The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporeal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.
Device Story
Disposable extracorporeal blood tubing set; connects patient vascular access to hemodialyzer. Components include pump tubing, air trap chambers, transducer filters, pressure monitoring lines, infusion lines, and clamps. Used during hemodialysis procedures with Nikkiso DBB-05 Hemodialysis Delivery System. Operates by facilitating blood flow, debris capture, medication/fluid infusion, blood sampling, and pressure monitoring. Used in clinical settings by healthcare professionals. Provides physical conduit for blood circuit; enables pressure monitoring via integral transducer protectors. Benefits patient by maintaining sterile, functional extracorporeal circuit for dialysis treatment.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, hemocompatibility, genotoxicity per ISO10993), sterility, pyrogenicity, and mechanical performance (blood pathway volume, flow dynamics, pump segment performance, occlusive clamping) per ANSI/AAMI RD17: 2007.
Indicated for patients requiring hemodialysis to provide extracorporeal blood access. No specific age or gender restrictions stated. Contraindications not explicitly listed.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Nipro® Blood Tubing Set with Transducer Protectors and Priming Set (K010264)
Reference Devices
Nikkiso DBB-05 Hemodialysis Delivery System (K061519)
{0}------------------------------------------------
K082719
NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)
# Section 6- 510(k) Summary
MAR 3 0 2009
- a. Company name, address NIKKISO CO., LTD. 43-2, Ebisu 3-Chome, Shibuya-ku Tokyo, 150-8677, Japan
- b. Contact Masashi Yoshida Manager Regulatory Affairs
- c. Date prepared
August 18, 2008
### d. Name of device
| Trade Name: | NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS |
|----------------------|--------------------------------------------------------|
| Common Name: | Hemodialysis system and accessories |
| Classification Name: | Hemodialysis system and accessories |
# e. Predicate devices
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to:
| 510(k): | K010264 |
|---------------|--------------------------------------------------------|
| Trade name: | Nipro® Blood Tubing Set with Transducer Protectors and |
| | Priming Set |
| Product code: | FIB, FJK |
### f. Description of the device
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS includes arterial and venous dialysis blood tubing (non-implanted blood access device) as described in 21 CFR 876.5820.
The devices are packaged together for convenient use during for hemodialysis procedures. Three models of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS (AV06A-P, AV06B-P, AV06C-P) are being manufactured for use with the Nikkiso DBB-05 Hemodialysis Delivery System (K061519).
The components of the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS include pump tubing, air trap chambers, transducer filters, pressure monitoring lines,
NIKKISO CO., LTD.
{1}------------------------------------------------
K082719
NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)
infusion lines for saline, access ports and tubing clamps which are all used to pump blood, retain and capture blood debris, infuse medications or fluids, sample blood, monitor pressure, and make connections to other devices.
Major materials used for the various components are Polyvinyl chloride (PVC), polycarbonate (PC), polypropylene (PP). Materials in direct contact with blood are Polycarbonate (PC), Polyvinyl chloride (PVC), Silicone, Polypropylene (PP) and Polytetrafluoroethylene (PTFE).
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are restricted for sale by or on the orders of a physician.
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are packaged sterile and labeled for single use only and non-pyrogenic. Sterility and non-pyrogenicity were confirmed by sterilization validation and pyrogenicity testing.
Biocompatibility tests were performed in compliance with ISO10993 requirements and include acute toxicity, intracutaneous reactivity, hemolysis testing, and implantation testing. The results of biocompatibility testing did not raise any question regarding biocompatibility, and support the equivalence to the predicate device.
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are designed for use with the Nikkiso DBB-05 Hemodialysis Delivery System (K061519).
Image /page/1/Figure/8 description: The image shows a diagram of an arterial and venous circuit. The arterial circuit includes an access port, shunt connector (red), air trap chamber, pump segment, and transducer protector line for pressure monitor. The venous circuit includes a shunt connector (blue), air trap chamber, access port, and transducer protector line for pressure monitor. The diagram also shows the connection to the hemodializer (red and blue).
Figure 1. NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS (AV06A-P)
NIKKISO CO., LTD.
{2}------------------------------------------------
K082719
NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)
# g. Indications for Use
#### Indication for Use
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporeal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.
# h. Statement of substantial equivalence
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro® Blood Tubing Set for Hemodialysis with Transducer Protectors and Priming Set (K010264) (Hereinafter "the Nipro tubing set").
