The Galaxy 55 multi-parameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3 lead /5 lead), SpO2. Respiration, Temperature, Capnography (CO2) & optional Gas module unit. It can also display the digital values of HR/PR, SpO2. RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EICO2, FiCO2, NoO. O2, EtAA and FiAA readings. The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Device Story
Galaxy 55 is a multi-parameter patient monitor for bedside use by professional healthcare providers. It acquires physiological signals including ECG (3/5 lead), respiration, NIBP, IBP, temperature, SpO2, CO2, and anesthetic gases. The device processes these inputs to display real-time waveforms and digital values (HR/PR, RR, blood pressure, gas concentrations) on an external 15"/17"/19" monitor. It features tabular/graphical trends (up to 72 hours), NIBP trend storage (240 readings), and alarm recall (last 24 alarms). It includes adjustable alarm limits with audible/visible alerts. Data can be printed via a two-channel thermal array recorder. Optional connectivity includes USB, RS232, infrared remote, and Ethernet. The device assists clinicians in continuous patient assessment, enabling timely clinical intervention based on monitored vital signs.
Indicated for monitoring vital signs of single adult, pediatric, or neonatal patients in clinical settings. Parameters include ECG, SpO2, respiration, temperature, capnography, NIBP, IBP, and gas monitoring. Not for home use.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo consisting of the letters 'L' and 'T' intertwined within a circle. The 'L' is positioned on the upper left, while the 'T' is on the lower right, with both letters sharing a horizontal stroke in the middle. The logo is rendered in black against a white background, giving it a bold and clear appearance.
# SEP 2 6 2008
## LARSEN & TOUBRO LIMITED
ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS -
Mysore Works, KIADB Industrial Area, Hebbal - Hootagalii, Mysore - 570 018 = Tel : (91) - 821 - 2402561 ● Fax : (91) - 821 - 2402468
E - Mail :
Hei :
Date: 20.02.2008 Page: 01 of 04
## 510(K) SUMMARY
(Per section 807.92 ✆)
| CONTACT DATA | | | |
|-------------------------------|------------------------------------------------------------------------------|-------|---------------------------------------------------------|
| Submitter's Name | Larsen & Toubro Limited | | |
| Address | KIADB Industrial Area. Hebbal Hootagalli.<br>Mysore 570018. Karnataka, INDIA | | |
| Telephone | 91-821-2405439 | Fax | 91-821-2402468 |
| Contact Person | A.B.Deshpande | Title | Head - Quality Assurance &<br>Management Representative |
| E-Mail address | DeshpandeAB@myw.ltindia.com | | |
| Date the summary was prepared | 20th Feb, 2008 | | |
Regd. Off : L & T House, Ballard Estate, P. O. Box 276, Mumbai 400 001 • Phone : 261 8181 / 82 • 262 0223 Website : www.intelug.com
23
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters 'LT' inside of a circle. The letters are stylized and appear to be connected. The logo is black and white and appears to be a company logo. Below the logo, there is some text, but it is not clear enough to read.
K082685
p2/4
ﺇﻟﻰ ﺍﻟﺘﻲ ﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1991 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
## LARSEN & TOUBRO LIMITED
- ELECTRICAL BUSINESS GROUP + ELECTRONIC PRODUCTS -
Mysore Works, KIADB Industrial Area, Hebbal - Hootagalii, Mysore - 570 018 ▲ Tel : (91) - 821 - 2402561 ● Fax : (91) - 821 - 2402468
E · Mail :
મેદા :
Date: 20.02.2008 Page: 02 of 04
| DEVICE | |
|---------------------|---------------------------|
| Trade name | GALAXY 55 |
| Common name | Patient Monitoring System |
| Classification name | Vital Signs Monitor |
| PREDICATE DEVICE IDENTIFICATION | | |
|-----------------------------------------|---------------------------------------------------------------------------------|-----|
| CFR21 Section | Product code (optional) | MWI |
| 870.2300 | | |
| Classification panel | Cardiovascular | |
| Device Class | Class II | |
| Legally marketed Comparison Device / K# | STAR 55 Patient Monitoring System (L&T Medical Equipments & systems) / K 080173 | |
Regd. Off : L & T House, Ballard Estate, P. O. Box 278, Mumbai 400 001 ● Phone : 261 6181 / 82 ● Fax : 91 - 22 - 262 0223
Website : 20 - 201 - 1 - 2 - Website : www.intebg.
.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo of Larsen & Toubro Limited. The logo consists of a stylized "LT" inside a circle above the company name. The text "LARSEN & TOUBRO LIMITED" is written in bold, uppercase letters.
