← Product Code [EHD](/productcode/EHD) · K082875

# PORTABLE X-RAY SYSTEM, MODEL DXR -1 (K082875)

_Exaro Co., Ltd. · EHD · Feb 5, 2009 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082875

## Device Facts

- **Applicant:** Exaro Co., Ltd.
- **Product Code:** [EHD](/productcode/EHD.md)
- **Decision Date:** Feb 5, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.1800
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Pediatric

## Indications for Use

The Portable X-ray System (Model: DXR-1) is intended to be used by trained dentists and dental technicians as extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

## Device Story

The DXR-1 is a portable, hand-held dental X-ray system. It operates on a 25.2V DC rechargeable Li-Polymer battery. The device integrates an X-ray generator, controller, and beam-limiting device into a single enclosure. It uses a high-frequency inverter to convert DC to AC for X-ray generation. The operator, typically a dentist or dental technician, uses the device to penetrate tooth and jaw structures, producing images on intraoral receptors (digital sensors or film). The system features a control panel and remote controller for exposure settings. By providing a portable source of radiation, it allows for diagnostic imaging in clinical settings, potentially improving workflow efficiency for dental practitioners.

## Clinical Evidence

Bench testing only. The device underwent performance evaluation and EMC testing in accordance with standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32. Results confirmed compliance with safety and performance requirements.

## Technological Characteristics

Portable hand-held X-ray system; 25.2V DC Li-Polymer battery; high-frequency inverter; 60kV fixed voltage; 2mA fixed current; 0.8mm focal spot; constant potential (DC) waveform; 20cm source-to-skin distance; 6cm cone diameter. Complies with IEC 60601-1, 60601-1-2, 60601-2-7, 60601-2-28, and 60601-2-32.

## Regulatory Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

## Predicate Devices

- Portable X-Ray System (Model: PORT-X II) ([K063121](/device/K063121.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0&2 f75

Image /page/0/Picture/1 description: The image shows the word "EXARO" in a stylized font. A line crosses through the "X" in the word. The letters are bold and have a textured appearance.

FEB - 5 2009

#### Exhibit 5 510(k) Summary

Portable X-Ray System / Model: DXR-1

1.

గా

## Submitter and US Official Correspondent

| Submitter:              | EXARO Co., Ltd.                       |
|-------------------------|---------------------------------------|
| Address:                | #1201 World Meridian Venture Center I |
|                         | 426-5 Gasan-Dong, Gumcheon-Gu         |
|                         | Seoul, 153-803, Korea                 |
| Official Correspondent: | Shin Kuk Yoo, BD Manager              |
| Telephone No.:          | 801-303-7440 (Ext. 202)               |
| Fax No.:                | 801-303-7455                          |
| Email:                  | skyone@LSKBioPartners.com             |

- Establishment Registration Number The firm will be registered and listed prior to 2. distribution of medical device.
Device Information Proprietary/Trade Name: Portable X-Ray System (Model: DXR - 1) Common/Usual Name: Portable X-Ray System Classification Name: Extraoral Source X-Ray System Product Code: EHD Device Class: Class II per regulation 21 CFR 872.1800

- Equivalent Legally Marketed Device 4.

| Manufacturer:   | GENORAY Co., Ltd.                                                              |
|-----------------|--------------------------------------------------------------------------------|
| Device Name:    | Portable X-Ray System (Model: PORT-X II)                                       |
| 510(k) Number:  | K063121 (Decision Date - Jan. 11, 2007)                                        |
| Classification: | Extraoral Source X-Ray System: EHD, Class II per<br>regulation 21 CFR 872.1800 |

#### Description of the Device ട്.

DXR-1, a portable dental X-ray system, operates on 25.2V DC supplied by a rechargeable Li-Polymer battery pack. The X-ray tubehead, controls and power source

{1}------------------------------------------------

are assembled into a single hand-held enclosure. The package includes a battery charger.

The potable X-ray system, DXR-1, being composed of X-ray generator, controller, and beam limiting device is designed to diagnose tooth and jaw through generated and controlled X-ray. The mechanical principle of DXR-1 starts from the generation of X-ray by high voltage electricity, which in turn penetrates tooth and jaw area after flowing through X-ray tube and produces X-ray images on X-ray receptors (i.e. chemical film or digital sensor)

This device contains a high frequency inverter that converts direct to alternating current, X-ray tubehead, electrical protective devices, and other elements. The DXR-1 produces sharp and clear images and prevents patients and dentists from radiation exposure with utilizing small dose of radiation.

#### Indications for use 6.

The portable X-ray system (Model: DXR - 1) is intended to be used by trained dentists and dental technicians as extraoral X-ray source for producing diagnostic X-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

#### Safety, EMC and Performance Data 7.

The compliance of DXR-1 will satisfy the applicable requirements of the Underwriters Laboratories Standard for Safety-UL/IEC 60601-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32. All required documents and reports will be submitted to the appropriate oversight agency to establish compliance with the applicable requirements.