Table 1 compares NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS to the Nipro tubing set. Both tubing lines have the same intended use, same configurations and same major materials. Both of the tubing lines are sterile and labeled as single use only and non-pyrogenic. Tests results from biocompatibility tests for the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS in compliance to ANSI/AAMI RD17: 2007 "Cardiovascular implants and artificial organs - Extracorporeal blood circuit for hemodialyzers, hemodiafilters, and hemofitters" (Hereinafter "RD17") and ISO10993 did not raise any concerns regarding biocompatibility. Nikkiso performed all tests required in RD17 for both NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS and the Nipro tubing set.
As shown in Table 1, both tubing lines met the requirements in RD17 including sterility, pyrogenicity and mechanical characteristic tests.
Based on the above considerations, Nikkiso concludes that the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro tubing set.
{3}------------------------------------------------
K082719
# nikkline blood tubing lines with transducer protectors PREMARKET NOTIFICATION 510(k)
# Table 1. Comparison table
.
・
| No. | Item | Proposed Device | Predicate Device |
|-----|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Indication for use | The NIKKLINE BLOOD TUBING LINES WITH<br>TRANSDUCER PROTECTORS are<br>disposable bloodlines intended to<br>provide extracorporeal access to a<br>patient's blood during hemodialysis.<br>The compatibility of available<br>configurations is the responsibility of<br>the physician or other licensed<br>practitioner. | The Nipro Blood Tubing Set with<br>Transducer Protector and Priming Sets<br>are disposable blood tubing lines<br>intended to provide extracorporeal<br>access to the patient's blood during<br>hemodialysis. The compatibility of<br>available configurations is the<br>responsibility of the physician in<br>charge |
| 2 | Configuration | The NIKKLINE BLOOD TUBING LINES<br>WITH TRANSDUCER PROTECTORS are<br>composed of various components such<br>as tubing, luer locks, connectors,<br>tubing clamps and air trap chambers<br>that are configured for connecting a<br>patient's vascular access system to a<br>dialyzer. The tubing lines components<br>are pre-configured and packaged as an<br>arterial line, a venous line, an infusion<br>lines, a saline line and pressure<br>monitoring lines for use with the<br>Nikkiso DBB-05 Hemodialysis<br>Delivery System (K061519). | The Nipro blood tubing lines are<br>composed of various components such<br>as tubing, luer locks, connectors,<br>clamps and drip chambers that are<br>configured for connecting a patient's<br>vascular access system to a dialyzer.<br>The tubing lines components are<br>pre-configured and packaged as an<br>arterial line, a venous line, an infusion<br>line, a saline line and pressure<br>monitoring lines for use with various<br>dialysis systems such as Fresenius,<br>Baxter, Cobe, Althin, etc. |
| 3 | Biological Safety | Meets ANSI/AAMI RD17: 2007 and<br>ISO10993.<br>Tests results from biocompatibility<br>tests for The NIKKLINE BLOOD<br>TUBING LINES WITH TRANSDUCER<br>PROTECTORS did not raise any<br>concerns regarding biocompatibility<br>including:<br>• Cytotoxicity<br>• Sensitization<br>• Irritation or intracutaneous reactivity<br>• Acute Systemic toxicity<br>• Hemocompatibility<br>• Genotoxicity | Meets ANSI/AAMI RD17 and<br>ISO10993. |
| 4 | Sterility<br>-method | Steam sterilization | Ethylene Oxide sterilization |
| | -SAL 10 $^{-6}$ | Assured SAL 10 $^{-6}$ by validation met<br>ISO11134 and EN554. | Assured SAL 10 $^{-6}$ by validation met<br>ISO11135-1 |
| 5 | Pyrogenicity | Non- pyrogenic | Non-pyrogenic |
| 6 | Single use or reuse | Single use only. | Single use only. |
| 7 | Structural integrity | Meets 4.4.1 Structural integrity AAMI<br>RD17: 2007. | Meets 4.4.1 Structural integrity AAMI<br>RD17: 2007. |
| 8 | Connection to<br>hemodialyzer, hemodiafilter<br>or hemofilter | Meets 4.4.2 Connection to<br>hemodialyzer, hemodiafilter or<br>hemofilter AAMI RD17: 2007. | Meets 4.4.2 Connection to<br>hemodialyzer, hemodiafilter or<br>hemofilter AAMI RD17: 2007. |
| No. | Item | Proposed Device | Predicate Device |
