The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
Device Story
ArthroCare System 12000 is a bipolar, high-frequency electrosurgical generator (Controller) used with disposable, bipolar, single-use Wands. Device operates in arthroscopic and orthopedic surgical settings; physician-operated. System delivers electrical energy to Wands to perform soft tissue ablation, resection, coagulation, and hemostasis. Output allows precise tissue management during joint procedures; benefits include controlled tissue removal and bleeding management. No complex AI/ML algorithms; standard electrosurgical power delivery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Bipolar high-frequency electrosurgical generator. System includes Controller and disposable bipolar Wands. Energy type: high-frequency electrical. Connectivity: standalone. Sterilization: not specified for generator; Wands are single-use.
Indications for Use
Indicated for soft tissue resection, ablation, coagulation, and hemostasis in arthroscopic and orthopedic procedures across all joints (ankle, elbow, hip, knee, shoulder, wrist). Specific procedures include ACL/PCL repair, acromioplasty, chondroplasty, meniscectomy, and various ligament/tendon releases.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
OCT 1 5 2008
82666
Page 173
## 510(k) Summary
# ArthroCare Corporation ArthroCare® System 12000
## General Information
| Submitter Name/Address: | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-3523 |
|------------------------------------|---------------------------------------------------------------------------|
| Establishment Registration Number: | 2951580 |
| Contact Person: | Valerie Defiesta-Ng<br>Director, Regulatory Affairs |
| Date Prepared: | September 11, 2008 |
**Device Description**
| Trade Name: | ArthroCare® System 12000 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| Generic/Common Name: | Electrosurgical Device and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation<br>Device and Accessories (Class II, 21 CFR<br>878.4400, Product Code GEI) |
**Predicate Devices**
| ArthroCare® System 12000 | K071709 |
|--------------------------|---------|
|--------------------------|---------|
### Product Description
The ArthroCare® System 12000 is a bipolar, high frequency, electrosurgical generator called the Controller that is intended to be used with a family of disposable, binolar, single use Wands.
{1}------------------------------------------------
Ko82666
Page 2 of 3
#### Intended Uses
The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft
tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
| | Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|---|----------------------------------------|--------------------------------------------------------------------------------------|
| | Ablation and Debridement | |
| • | ACL/PCL | Knee |
| • | Acromioplasty | Shoulder |
| • | Articular Cartilage | All Joints |
| • | Bursectomy | All Joints |
| • | Chondroplasty | All Joints |
| • | Fascia | All Joints |
| • | Ligament | All Joints |
| • | Notchplasty | Knee |
| • | Scar Tissue | All Joints |
| • | Soft Tissue | All Joints |
| • | Subacromial Decompression | Shoulder |
| • | Synovectomy | All Joints |
| • | Tendon | All Joints |
| | Excision and Resection | |
| • | Acetabular Labrum | Hip |
| • | Articular Labrum | All Joints |
| • | Capsule | All Joints |
| • | Capsular Release | Knee |
| • | Cartilage Flaps | Knee |
| • | Cysts | All Joints |
| • | Discoid Meniscus | Knee |
| • | Frozen Shoulder Release | Shoulder |
| • | Glenoidale Labrum | Shoulder |
| • | Lateral Release | Knee |
| • | Ligament | All Joints |
| • | Loose Bodies | All Joints |
| • | Meniscal Cystectomy | Knee |
| • | Meniscectomy | Knee |
{2}------------------------------------------------
Kof 2666
| Continued | |
|-----------|--|
| | |
| | |
| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| • Plica Removal | All Joints |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Synovial Membrane | All Joints |
| • Tendon | All Joints |
| • Triangular Fibrocartilage (TFCC) | Wrist |
| • Villusectomy | Knee |
| Coagulation | |
| • ACL/PCL | Knee |
| • Articular Cartilage | All Joints |
| • Carpal Ligaments | Wrist |
| • Glenohumeral Capsule | Shoulder |
| • Ligament | All Joints |
| • Medial Retinaculum | Knee |
| • Rotator Cuff | Shoulder |
| • Tendon | All Joints |
| • Wrist Tendons | Wrist |
#### Substantial Equivalence
This Special 510(k) proposes a modification in the performance specifications and labeling for the ArthroCare System 12000, which was previously cleared in K071709 on August 7, 2007. The indications for use, technology, principle of operation, materials, packaging, and sterilization parameters of the ArthroCare System 12000 remain the same as in the predicate cleared 510(k).
#### Summary of Safety and Effectiveness
The modified ArthroCare System 12000, as described in this Special 510(k), is substantially equivalent to the predicate device. The proposed modifications in performance specifications and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the System.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus symbol.
Public Health Service
Image /page/3/Picture/3 description: The image shows a date, "OCT 1 5 2008". The date is written in a simple, sans-serif font. The letters and numbers are all capitalized and evenly spaced. The text is clear and easy to read.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ArthroCare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523
Re: K082666
Trade/Device Name: ArthroCare System 12000 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 25, 2008 Received: September 29, 2008
Dear Ms. Defiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Ms. Valerie Defiesta-Ng
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Prosmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kw82666
Page 1 of 2
# Indications for Use Statement
ArthroCare System 12000 Device Name
K_____________________________________________________________________________________________________________________________________________________________________________ 510(k) Number:
Indications for Use:
The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| Ablation and Debridement | |
| • ACL/PCL | Knee |
| • Acromioplasty | Shoulder |
| • Articular Cartilage | All Joints |
| • Bursectomy | All Joints |
| • Chondroplasty | All Joints |
| • Fascia | All Joints |
| • Ligament | All Joints |
| • Notchplasty | Knee |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Subacromial Decompression | Shoulder |
| • Synovectomy | All Joints |
| • Tendon | All Joints |
| Excision and Resection | |
| • Acetabular Labrum | Hip |
| • Articular Labrum | All Joints |
| • Capsule | All Joints |
| • Capsular Release | Knee |
| • Cartilage Flaps | Knee |
| • Cysts | All Joints |
| • Discoid Meniscus | Knee |
| • Frozen Shoulder Release | Shoulder |
| • Glenoidale Labrum | Shoulder |
| • Lateral Release | Knee |
| • Ligament | All Joints |
| • Loose Bodies | All Joints |
| • Meniscal Cystectomy | Knee |
| • Meniscectomy | Knee |
əqmini (1)01 gift uoi
{6}------------------------------------------------
s.666
Continued
| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| • Plica Removal | All Joints |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Synovial Membrane | All Joints |
| • Tendon | All Joints |
| • Triangular Fibrocartilage (TFCC) | Wrist |
| • Villusectomy | Knee |
| Coagulation | |
| • ACL/PCL | Knee |
| • Articular Cartilage | All Joints |
| • Carpal Ligaments | Wrist |
| • Glenohumeral Capsule | Shoulder |
| • Ligament | All Joints |
| • Medial Retinaculum | Knee |
| • Rotator Cuff | Shoulder |
| • Tendon | All Joints |
| • Wrist Tendons | Wrist |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-the-Counter Use
D.A.
X
(Division Sign-Off) (Division of General, Restorative, and Neurological Devices
**510(k) Number** k082666
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.