The Portable X-ray System (Model: DXR-1) is intended to be used by trained dentists and dental technicians as extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.
Device Story
The DXR-1 is a portable, hand-held dental X-ray system. It operates on a 25.2V DC rechargeable Li-Polymer battery. The device integrates an X-ray generator, controller, and beam-limiting device into a single enclosure. It uses a high-frequency inverter to convert DC to AC for X-ray generation. The operator, typically a dentist or dental technician, uses the device to penetrate tooth and jaw structures, producing images on intraoral receptors (digital sensors or film). The system features a control panel and remote controller for exposure settings. By providing a portable source of radiation, it allows for diagnostic imaging in clinical settings, potentially improving workflow efficiency for dental practitioners.
Clinical Evidence
Bench testing only. The device underwent performance evaluation and EMC testing in accordance with standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32. Results confirmed compliance with safety and performance requirements.
Indicated for use by trained dentists and dental technicians as an extraoral X-ray source for diagnostic imaging in adult and pediatric patients using intraoral image receptors.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
Predicate Devices
Portable X-Ray System (Model: PORT-X II) (K063121)
Submission Summary (Full Text)
{0}------------------------------------------------
K0&2 f75
Image /page/0/Picture/1 description: The image shows the word "EXARO" in a stylized font. A line crosses through the "X" in the word. The letters are bold and have a textured appearance.
FEB - 5 2009
#### Exhibit 5 510(k) Summary
Portable X-Ray System / Model: DXR-1
1.
గా
## Submitter and US Official Correspondent
| Submitter: | EXARO Co., Ltd. |
|-------------------------|---------------------------------------|
| Address: | #1201 World Meridian Venture Center I |
| | 426-5 Gasan-Dong, Gumcheon-Gu |
| | Seoul, 153-803, Korea |
| Official Correspondent: | Shin Kuk Yoo, BD Manager |
| Telephone No.: | 801-303-7440 (Ext. 202) |
| Fax No.: | 801-303-7455 |
| Email: | skyone@LSKBioPartners.com |
- Establishment Registration Number The firm will be registered and listed prior to 2. distribution of medical device.
Device Information Proprietary/Trade Name: Portable X-Ray System (Model: DXR - 1) Common/Usual Name: Portable X-Ray System Classification Name: Extraoral Source X-Ray System Product Code: EHD Device Class: Class II per regulation 21 CFR 872.1800
- Equivalent Legally Marketed Device 4.
| Manufacturer: | GENORAY Co., Ltd. |
|-----------------|--------------------------------------------------------------------------------|
| Device Name: | Portable X-Ray System (Model: PORT-X II) |
| 510(k) Number: | K063121 (Decision Date - Jan. 11, 2007) |
| Classification: | Extraoral Source X-Ray System: EHD, Class II per<br>regulation 21 CFR 872.1800 |
#### Description of the Device ട്.
DXR-1, a portable dental X-ray system, operates on 25.2V DC supplied by a rechargeable Li-Polymer battery pack. The X-ray tubehead, controls and power source
{1}------------------------------------------------
are assembled into a single hand-held enclosure. The package includes a battery charger.
The potable X-ray system, DXR-1, being composed of X-ray generator, controller, and beam limiting device is designed to diagnose tooth and jaw through generated and controlled X-ray. The mechanical principle of DXR-1 starts from the generation of X-ray by high voltage electricity, which in turn penetrates tooth and jaw area after flowing through X-ray tube and produces X-ray images on X-ray receptors (i.e. chemical film or digital sensor)
This device contains a high frequency inverter that converts direct to alternating current, X-ray tubehead, electrical protective devices, and other elements. The DXR-1 produces sharp and clear images and prevents patients and dentists from radiation exposure with utilizing small dose of radiation.
#### Indications for use 6.
The portable X-ray system (Model: DXR - 1) is intended to be used by trained dentists and dental technicians as extraoral X-ray source for producing diagnostic X-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.