EMS test was performed by SGS Testing Korea Co., Ltd. for DXR-1 in accordance with Standard EN/IEC 60601-1-2. All test results were complied with the requirements.

#### 8. Safety and Effectiveness, comparison to Predicate

The result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate devices.

{2}------------------------------------------------

### Substantial Eguivalence Chart တံ

. . . . . .

| Company                        | GENORAY Co. Ltd.                                                                                                                                                                                                                                | EXARO Co., Ltd.                                                                                                       |      |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------|
| Model                          | PORT-X II                                                                                                                                                                                                                                       | DXR-1                                                                                                                 |      |
| 510(k) No                      | K063121                                                                                                                                                                                                                                         | Not assigned yet                                                                                                      |      |
| Energy                         | Rechargeable 22.2V, DC Lithium                                                                                                                                                                                                                  | Rechargeable 25.2V, DC                                                                                                |      |
| Source                         | Polymer battery pack                                                                                                                                                                                                                            | Lithium Polymer Battery pack                                                                                          |      |
| Expose Time                    | 0.01-2.0 seconds, 0.01 increments                                                                                                                                                                                                               | 0.01-1.6 seconds, 0.01 increments                                                                                     |      |
| Time Accuracy                  | ±(10%+1ms)                                                                                                                                                                                                                                      | ±(10%+1ms)                                                                                                            |      |
| Heat Capacity                  | 8.5 KHU                                                                                                                                                                                                                                         | 8.5 KHU                                                                                                               |      |
| Power Output                   | 100 W                                                                                                                                                                                                                                           | 180 W                                                                                                                 |      |
| mA                             | 2mA fixed                                                                                                                                                                                                                                       | 2mA fixed                                                                                                             |      |
| kVp                            | 60kV fixed                                                                                                                                                                                                                                      | 60kV fixed                                                                                                            |      |
| Focal Spot                     | 0.8 mm                                                                                                                                                                                                                                          | 0.8 mm                                                                                                                |      |
| Wave Form                      | Constant Potential (DC)                                                                                                                                                                                                                         | Constant Potential (DC)                                                                                               |      |
| Safety, EMC and<br>performance | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-7, IEC 60601-2-28, IEC<br>60601-2-32                                                                                                                                                                 | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-7, IEC 60601-2-28, IEC<br>60601-2-32                                       |      |
|                                | Source to skin<br>distance                                                                                                                                                                                                                      | 20cm                                                                                                                  | 20cm |
|                                | Cone Diameter                                                                                                                                                                                                                                   | 7cm                                                                                                                   | 6 cm |
| User Interface                 | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor<br>with display.                                                                                                                           | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor<br>with display. |      |
| Exposure switch                | Control panel and remote controller                                                                                                                                                                                                             | Control panel and remote<br>controller                                                                                |      |
| Tubehead<br>Mounting           | Yes                                                                                                                                                                                                                                             | Yes                                                                                                                   |      |
| Intended Use                   | Intended to use by trained dentists and dental technicians as an extra-<br>oral x-ray source for producing diagnostic x-ray images using intra-oral<br>image receptors or film. It is intended to use for both adult and pediatric<br>subjects. |                                                                                                                       |      |

.

:

.

.

 : 

. .

:

.

9.

. . . . . . . . . . .

:

: 六六十四十四

:

·

、

.

·

.

:

:

{3}------------------------------------------------

#### 10. · Conclusion

.

:

:

:

:

. . .

In reference to the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807 and the comparison information provided substantial equivalent chart above, the EXARO Co., Ltd., believes that the portable X-ray system (Model: DXR-1) is safe and effective and substantially equivalent to the predicate device (Model: PORT-X II).

and the comments of the comments of the comments of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the con

. .

:

:

·

:

.

:

:

:

.

、

:

.

.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 5 2009

EXARO Co., Ltd. % Mr. Shin Kuk Yoo LSK BioPartners, Inc. 215 South State Street, Suite 100B SALT LAKE CITY UT 84111

Re: K082875

Trade/Device Name: Portable X-Ray System (Model: DXR-1) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD and MUH Dated: January 29, 2009

Received: February 2, 2009

## Dear Mr. Yoo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Janine M. Morris

anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health 

{6}------------------------------------------------

Exhibit 4 Indications for Use

lications for use

510(k) number (if known): _KO82 875

Device Name: Portable X-Ray System (Model: DXR-1) Indications for Use:

> The Portable X-ray System (Model: DXR-1) is intended to be used by trained dentists and dental technicians as extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use

(Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helen Lewin

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 0810875 510(k) Number

15/336

---

**Source:** [https://fda-staging.innolitics.com/device/K082875](https://fda-staging.innolitics.com/device/K082875)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
