K082947 · Lensx Lasers, Inc. · OOE · Aug 14, 2009 · Ophthalmic
Device Facts
Record ID
K082947
Device Name
LENSX 550 LASER SYSTEM
Applicant
Lensx Lasers, Inc.
Product Code
OOE · Ophthalmic
Decision Date
Aug 14, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LenSx 550 Laser System is indicated for anterior capsulotomy during cataract surgery.
Device Story
LenSx 550 Laser System is an ophthalmic surgical laser for anterior capsulotomy. It focuses low-energy infrared femtosecond laser pulses into the anterior lens. A computer-controlled scanning system directs the beam in a 3D pattern to create incisions via tissue photodisruption. The system uses an off-the-shelf disposable contact lens to fix the eye relative to the delivery system. Operated by a surgeon in an operating room, the device provides precise, automated capsulotomy creation. The surgeon monitors the procedure via an operating microscope. By replacing manual capsulotomy with a laser-created incision, the device aims to improve the uniformity, accuracy, and predictability of the capsular opening, potentially facilitating safer intraocular lens placement and reducing complications like radial tears.
Clinical Evidence
Prospective, single-center clinical trial evaluating anterior capsulotomy performance during cataract/IOL surgery. Primary endpoints included successful capsulotomy creation, centering, and absence of complications. Results: 100% success rate in capsulotomy completion; no radial tears observed intraoperatively or postoperatively; all IOLs centered; no posterior capsule tears. Bench testing confirmed uniform, accurate, and predictable capsulotomy size/depth in porcine eyes and plastic substrates, with edge smoothness comparable to manual techniques.
Technological Characteristics
Ophthalmic femtosecond laser system; infrared light source; computer-controlled 3D scanning system; disposable contact lens interface. Class II device (21 CFR 886.4390).
Indications for Use
Indicated for anterior capsulotomy during cataract surgery in patients requiring the procedure.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
# 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
# SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON, AND DATE SUMMARY PREPARED
| a. | Applicant: | LenSx Lasers, Inc.<br>33 Journey, Suite 175<br>Aliso Viejo, CA 92656<br>Tel: (949) 360-6010<br>Fax: (949) 612-1843 |
|----|-----------------|--------------------------------------------------------------------------------------------------------------------|
| b. | Contact Person: | Judy Gordon, D.V.M.<br>ClinReg Consulting Services, |
Inc. 733 Bolsana Drive Laguna Beach, CA 92651 judy@clinregconsulting.com Tel: (949) 715-0609 Fax: (949) 715-0610
# NAME OF DEVICE, INCLUDING TRADE NAME AND CLASSIFICATION NAME
| a. | Trade/Proprietary Name: | LenSx 550 Laser System |
|----|-------------------------|------------------------------------------------------------|
| b. | Common/Usual Name: | LenSx 550 Laser System |
| c. | Device Classification: | Ophthalmic Femtosecond Laser, Class II,<br>21 CFR 886.4390 |
| d. | Product Code: | OOE |
ರ. Product Code:
K082947- SUMMARY OF SAFETY AND EFFECTIVENESS
..
{1}------------------------------------------------
#### PREDICATE DEVICES
| 510(k) # | TRADE NAME | MANUFACTURER |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------|
| K001498 | The Fugo Blade | Medisurg Research and<br>Management Corp. |
| K040005 | Profile 3000 laser System and<br>Delivery Devices and<br>Accessories | Sciton, Inc. |
| K052806 | Burane Laser System | WaveLight Laser Technologie,<br>AG |
| K021548 | Fotona Dualis<br>Nd:YAG/Er:YAG Laser System | Fontona D.D. |
| K033354 | FEMTEC Laser Microkeratome | 20/10 Perfect Vision Optische<br>Gerate GMBH |
| K993153, K001211,<br>K002890, K013941,<br>K031960, K041893,<br>K060372, K063682, and<br>K073404 | IntraLase FS Lasers | IntraLase Corp. |
### DEVICE DESCRIPTION
The LenSx 550 Laser is an ophthalmic surgical laser intended for use in patients requiring an anterior capsulotomy. The LenSx 550 Laser focuses a beam of low energy pulses of infrared light into the anterior lens. Consistent with commercially available femtosecond lasers used for lamellar resection of the cornea (GEX, 878.4810), the cutting effect is produced by scanning individual pulses to produce a continuous incision. The location of the tissue photodisruption is controlled by moving the focus of the laser beam to the desired location. A computer-controlled scanning system directs the laser beam throughout a three-dimensional pattern to produce an incision. Finally, the laser pulses are delivered through an off-the-shelf disposable contact lens that contacts the cornea and fixes the eye with respect to the delivery system.
#### STATEMENT OF INTENDED USE
The LenSx 550 Laser System is indicated for anterior capsulotomy during cataract surgery.
{2}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS COMPARISON
The LenSx 550 laser shares a similar intended use, and indication for use as the Fugo Blade by Medisurg Research and Management Corp., the Profile 3000 Laser System (K040005), the Burane Laser System (K052806) and the Fontona Dualis Nd:YAG/Er:YAG Laser System (K021548). The LenSx 550 mode of operation and the technology utilized to create the cutting action are similar to the 20/10 Perfect Vision FemTec Laser Microkeratome, and the IntraLase FS Lasers and is therefore substantially equivalent to these legally marketed predicate devices.
## BRIEF SUMMARY OF NONCLINICAL TESTS AND RESULTS
LenSx has developed the LenSx 550 for use in anterior capsulotomy, employing the precision and control of femtosecond lasers for this procedure. Pre-clinical and clinical testing has included bench testing and a prospective, single-center clinical trial.
#### Summary of Pre-Clinical Testing
Testing and analyses performed included accuracy and reproducibility of capsulotomy size and depth in porcine eyes and plastic substrates, as well as scanning electron microscopy. The data demonstrated the LenSx 550 produces anterior capsulotomies that are uniform, accurate and predictable in size over a wide range of depths. The capsular edge created by the LenSx 550 is also at least as smooth as that created during manual anterior capsulotomy.
#### Summary of Clinical Testing
A clinical trial of the LenSx 550 Laser was conducted to evaluate the performance of this laser system in the creation of anterior capsulotomy during cataract/IOL surgery. Anterior capsulotomy was successfully performed in all eyes using the LenSx 550 Laser. i.e., the capsulotomy was complete in all eyes, with no radial tears observed intraoperatively or postoperatively, and with an intraocular lens placed in the capsular bag. Postoperatively, the course of follow-up in the study population was unremarkable, The intraocular lens was centered in all study cyes, and no posterior capsule tears were observed. All capsulotomies were judged to be well centered by the surgeon using visual inspection in the operating microscope.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
LenSx Lasers, Inc. c/o Judy Gordon, D.V.M. Regulatory Consultant to LenSx Lasers, Inc. ClingReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, CA 92651
AUG 1 4 2009
Re: K082947
Trade/Device Name: LenSx 550 Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: II Product Code: OOE Dated: July 8, 2009 Received: July 9, 2009
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Judy Gordon, D.V.M.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K082947
LenSx 550 Laser System Device Name(s):
Indications for Use:
The LenSx 550 Laser System is indicated for anterior capsulotómy during cataract surgery.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ 1__ of __ 1
Bruce Drum
(Division Sign-Off) Division of Ophthalmie, Neurological and Ear, Nose and Throat Devices
510(k) Number K082947
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.