#### Safety, EMC and Performance Data 7.
The compliance of DXR-1 will satisfy the applicable requirements of the Underwriters Laboratories Standard for Safety-UL/IEC 60601-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32. All required documents and reports will be submitted to the appropriate oversight agency to establish compliance with the applicable requirements.
EMS test was performed by SGS Testing Korea Co., Ltd. for DXR-1 in accordance with Standard EN/IEC 60601-1-2. All test results were complied with the requirements.
#### 8. Safety and Effectiveness, comparison to Predicate
The result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate devices.
{2}------------------------------------------------
### Substantial Eguivalence Chart တံ
. . . . . .
| Company | GENORAY Co. Ltd. | EXARO Co., Ltd. | |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------|
| Model | PORT-X II | DXR-1 | |
| 510(k) No | K063121 | Not assigned yet | |
| Energy | Rechargeable 22.2V, DC Lithium | Rechargeable 25.2V, DC | |
| Source | Polymer battery pack | Lithium Polymer Battery pack | |
| Expose Time | 0.01-2.0 seconds, 0.01 increments | 0.01-1.6 seconds, 0.01 increments | |
| Time Accuracy | ±(10%+1ms) | ±(10%+1ms) | |
| Heat Capacity | 8.5 KHU | 8.5 KHU | |
| Power Output | 100 W | 180 W | |
| mA | 2mA fixed | 2mA fixed | |
| kVp | 60kV fixed | 60kV fixed | |
| Focal Spot | 0.8 mm | 0.8 mm | |
| Wave Form | Constant Potential (DC) | Constant Potential (DC) | |
| Safety, EMC and<br>performance | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-7, IEC 60601-2-28, IEC<br>60601-2-32 | IEC 60601-1, IEC 60601-1-2, IEC<br>60601-2-7, IEC 60601-2-28, IEC<br>60601-2-32 | |
| | Source to skin<br>distance | 20cm | 20cm |
| | Cone Diameter | 7cm | 6 cm |
| User Interface | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor<br>with display. | Exposure time: up, down.<br>Selection buttons of parts of teeth,<br>adult and child, film and sensor<br>with display. | |
| Exposure switch | Control panel and remote controller | Control panel and remote<br>controller | |
| Tubehead<br>Mounting | Yes | Yes | |
| Intended Use | Intended to use by trained dentists and dental technicians as an extra-<br>oral x-ray source for producing diagnostic x-ray images using intra-oral<br>image receptors or film. It is intended to use for both adult and pediatric<br>subjects. | | |
.
:
.
.
:
. .
:
.
9.
. . . . . . . . . . .
:
: 六六十四十四
:
·
、
.
·
.
:
:
{3}------------------------------------------------
#### 10. · Conclusion
.
:
:
:
:
. . .
In reference to the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807 and the comparison information provided substantial equivalent chart above, the EXARO Co., Ltd., believes that the portable X-ray system (Model: DXR-1) is safe and effective and substantially equivalent to the predicate device (Model: PORT-X II).
and the comments of the comments of the comments of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the con
. .
:
:
·
:
.
:
:
:
.
、
:
.
.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 5 2009
EXARO Co., Ltd. % Mr. Shin Kuk Yoo LSK BioPartners, Inc. 215 South State Street, Suite 100B SALT LAKE CITY UT 84111
Re: K082875
Trade/Device Name: Portable X-Ray System (Model: DXR-1) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD and MUH Dated: January 29, 2009
Received: February 2, 2009
## Dear Mr. Yoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Exhibit 4 Indications for Use
lications for use
510(k) number (if known): _KO82 875
Device Name: Portable X-Ray System (Model: DXR-1) Indications for Use:
> The Portable X-ray System (Model: DXR-1) is intended to be used by trained dentists and dental technicians as extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.
Prescription Use _ V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helen Lewin
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 0810875 510(k) Number
15/336
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.