K082712 · Theken Spine, LLC · MAX · Jan 2, 2009 · Orthopedic
Device Facts
Record ID
K082712
Device Name
THEKEN SPINE VU EPOD AND VU LPOD SYSTEM
Applicant
Theken Spine, LLC
Product Code
MAX · Orthopedic
Decision Date
Jan 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an intervertebral body fusion device, the Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is indicated for use with autograft only. The Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System. Degenerative disc disease (DDD) is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. When used as a vertebral body replacement (VBR) the Vu ePOD and Vu LPOD VBR System is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (ie. fracture). The Vu ePOD and Vu LPOD VBR System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device is indicated for use with autograft or allograft only. The Vu ePOD and LPOD VBR System is intended for use with supplemental internal spinal fixation such as the BodyForm Thoracolumbar Fixation System or the Coral Spinal System.
Device Story
PEEK OPTIMA-LT cages with fenestrations and toothed ridges; straight (ePOD) or curved (LPOD) geometry. Implanted surgically to provide intervertebral body fusion or vertebral body replacement. Toothed ridges engage vertebral endplates to resist rotation and migration. Used with autograft (fusion) or autograft/allograft (VBR). Requires supplemental internal spinal fixation (e.g., Coral or BodyForm systems). Restores biomechanical integrity of spinal column. Used in clinical settings by surgeons.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F-2077 and ASTM F-2267, including static axial compression, static compression shear, dynamic axial compression, dynamic compression shear, subsidence, and expulsion. No clinical data provided.
Technological Characteristics
Materials: PEEK OPTIMA-LT. Design: Cages with fenestrations and toothed ridges for endplate engagement. Geometry: Straight (ePOD) or curved (LPOD). Standards: ASTM F-2077, ASTM F-2267. Energy source: None (passive implant). Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at 1-2 contiguous levels (L2-S1), with or without Grade 1 spondylolisthesis, requiring fusion; or for patients with collapsed/unstable vertebral bodies in the thoracolumbar spine (T1-L5) due to tumor or trauma requiring vertebral body replacement (VBR).
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
### 510(k) Summary (21 CFR Part 807.92)
Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices
MAX 888.3080 (Class II) - Intervertebral body fusion device
Theken Surgical REVEAL VBR System, K050058
Theken Spine Vu ePOD System, K071031 Theken Spine Vu aPOD System, K080822
MQP 888.3060 (Class II) - Spinal intervertebral body fixation orthosis
Intervertebral body fusion device
#### A. Submitter Information
| Submitter's Name: | Theken Spine, LLC |
|-------------------------------|---------------------------------------|
| Establishment Registration #: | 1530901 |
| Address: | 1800 Triplett Blvd. Akron, Ohio 44306 |
| Telephone Number: | 330-475-8600 |
| Fax Number: | 330-773-7697 |
| Contact Person: | Dale Davison |
| Date Prepared: | 9/12/2008 |
JAN - 2 2009
#### B. Device Information
Trade Name:
Common Name:
Classification:
Predicate Devices:
Device Description:
Summary of Test Data:
Intended Use:
The Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are comprised of PEEK OPTIMA-LT cages with fenestrations and radii on all sides and toothed ridges. The toothed ridges of the concave cages engage with the superior and inferior end plates of the neighboring vertebral bodies to resist rotation and migration. The Vu ePOD is a straight shaped cage while the Vu LPOD has a curved shape. The Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices may be used individually or in a pair depending on the surgical need.
Mechanical testing of the subject device consisted of static axial compression, static compression shear, dynamic axial compression, dynamic compression shear, subsidence and expulsion. Testing was conducted in accordance with ASTM F-2077 and ASTM F-2267. The construct performed as designed and met or exceeded all product specifications.
When used as an intervertebral body fusion device the Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is indicated for use with autograft only. The Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System.
Degenerative disc disease (DDD) is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
Page 1 of 2
{1}------------------------------------------------
When used as a vertebral body replacement (VBR) the Vu ePOD and Vu LPOD VBR System is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (ie. fracture).
The Vu ePOD and Vu LPOD VBR System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device is indicated for use with autograft or allograft only.
The Vu ePOD and LPOD VBR System is intended for use with supplemental internal spinal fixation such as the BodyForm Thoracolumbar Fixation System or the Coral Spinal System.
#### C. Substantial Equivalence
The technological characteristics of the Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are similar to the predicate device REVEAL Vertebral Body Replacement (K050058), the Theken Spine Vu ePOD System (K071031) and the Theken Spine Vu aPOD System (K080822).
The subject device similarities include:
- The same indications for use as the Vu aPOD �
- The same basic design as the Vu ePOD and Vu LPOD .
- The same operating principle
- The same materials
- Used in conjunction with supplemental fixation
- The same manufacturing environment
- The same sterilization process
- The same packaging configurations
Theken Spine believes that sufficient evidence exists to reasonably conclude that the Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are substantially equivalent to the predicate device REVEAL Vertebral Body Replacement (K050058), the Theken Spine Vu ePOD System (K071031) and the Theken Spine Vu aPOD System (K080822). This is based on the design concept, the use of established, known materials, feature comparisons, mechanical testing, indications for use, pre-production quality assurance planning and engineering analysis. All implants represent a basic design concept in terms of safety and effectiveness, and differ only in minor details.
Page 2 of 2
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is simple, using a black and white color scheme.
#### Public Health Service
Food and Drug. Administration 9200 Corporate Boulevard Rockville MD. 20850
Theken Spine, LLC % Mr. Dale Davison 1800 Triplett Blvd Akron, OH 44306
JAN - 2 2009
Re: K082712
Trade/Device Name: Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: II Product Code: MAX, MQP Dated: December 30, 2008 Received: December 31, 2008
Dear Mr. Davison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 -- Mr. Dale Davison
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Mark M. Milliman
- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{4}------------------------------------------------
# Indications for Use
### 510(k) Number (if known): K082712
## Device Name: Theken Spine Vu ePOD and Vu LPOD System
Indications for Use:
When used as an intervertebral body fusion device, the Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The device is indicated for use with autograft only. The Vu ePOD and Vu LPOD Intervertebral Body Fusion Devices are intended for use with supplemental fixation such as the Coral Spinal System or the BodyForm Thoracolumbar Fixation System.
Degenerative disc disease (DDD) is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.
When used as a vertebral body replacement (VBR) the Vu ePOD and Vu LPOD VBR System is indicated for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged or otherwise unstable vertebral body due to tumor or trauma (ie. fracture).
The Vu ePOD and Vu LPOD VBR System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device is indicated for use with autograft or allograft only.
The Vu ePOD and LPOD VBR System is intended for use with supplemental internal spinal fixation such as the BodyForm Thoracolumbar Fixation System or the Coral Spinal System.
Prescription Use *X*
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
MARYANN H. LLOYD FOR M.MELKERSON
Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K082712
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.