ApneiCare Connection Center/Internet Analysis is a software device, intended to be used as an aid in the diagnosis of sleep and respiratory related sleep disorders. ApneiCare Connection Center/Internet Analysis is used for analysis (semi-automatic and manual rescoring), display, redisplay (retrieve), edit, summarize, to generate user defined reports, networking and managing of data received from devices that are typically used to evaluate sleep- and sleep related respiratory disorders. ApneiCare Connection Center/Internet Analysis software also provides scoring of desaturation events as they related to sleep breathing disorders. ApneiCare Connection Center/Internet Analysis is intended to be used as an information and decision management tool to import/record and transmit/transfer data typically captured to evaluate sleep and respiratory related sleep disorders. ApneiCare Connection Center/Internet Analysis itself is not a diagnostic tool. It is an information and decision management tool which allows medical personnel to upload, view, and score data related to a sleep study and provide output reports to a physician. The use of this software is to be under the supervision of a physician, sleep technologies, or clinician.
Device Story
Client-server software system for sleep study data management; Connection Center (PC-based) binds patient info to study data from external sleep devices; transmits data via encrypted HTTPS to Internet Analysis server. Server performs mathematical data processing; identifies waveform patterns (SpO2, pulse); scores desaturation and respiratory events using DPAR objects; provides manual rescoring and visualization tools. Used in home care, clinics, and hospitals by clinicians/technicians. Output reports aid physicians in diagnosing sleep disorders; facilitates data management, viewing, and reporting.
Clinical Evidence
Bench testing only. Comparative testing demonstrated equivalence to predicate devices.
Technological Characteristics
Client-server architecture; Connection Center runs on Windows XP/Vista (.NET 2.0); Internet Analysis server runs on Windows Server 2008, SQL Server 2005, and .NET 3.5. Data transmission via 128-bit HTTPS. No patient-contact materials.
Indications for Use
Indicated for all individuals undergoing evaluation for sleep and respiratory-related sleep disorders. Used under supervision of physician, sleep technologist, or clinician. No contraindications.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
K082968
### Non-Confidential Summary of Safety and Effectiveness
Page 1 of 4 30-Dec-2008
# JAN - 2 2009
### Company
ApneiCare, LLC PO Box 418 Worthington, OH 43085
#### Official Contact:
ApneiCare, LLC Craig Pickerill, Vice President craig@apneicare.com P 614-774-6003
# Proprietary or Trade Name:
ApneiCare Connection Center / Internet Analysis
#### Common/Usual Name:
Polysomnography Scoring Software
# Classification Information:
Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: Class II Product Code: MNR
Device: ApneiCare Connection Center / Internet Analysis
Predicate Devices: Vitascore BV - Vitascore - 510(k) K072014 Respironics - ProFox Software - 510(k) K001708
# Device Description:
ApneiCare Connection Center / Internet Analysis is composed of two components: The Connection Center and the Internet Analysis Application. The ApneiCare Connection Center and the ApneiCare Internet Analysis Application function in a client server relationship.
The purpose of ApneiCare Connection Center is to download stored data from supported devices that are typically used to evaluate sleep and sleep related respiratory disorders. It allows users to
{1}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness Page 2 of 4 30-Dec-2008
export sleep data, and interact with the ApneiCare Internet Analysis to run analysis reports on sleep studies.
ApneiCare Connection Center will operate under Windows XP and Windows Vista; but more importantly is built on Microsoft .NET Framework. The ApneiCare applications have been designed on Microsoft .NET Framework to take advantage of the latest laver of abstraction to allow for Operating System changes in the future while lowering the probability of the application being negatively impacted by operating system changes. "ApneiCare Connection Center is responsible for binding patient information to the study and downloading the data off of the source device. The ApneiCare Connection Center is a small program on the PC with functionality limited to that of general purpose data entry, storage and communication, it performs no data manipulation functions. The data is then communicated to ApneiCare Internet Analysis via an Internet connection, where the sleep analysis and reporting is performed. The ApneiCare Connection Center has built in features to insure that required patient information is entered and that internet connectivity is available. All Internet activity is encrypted and secure. Each transaction between ApneiCare Connection Center and ApneiCare Internet Analysis is conducted over at least 128-bit HTTPS using private and public encryption.
Saved tests can be converted into ASCII format for importing to spreadsheets or databases. The patient information consists of the patient name and address, an ID number, patient insurance and other relevant information. The ApneiCare Internet Analysis Application provides information pertaining to the waveform patterns, such as those of the SpO2 and pulse recording, to aid physicians with the diagnosis of sleep and breathing disorders. Individual desaturation and respiratory events in series (cycles) are scored by algorithm. These patterns are identified using DPAR objects. In addition, to the Internet Analysis the ApneiCare Internet Analysis DPAR Case Viewer application module may also be used to manually score events and visually compare and analyze across multiple channels of data.
