InterGro DBM products (Putty, Paste, Plus) are to be used for filling bony voids or gaps in the extremities and pelvis that are not intrinsic to the bony stability of the structure, and as an autograft extender in the spine. InterGro Plus may also be used as a bone void filler in the spine (posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. InterGro DBM may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm2. The amount of InterGro DBM products to be used should be based on the type of procedure and size of the graft site.
Device Story
InterGro DBM is a resorbable, osteoconductive, and osteoinductive bone graft substitute derived from human allograft demineralized cortical bone matrix (DBM). It is processed via granulation and lyophilization. Some versions incorporate calcium salt granules for radiopacity and structural strength. The DBM is suspended in a biocompatible, semi-viscous lipid carrier. Provided ready-to-use in putty, paste, or plus consistencies for single-patient application. Used by surgeons to fill bony voids or gaps in extremities, pelvis, spine, and craniofacial sites. The device resorbs over time, replaced by host bone during healing. Osteoinductive potential is verified in vitro via C2C12 myoblast alkaline phosphatase assay, correlated with athymic rat muscle implantation models. Viral inactivation of donor tissue is validated through a panel of representative viruses.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, viral inactivation validation, and animal model performance evaluation (radiographic and histological). Osteoinductive potential assessed via in vitro C2C12 bioassay (correlated with rat model, r=0.88, p<0.0005).
Technological Characteristics
Resorbable bone graft substitute; components: demineralized cortical bone matrix (DBM), semi-viscous lipid carrier, and optional calcium salt granules. Processing: aseptic, lyophilized. Standards: AATB, ASTM, and FDA guidance documents used for material certification and performance testing.
Indications for Use
Indicated for patients requiring bone void filling in extremities, pelvis, and spine (posterolateral); autograft extender in spine; craniofacial defects/craniotomies ≤ 25cm². Applicable to surgically created or traumatic osseous defects.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
DynaGraft II
DBX
InterGro DBM
Submission Summary (Full Text)
{0}------------------------------------------------
K082793
# APR – 2 2009
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92
| Sponsor: | Biomet Spine/Biomet Trauma<br>100 Interpace Parkway<br>Parsippany NJ 07054 |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 2242816 |
| Manufacturing Site | Interpore Cross International<br>181 Technology Dr<br>Irvine CA 92618 |
| Manufacturing Site<br>Registration<br>Number: | 2029012 |
| Contact: | Kathleen Ideo<br>Regulatory Associate<br>Interpore Cross International<br>181 Technology Dr<br>Irvine CA 92618<br>(949) 453-3200 |
| Date Prepared: | February 17, 2009 |
| Trade/Proprietary Name: | InterGro DBM (Putty, Paste, Plus) |
| Common/Usual Name: | DBM Bone Graft Substitute |
| Classification Name: | Bone Grafting Material |
| Device Classification: | Class II |
| Regulation Number | 21 CFR 888.3045 |
| Product Codes | MQV, MBP, GXP |
| Predicate Device: | DynaGraft II, DBX, InterGro DBM |
| Device Formulation: | InterGro DBM (Putty, Paste, Plus) |
| Performance: | Performance<br>standards<br>applicable<br>DBM-based<br>to |
¹⁴
pg 1 of 3
{1}------------------------------------------------
products have not been published by the FDA. Interpore Cross International intends to manufacture and package this device according to the regulations and standards that are appropriate to the risk that Class II devices reasonably present. Voluntary performance standards, such as materials certifications, in-house SOP's, FDA Guidance Documents, AATB Standards and/or ASTM Standards are used as appropriate.
InterGro DBM is a resorbable, osteoconductive, and Device Description: osteoinductive bone graft substitute that resorbs and is replaced with bone during the healing process. Its main component, demineralized cortical bone matrix (DBM), is derived from donor human tissue (allograft bone) and contains various growth factors including osteoinductive proteins. The DBM has been granulated, lyophilized and aseptically processed. In some versions of the product, calcium salt granules shall be incorporated to provide additional radiopacity, osteoconduction, and enhanced structural strength. The carrier for InterGro DBM is a resorbable, biocompatible, semi-viscous lipid. InterGro DBM is provided ready-to-use in various physical consistencies. It is packaged in various sizes by volume for single patient use.
> InterGro DBM products (Putty, Paste, Plus) are to be used for filling bony voids or gaps in the extremities and pelvis that are not intrinsic to the bony stability of the structure, and as an autograft extender in the spine. InterGro Plus may also be used as a bone void filler in the spine (posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. InterGro DBM may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm2. The amount of InterGro DBM products to be used should be based on the type of procedure and size of the graft site.
Indications for Use:
\$\frac{+5}{pg}\$ 2 of 3
{2}------------------------------------------------
Substantial Equivalence Viral Inactivation Validation: The methods for processing the DBM contained in InterGro DBM were Information evaluated for their viral inactivation potential. A select panel of viruses representing various viral types, sizes, shapes, and genomes were evaluated. The viral inactivation testing demonstrated suitable viral inactivation potential of the processing methods for a wide range of potential human viruses.
> Osteoinductive Potential: Each lot of DBM incorporated into InterGro DBM is assayed for its osteoinductive potential. The assay measures the alkaline phosphatase production of a myoblast cell line (C2C12) in the presence of human DBM compared to positive and negative controls (osteoinductive index). Results of the assay have been correlated with results from implantation of DBM into athymic rat muscle, which demonstrated a correlation coefficient of 0.88 (p<0.0005) and accurately predicted the in vivo osteoinductivity in 20 donor lots. '
> The combination of DBM, the carrier and, in some formulations, ceramic granules has not been evaluated for osteoinductivity; therefore it is unknown to what extent the formulation components may alter the osteoinductive character of the DBM. Additionally, it is unknown how osteoinductivity of the DBM component, measured via the in-vitro "C2C12" bioassay will correlate with human clinical performance of InterGro DBM products.
> 1Han B. Tang B. and Nimni M. Quantitative and Sensitive in vitro Assay for Osteoinductive Activity of Demineralized Bone Matrix. J Ortho Res, 2003, 21:648-54.
> Product Performance Testing: Performance of InterGro DBM was evaluated in an animal model by radiographic and histological methods.
> The safety and effectiveness of InterGro DBM is adequately supported by the substantial equivalence information, materials data, and testing results provided within this premarket notification.
> InterGro DBM was found to be substantially equivalent to the predicate devices based on the intended use, base materials, select performance properties, and use of a handling material.
Conclusion: .
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces to the left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Biomet Spine/Biomet Trauma % Ms. Kathleen Ideo 100 Interpace Parkway Parsippany, New Jersey 07054
Re: K082793
Trade/Device Name: InterGro® DBM (Putty, Paste, Plus) Regulation Number: 21 CFR 888,3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV, MBP, GXP Product March 25, 2009 Received: March 26, 2009
Dear Ms. Ideo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
APR - 2 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 -- Ms. Kathleen Ideo
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K082793
Device Name: InterGro® DBM (Putty, Paste, Plus)
Indications for Use:
InterGro DBM products (Putty, Paste, Plus) are to be used for filling bony voids or gaps in the extremities and pelvis that are not intrinsic to the bony stability of the structure, and as an autograft extender in the spine. InterGro Plus may also be used as a bone void filler in the spine (posterolateral spine). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to bone. InterGro DBM may also be used for filling craniofacial defects and craniotomies that are no larger than 25cm². The amount of InterCro DBM products to be used should be based on the type of procedure and size of the graft site.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Hubert Dinhup
(Division Sign-Off) for MXH
Division of Gene Division
and Neurological Devices
510(k) Number K082793
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.