneously breathing a minimum tidal volume of 3.5cc/kg of predicted body weight. The device is designed for continuous applications such as patient ambulation, physical therapy, occupational therapy, respiratory therapy, and other rehabilitation efforts in an institutional environment. The device is intended for operation by trained personnel under the direction of a physician.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) ห Schuts
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
: 510(k) Number: K082982
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.