K082858 · Alcavis International, Inc. · FQH · May 27, 2009 · General Hospital
Device Facts
Record ID
K082858
Device Name
EXSEPT WC ANTIMICROBIAL SKIN AND WOUND CLEANSER
Applicant
Alcavis International, Inc.
Product Code
FQH · General Hospital
Decision Date
May 27, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OTC: ExSept WC Skin and Wound Cleanser is intended for cleansing and removal of dirt, debris and foreign material from minor skin abrasions, minor lacerations, minor irritations, minor cuts, and intact skin. Professional Use: Under supervision of healthcare professionals, ExSept WC Skin and Wound Cleanser is intended for use for mechanical cleansing, debridement and removal of foreign material and debris from exudating and/or dirty wounds, such as stage I-IV pressure ulcers, diabetic foot ulcers, post surgical wounds, first and second degree burns, grafted and donor sites, and catheter exit sites.
Device Story
ExSept WC Skin and Wound Cleanser is a liquid solution used for mechanical cleansing of wounds. It removes dirt, debris, and foreign material from epidermal and dermal wounds. Supplied in pump-spray or pour bottles; applied by patients (OTC) or healthcare professionals (clinical setting). The solution acts mechanically to clear wound beds and surrounding skin. Benefits include effective debridement and removal of contaminants, facilitating wound care management.
Clinical Evidence
Bench testing only. In-vitro and in-vivo studies performed to demonstrate efficacy, safety, and biocompatibility. Stability studies confirm a 30-month shelf life at room temperature.
Technological Characteristics
Clear to slight yellow liquid solution with slight chlorine odor. Supplied in 100ml, 200ml, 250ml, and 500ml containers. Mechanical cleansing principle. No software or electronic components.
Indications for Use
Indicated for OTC use on minor skin abrasions, lacerations, irritations, cuts, and intact skin. Indicated for professional use on exudating or dirty wounds including stage I-IV pressure ulcers, diabetic foot ulcers, post-surgical wounds, first/second degree burns, grafted/donor sites, and catheter exit sites.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Anasept Animicrobial Skin and Wound Cleanser (K073547)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 2
K082858 ExSept WC Skin and Wound Cleanser
# 510(k) Summary
27 2009 MAY
| Sponsor: | Alcavis International<br>8322 Helgerman Court<br>Gaithersburg, MD 20877 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gary J. Mishkin<br>Vice President, Research and Development<br>Tele: (301) 330-7597<br>Fax: (301) 330-6432 |
| Date Prepared: | September 25, 2008 |
| Device Name: | Proprietary Name: ExSept WC Skin and Wound Cleanser<br>Common Name: Wound Cleanser<br>Classification Name: Dressing, Wound, Drug |
| Predicate Devices: | ExSept WC Wound Cleaner, K061467<br>SilvaKlenz Skin and Wound Cleanser, K063069<br>Dermacyn Wound Cleanser, K042729<br>Anasept Animicrobial Skin and Wound Cleanser, K073547 |
Device Description: ExSept WC Skin and Wound Cleanser is a clear to slight yellow, slight chlorine odor, solution which is intended for the mechanical cleansing of debris and foreign material from exudating/dirty epidermal and dermal wounds. It is supplied in a 200 ml pump-spray and 100 ml, 250 ml and 500 ml pour bottles. ExSept WC Skin and Wound Cleanser has a 30 month shelf life.
### Intended Use:
1
OTC: ExSept WC Skin and Wound Cleanser is intended for cleansing and removal of dirt, debris and foreign material from minor skin abrasions, minor lacerations, minor irritations, minor cuts, and intact skin.
Professional Use: Under supervision of healthcare professionals, ExSept WC Skin and Wound Cleanser is intended for use for mechanical cleansing, debridement and removal of foreign material and debris from exudating and/or dirty wounds, such as stage I-IV pressure ulcers, diabetic foot ulcers, post surgical wounds, first and second degree burns, grafted and donor sites, and catheter exit sites.
- Substantial Equivalence: ExSept WC Skin and Wound Cleanser is substantially equivalent in the cleansing functions and intended uses to the predicate devices ExSept WC, SilvaKlenz, Dermacyn
{1}------------------------------------------------
K082858
Page 2 of 2
K082858 ExSept WC Skin and Wound Cleanser
and Anasept. All predicate devices use a mechanical action to remove foreign material and other debris from the wounds and surrounding skin.
- Testing: Numerous in-vitro and in-vivo studies have been performed to demonstrate the efficacy, safety and biocompatibility of the ExSept WC Skin and Wound Cleanser for the indications for use. Stability studies show a shelf life of 30 months for the ExSept WC Skin and Wound Cleanser, when stored at room temperature.
Conclusion: Based upon the information in the 510(k) submission, ExSept WC Skin and Wound Cleanser will perform as intended as a wound cleanser. ExSept WC Skin and Wound Cleanser is substantially equivalent in its actions and functions as the marketed devices ExSept WC, SilvaKlenz, Dermacyn and Anasept.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle.
## MAY 27 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alcavis International, Incorporated % Mr. Gary J. Mishkin Vice President, Research and Development 8322 Helgerman Court Gaithersburg, Maryland 20877
Re: K082858
Trade/Device Name: ExSept WC Skin and Wound Cleanser Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: II Product Code: FQH Dated: April 17, 2009 Received: April 20, 2009
Dear Mr. Mishkin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting
{3}------------------------------------------------
Page 2-Mr. Mishkin
(reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
F-R Pete D. Rummmo mfit
Mark N. Mallem O.K.
Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
K082858 ExSept WC Skin and Wound Cleanser
# Indications for Use
### 510(k) Number (if known):K082858
Device Name: ExSept WC Skin and Wound Cleanser
#### Indications For Use:
- OTC: ExSept WC Skin and Wound Cleanser is intended for cleansing and removal of dirt, debris and foreign material from minor skin abrasions, minor lacerations, minor irritations, minor cuts, and intact skin.
- Professional Use: Under supervision of healthcare professionals, ExSept WC Skin and Wound Cleanser is intended for use for mechanical cleansing, debridement and removal of foreign material and debris from exudating and/or dirty wounds, such as stage I-IV pressure ulcers, diabetic foot ulcers, post surgical wounds, first and second degree burns, grafted and donor sites, and catheter exit sites.
Prescription Use X ( Part 21 CFR 801 Subpart D) Over-The-Counter Use _ X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Keane for MXM
(Division Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K062656
Page 1 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.