ELECTRONICS DIVISION - ELECTRONIC PRODUC Mysore Campus, KIADB Industrial Area, Hebbal · Hootagalli, Mysore · 570 018 · Tel: +91 (821) 2402468
E · Mail :
Date: 20.02.2008 Page: 03 of 04
#### વિદ્યા :
### DEVICE DESCRIPTION
GALAXY 55 is a multi-parameter patient monitoring system for continuous monitoring of the physiological parameters ECG (3/5 lead). Respiration, NIBP. TBP. Temperature. SpO2. CO2 & Gas monitoring.
GALAXY 55 is a 8-channel monitor with 15"/17"/19" external monitor display capable of displaying ECG. Respiration. SpO2, CO2. digital values of HR/PR. SpO2, RR. Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic, Diastolie and Mean), Temperature, EiCO2. FiCO2. NoO. On, EiAA and FiAA readings. It has selective 24148172 Hours tabular and graphical trends. It has special feature of NIBP having a trend of storing last 240 readings. It has Alarm Recall facility with last 24 patient alarms details. It has a two-channel thermal array recorder for printing of Tabular trends & waveforms. It has got optional communication features -USB. RS232, Infrared remote and Ethernet. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
### INTENDED USE OF THE DEVICE
The Galaxy 55 multi-parameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3 lead /5 lead), SpO2. Respiration, Temperature, Capnography (CO2) & optional Gas module unit. It can also display the digital values of HR/PR, SpO2. RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean). Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EICO2, FiCO2, NoO. O2, EtAA and FiAA readings,
The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Regd. Off:L & T House, Ballard Estate, P.O.Box 278, Mumbai 400 001 ● Phone:+91(22)6762668 ● Fax: 191(22)67625858 Website : www.LNTEBG.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Larsen & Toubro Limited. The logo consists of a stylized symbol above the company name. The symbol is a circle with the letters 'L' and 'T' intertwined inside. The company name is written in bold, uppercase letters.
- ELECTRICAL BUSINESS GROUP - ELECTRONIC PRODUCTS -
Mysore Works, KIADB Industrial Area, Hebbal - Hoolagali, Mysore - 570 018 ● Tel : (91) - 821 - 24022661 ● Fax : (91) - 821 - 2402468
E - Mail :
ମ୍ପରା :
Date: 20.02.2008 Page: 04 Of 04
## TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
Device: Larsen & Toubro limited make GALAXY 55 Patient Monitoring System. Predicate device:
- STAR 55 patient Monitoring System (Make: L&T Medical Equipments & . systems) / K 080173
The parameters available with the Larsen & Toubro Limited make GALAXY 55 Patient monitoring system are available with the predicate devices - Larsen & Toubro Limited make STAR 55 patient monitoring system
Comparison of all the parameters of GALAXY 55 to that of the predicate devices is given in the "Predicate device comparison table" document.
### Compliance to standards:
The following international standards are referred. IEC 60601-1 Medical Electrical safety IEC 60601-1-2 EMC compliance
### Conclusion:
Based on the Technological characteristics of GALAXY 55 and its comparison with that of predicate devices Star 55, Larsen & Toubro Limited believes that their device is substantially equivalent to this predicate Monitor and doesn't pose any additional risk on safety & effectiveness of the device.

(N Ravindran) Head - Design & Development
Regd. Off : L & T House, Ballard Estate, P. O. Box 276, Mumbai 400 001 ● Phone : 261 8181 / 82 • Fax : 91 - 22 - 262 0223 Website : www.intebg.com
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, depicted with stylized lines.
#### Public Health Service
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 6 2008
Larsen & Toubro Limited c/o Underwriters Laboratories, Inc. (Third Party Review) Mr. Ned Devine Senior Staff Engineer 333 Pfingsten Road Northbrook, IL 60062
Rc: K082685
Trade/Device Name: Galaxy 55 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: Scptember 2, 2008 Received: September 15, 2008
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with
{5}------------------------------------------------
Page 2 - Mr. Ned Devine
all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Dura R. Holmes
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device name: Galaxy 55
Indication for use:
The Galaxy 55 multi-parameter Patient Monitoring system is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside along with the appropriate accessories mentioned / supplied with the unit. Vital signs parameters include ECG (3 lead /5 lead), SpO2, Respiration, Temperature, Capnography (CO2) & optional Gas module unit. It can also display the digital values of HRPR, SpO2, RR, Non-Invasive Blood Pressure (Systolic, Diastolic and Mean), Invasive Blood Pressure (Systolic, Diastolic and Mean), Temperature, EtCO2, FiCO2, N2O, O2, EtAA and FiAA readings.
The user, responsible to interpret the monitored data made available, will be a professional health care provider. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The monitor is not intended for home use.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The -Counter Use (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Jenkins
(Division Sign-Off) Division of Cardiovascular Devices
K082685 510(k) Number.
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.