← Product Code [LLZ](/productcode/LLZ) · K082684

# 4D VIEWER RADIOLOGY, MODEL USWP-100A (K082684)

_Toshiba America Medical Systems, In.C · LLZ · Oct 14, 2008 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082684

## Device Facts

- **Applicant:** Toshiba America Medical Systems, In.C
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Oct 14, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

The intended use is displaying and analyzing ultrasound images for medical diagnosis in obstetric/gynecology and general examinations.

## Device Story

USWP-100A 4D Viewer-Radiology is a software package for existing workstations; receives 4D ultrasound volume data sets via DICOM from Toshiba ultrasound systems. Software provides measurement packages previously cleared on predicate device. Used by physicians in clinical settings to display and analyze ultrasound images for medical diagnosis. Output allows clinicians to perform measurements and review 4D imaging data, supporting clinical decision-making in OB/GYN and general radiology examinations.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on the device's identical intended use and measurement functionality to the predicate device.

## Technological Characteristics

Software-based image processing system. Operates on existing workstations receiving DICOM volume data. Implements measurement packages for 4D ultrasound data. Connectivity via DICOM standard. No hardware components; software-only package.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Toshiba Aplio XG; SSA-790A V2.2 ([K081065](/device/K081065.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kc. E 2624

OCT 1 4 2008

## Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) USWP-100A, 4D Viewer-Radiology

# 510(k) Summary

| Date:                                 | September 5, 2008                                            |
|---------------------------------------|--------------------------------------------------------------|
| Submitter's Name:                     | Toshiba America Medical Systems, Inc.                        |
| Submitter's Address:                  | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068 |
| Submitter's Contact:                  | Paul Biggins, Director Regulatory Affairs,<br>(714)730-5000  |
| Establishment Registration<br>Number: | 2020563                                                      |
| Device Proprietary Name:              | USWP-100A; 4D Viewer-Radiology                               |
| Common Name:                          | System, Image Processing, Radiological                       |
| Regulatory Class:                     | II [21CFR820.2050, .1560, .1550]                             |
| Performance Standard:                 | None                                                         |
| Predicate Device(s):                  | Toshiba Aplio XG; SSA-790A V2.2 [k081065]                    |
| Reason for Submission                 | New Device                                                   |

### Description of this Device:

The USWP-100A is a software package that may be applied to an existing workstation that is capable of receiving volume data sets via DICOM from Toshiba Ultrasound systems that are configured for 4D imaging. This nofitware consists of measurement packages that were cleared on the predicate device, SSA-790A, AplioXG Diagnostic Ultrasound System.

#### Summary of Intended Uses:

The intended use is displaying and analyzing ultrasound images for medical diagnosis in obstetric/gynecology and general examinations.

### Safety and Effectiveness Concerns:

This software package is designed and manufactured in compliance with the Quality System Regulation. This device does not offer new indications or functions that have not been cleared via the predicate device, nor does it offer functions that are not already available through 510(k) cleared devices that are commercially available in the U.S.

{1}------------------------------------------------

## Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) USWP-100A, 4D Viewer-Radiology

#### Substantial Equivalence:

This software package provides information to the user that is similar to that which is provided by the predicate device. The information is obtained in a manner that is similar to the predicate device. Additionally, the indications for use and intended uses are identical to those included with the predicate device. There are no new indications for use that are not already available in devices already marketed in the U.S.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo consists of a circular seal with the department's name around the perimeter. Inside the circle is a stylized symbol featuring three curved lines above a pair of wavy lines, representing the department's mission related to health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 4 2008

Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K082684

Trade/Device Name: USWP-100A, 4D Viewer-Radiology Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 30, 2008 Received: October 1, 2008

Dcar Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jaque In Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K082684

Device Name: USWP-100A, 4D Viewer-Radiology

Indications for Use:

When used by a qualified physician, this device is intended for displaying and analyzing ultrasound images for medical diagnosis in obstetric/gynecology and general examinations.

This device is not intended for diagnosis of lossy compressed images.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

Arunthirly

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K082684](https://fda-staging.innolitics.com/device/K082684)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