| 9 | Connection to vascular access device | NIKKLINE BLOOD TUBING LINES WITH<br>TRANSDUCER PROTECTORS<br>Meets 4.4.3 Connection to vascular<br>access device AAMI RD17: 2007. | The Nipro Tubing Set (K010264)<br>Meets 4.4.3 Connection to vascular<br>access device AAMI RD17: 2007. |
| 10 | Connection to ancillary components | Meets 4.4.4 Connection to ancillary<br>components AAMI RD17: 2007. | Meets 4.4.4 Connection to ancillary<br>components AAMI RD17: 2007. |
| 11 | Color coding | Meets 4.4.5 Color coding AAMI<br>RD17: 2007.<br>The arterial patient-connection is red,<br>the venous patient-connection is blue. | Meets 4.4.5 Color coding AAMI<br>RD17: 2007.<br>The arterial patient-connection is red,<br>the venous patient-connection is blue. |
| 12 | Needle access ports | Meets 4.4.6.1 Needle access ports<br>AMMI RD17: 2007. | Meets 4.4.6.1 Needle access ports<br>AAMI RD17: 2007. |
| 16 | Transducer protector | Meets 4.4.9.1 Integral transducer<br>protectors AAMI RD17: 2007. | Using Borla's transducer protector<br>(K000702). |
| 20 | Raw materials | PVC, PC, PP, Silicone are used for the<br>components in the product. PBT and<br>PTFE are used specifically for the<br>transducer protectors. | PVC, PE, PC, PP, POM are used for<br>the components in the product. PBT<br>and PTFE are used specifically for<br>Borla's transducer protectors. |
NIKKISO CO., LTD.
・
{4}------------------------------------------------
K082719
#### NIKKLINE BLOOD TUBING LINES WITH TRANSIDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)
#### Table 1. Comparison table (continued)
PB.I. : Polybutylene terephthalate
: Polycarbonate PC
: Polyethylene PE
- POM : Polyoxymethylene
PP : Polypropylene
: Polytetrafluoroethylene PTFE
: Polyvinyl chloride PVC
Silicone : Silicone Rubber
### Note:
Tests including blood pathway volume, blood pathway flow dynamics, pump segment performance and occlusive clamping of tubing for the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS were performed successfully in accordance with AAMI RD17: 2007. Since the Nipro Tubing Set (K010264) was cleared by FDA and subjected to the same tests by Nipro, Nikkiso concludes that the test results for the Nipro Tubing Set and the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are substantially equivalent.
{5}------------------------------------------------
K082719
NIKKIJINE BLQQID TUBING LINES WITH TRANSDUCER PROTECTORS PREMARKET NOTIFICATION 510(k)
#### i. Conclusion
Based on the above discussion and enclosed sections regarding substantial equivalence to predicate devices, NIKKISO CO., LTD. concludes that the NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS is substantially equivalent to the Nipro® Blood Tubing Set with Transducer Protectors and Priming Set (K010264) and does not raise any new questions regarding safety or effectiveness.
{6}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and body. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAR 3 0 2009
Nikkiso Co. Ltd. c/o Dr. Fumiaki Kanai MIC International 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0035 JAPAN
Re: K082719
Trade/Device Name: NIKKLINE Blood Tubing Lines with Transducer Protectors Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KOC Dated: March 11, 2009 Received: March 17, 2009
Dear Dr. Kanai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K082719
Device Name: NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS
#### Indication for Use
The NIKKLINE BLOOD TUBING LINES WITH TRANSDUCER PROTECTORS are disposable bloodlines intended to provide extracorporcal access to a patient's blood during hemodialysis. The compatibility of available configurations is the responsibility of the physician or other licensed practitioner.
Image /page/8/Picture/5 description: The image shows the words "Prescription Use" in bold font, followed by the text "(Per 21 CFR 801 Subpart D)" in a smaller font. There is a large "X" mark above a horizontal line, which is to the right of the words "Prescription Use". The text indicates that the product is intended for prescription use and is subject to the regulations outlined in 21 CFR 801 Subpart D.
AND/OR
Over-the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|----------------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and Radiological Devices, | |
| 510(k) Number | K082719 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.