The ApneiCare Internet Analysis Application is primarily a mathematical data processing function that reports cycling respiratory events across multiple channels of data. The functionality of the ApneiCare Internet Analysis Application is not time critical or real time. Therefore delays in internet transfer do not impact functionality. The ApneiCare Internet Analysis Application also performs a data management functions which allows the various authorized users of the system to access, view, report and analyze individual cases. The ApneiCare Connection Center is compiled for .NET 2.0, and dependent upon the .NET Framework as the OTS. .NET 2.0 is provided by Microsoft and contains the CLR (Common Language Runtime) which has abstracted ApneiCare Connection Center away from specific Microsoft Windows Operating System Versions. The ApneiCare Internet Analysis application is a server application, runs on Windows Server 2008 and SQL Server 2005, and .NET 3.5, it is managed in a centralized location by ApneiCare.
{2}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness Page 3 of 4 30-Dec-2008
# Indications for Use:
ApneiCare Connection Center/Internet Analysis is a software device, intended to be used as an aid in the diagnosis of sleep and respiratory related sleep disorders. ApneiCare Connection Center/Internet Analysis is used for analysis (semi-automatic and manual rescoring), display, redisplay (retrieve), edit, summarize, to generate user defined reports, networking and managing of data received from devices that are typically used to evaluate sleep- and sleep related respiratory disorders. ApneiCare Connection Center/Internet Analysis software also provides scoring of desaturation events as they related to sleep breathing disorders.
ApneiCare Connection Center/Internet Analysis is intended to be used as an information and decision management tool to import/record and transmit/transfer data typically captured to evaluate sleep and respiratory related sleep disorders.
ApneiCare Connection Center/Internet Analysis itself is not a diagnostic tool. It is an information and decision management tool which allows medical personnel to upload, view, and score data related to a sleep study and provide output reports to a physician.
The use of this software is to be under the supervision of a physician, sleep technologies, or clinician.
| Attributes | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use<br>General | ApneiCare Connection Center/Internet Analysis is intended to<br>be used as an aid in the diagnosis of sleep and respiratory<br>related sleep disorders. The ApneiCare Connection Center<br>and ApneiCare Internet Analysis Application is intended to be<br>used as an information and decision management tool to<br>import/record and transmit/transfer data typically captured to<br>evaluate sleep and respiratory related sleep disorders. |
| Intended use<br>Specific | ApneiCare Connection Center/Internet Analysis is intended to<br>be used as an aid in the diagnosis of sleep and respiratory<br>related sleep disorders. The ApneiCare Connection Center<br>and ApneiCare Internet Analysis Application is intended to be<br>used as an information and decision management tool to<br>import/record and transmit/transfer data typically captured to<br>evaluate sleep and respiratory related sleep disorders. |
| Environments of use | home care, nursing home, sub-acute institutions, sleep clinics<br>or hospitals |
| Patient Population | All individuals |
| Contraindications | None |
| Prescription | Prescription |
| Materials | None in patient contact |
| Performance | Comparative testing demonstrate equivalence to the predciate |
#### Device Attributes:
{3}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness Page 4 of 4 30-Dec-2008
# Differences Between Other Legally Marketed Predicate Devices
The is viewed as substantially equivalent to the following predicate devices - Vitascore K072014 ProFox Software K001708.
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ApneiCare, LLC C/O Mr. Paul E. Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134-2958
. .
Re: K082968
Trade/Device Name: ApneiCare Connection Center/Internet Analysis Regulation Number: 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: October 2, 2008 Received: October 6, 2008
IAN - 2 2009
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clas
Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{6}------------------------------------------------
Page 1 of 1
# 510(k) Number:
# K082968
Device Name:
ApneiCare Connection Center / Internet Analysis
# Indications for Use:
ApneiCare Connection Center/Internet Analysis is a software device, intended to be used as an aid in the diagnosis of sleep and respiratory related sleep disorders. ApneiCare Connection Center/Internet Analysis is used for analysis (semi-automatic and manual rescoring), display, redisplay (retrieve), edit, summarize, to generate user defined reports, networking and managing of data received from devices that are typically used to evaluate sleep- and sleep related respiratory disorders. ApneiCare Connection Center/Internet Analysis software also provides scoring of desaturation events as they related to sleep breathing disorders.
ApneiCare Connection Center/Internet Analysis is intended to be used as an information and decision management tool to import/record and transmit/transfer data typically captured to evaluate sleep and respiratory related sleep disorders.
ApneiCare Connection Center/Internet Analysis itself is not a diagnostic tool. It is an . information and decision management tool which allows medical personnel to upload, view, and score data related to a sleep study and provide output reports to a physician. The use of this software is to be under the supervision of a physician, sleep technologies, or clinician.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K082968
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.