The MedX LPT 200 single diode tethered laser is a part of the infrared console system, as per 21 CFR 890.5500. The device is used when heat is indicated for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis. The MedX LPS 200 single diode Portable Laser is a rechargeable infrared device, as per 21 CFR 890.5500. The device is used when heat is indicated for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
Device Story
MedX LPT 200 (tethered) and LPS 200 (portable) are infrared laser devices; utilize 808 nm GaAlAs semiconductor laser diodes; emit ~200 mW power. LPT 200 powered by MedX 1100 console; LPS 200 contains rechargeable battery. Both devices deliver therapeutic heat to tissue; LPT 200 supports continuous or pulsed modes (8, 146, 1000 Hz); LPS 200 operates in continuous mode. Devices feature automatic shut-off after treatment duration. Used in clinical settings to provide adjunctive therapy for pain relief and muscle relaxation. Healthcare providers apply laser to target area; thermal energy increases local blood circulation. Benefits include temporary relief of minor musculoskeletal pain and stiffness.
Clinical Evidence
No new clinical trials performed. Evidence relies on 30+ years of research/clinical experience with therapeutic heating modality and bench testing (mechanical, electrical, thermal, EMC, irradiation distribution).
Technological Characteristics
808 nm GaAlAs semiconductor laser diodes; 200 mW output; continuous or pulsed (8, 146, 1000 Hz) modes; tethered (console-powered) or portable (battery-powered) form factors; automatic shut-off timer; firmware-controlled.
Indications for Use
Indicated for patients requiring temporary increase in local blood circulation, relief of minor muscle/joint aches, pains, stiffness, muscle relaxation, and relief of muscle spasms or pain/stiffness associated with arthritis.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
#### DISCUSSION OF SAFETY AND EFFECTIVENESS 3.
# A. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA
| Model No. /Names: | MedX LPT 200 Tethered Laser &<br>MedX LPS 200 Portable Laser |
|--------------------|---------------------------------------------------------------------------------------------|
| Classification: | Lamp Infrared, Heating Category ILY<br>Physical Medicine Device, 21 CFR 890.5500 (Class II) |
| Predicate Devices: | MedX 1100 Console, LCS 100 Portable Laser and<br>LCT 100 Tethered Laser (K032231) |
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
## Description of Devices
The MedX LPT 200 is a single diode. 200 mW tethered laser powered and controlled by the MedX 1100 Console. The MedX LPS 200 Portable Laser is a single diode, 200 mW laser that has embedded control and contains a rechargeable battery. The two devices have substantially equivalent output parameters. Both of these products use 808 nm near infrared laser diodes. Additionally these devices have been designed to produce therapeutic heating for the same indications of use as described in the ILY category. Therefore the requested indications of use for these two devices; when heat is indicated for the temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, and minor pain and stiffness associated with arthritis.
Both of the MedX Laser devices in this submission utilize a single semiconductor laser diode emitting approximately 200 mW (maximum + or - 10%) at the end of the fiber optic light guide. Both laser devices use 808 nm GaAlAs (galliumaluminum-arsenide) laser diodes.
The LPS 200 Portable Laser produces only continuous mode light at 808 nm. The tethered MedX laser is powered by the MedX 1100 console and is capable of continuous duty cycle or pulsing at 8, 146 and 1000 Hz at 80% duty cycle. The console and portable automatically turns itself off after the set treatment time has been delivered.
MedX Electronics Inc.
{1}------------------------------------------------
### Technological Characteristics Summary
The technological characteristics are a combination of the predicate Device - K032231. The indications for use are identical to these predicate products
Testing for both the MedX 1100 Console and the tethered MedX LPT 200 Laser and the MedX 200 Portable Laser have been carried out in the following areas: mechanical, electrical, firmware, thermal safety, environmental conditions and electromagnetic compatibility, temperature control and irradiation distribution patterns. The MedX 1100 Console and both laser devices have been found to be safe in all areas for the intended use referenced in this submission.
### Discussion of Non-Clinical and Clinical Data
Clinical practice in Canada has demonstrated the ability of the MedX 1100 Console and tethered laser device as well as the MedX 200 Portable Laser to warm the surface temperature and provide relief of minor aches and pains associated with minor muscle and tissue injury. Research was conducted for this specific 510(k) submittal specifically demonstrating heating. See Appendix I for heat testing data.
No new clinical trials were performed on the submitted products for this 510(k) submission. With over 30 years of research and clinical experience with therapeutic heating, the modality has demonstrated to be safe as adjunctive therapy where heating is indicated.
# Conclusions Demonstrating Safety, Effectiveness and Performance
The testing carried out for the MedX LPT 200 powered and controlled by the MedX 1100 Console and the MedX 200 Portable Laser indicates that they meet design and performance functional requirements. Clinical practice and academic research demonstrate the successful use of the infrared lamp when heat is indicated for temporary increase in local blood circulation, temporary relief of minor aches and pains associated with minor muscle and minor joint pain associated with arthritis. The proposed device meets the requirements of international and US medical electrical equipment standards for safety, and key performance and safety requirements.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------MedX Electronics Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### OCT 01 2008
MedX Electronics, Inc. % Intertek Testing Services Mr. Daniel W. Lehtonen Senior Staff Engineer -- Medical Devices 2307 East Aurora Road Twinsburg, Ohio 44087
Rc: K082707
Trade/Device Name: MedX LPT 200 Tethered Laser MedX LPS 200 Portable Laser Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp. Regulatory Class: II Product Code: ILY Dated: September 15, 2008 Received: September 16, 2008
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Daniel W. Lehtonen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillancc and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 1.3 STATEMENT OF INTENDED USE
510 (k) Number: ______________________________________________________________________________________________________________________________________________________________
### Device Names: MedX LPT 200 Tethered Laser & MedX LPS 200 Portable Laser
# Indications for Use
The MedX LPT 200 single diode tethered laser is a part of the infrared console system, as per 21 CFR 890.5500. The device is used when heat is indicated for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
The MedX LPS 200 single diode Portable Laser is a rechargeable infrared device, as per 21 CFR 890.5500. The device is used when heat is indicated for temporary increase in local blood circulation, temporary relief of minor muscle and joint aches, pains and stiffness and relaxation of muscles; for muscle spasms, minor pain and stiffness associated with arthritis.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-the-counter Use<br>(21 CFR 801 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K082707 | (Division Sign-off)<br>Division of General Restorative Devices<br>510(k) Number | |
|---------------|---------|---------------------------------------------------------------------------------|--|
|---------------|---------|---------------------------------------------------------------------------------|--|
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
MedX Electronics Inc.
MedX LPT 200 (Tethered) & LPS 200 (Portable) Laser 510 (k) Submission - July 23 08
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.