K082654 · Respironics, Inc. · KNT · Nov 25, 2008 · Gastroenterology, Urology
Device Facts
Record ID
K082654
Device Name
ENTERAL ONLY EXTENSION SET
Applicant
Respironics, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Nov 25, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which reduce the risk of accidental connection of an I.V. system to the enteral system, or the enteral system to an I.V. system.
Device Story
Enteral Only Extension Set consists of flexible PVC tubing with specialized enteral connectors (catheter tip, oral syringe connector, or luer lock). Device connects existing nasogastric/gastric feeding tubes to delivery systems like pumps or syringes. Safety connectors prevent accidental misconnection between enteral and IV systems. Variations include stop cocks or Y-sites for auxiliary equipment attachment. Used in clinical settings by healthcare providers to facilitate enteral nutrition delivery. Benefits patient by mitigating risk of IV/enteral cross-connection errors.
Clinical Evidence
Bench testing only. Design verification tests performed based on risk analysis and product requirements; all tests met acceptance criteria.
Technological Characteristics
Materials: PVC tubing; connectors/adaptors of PVC, ABS, PE, PP, and PC. Biocompatibility evaluated per ISO 10993-1. Form factor: flexible tubing with various enteral/luer connectors, stop cocks, or Y-sites. Sterile, single-use disposable.
Indications for Use
Indicated for patients requiring nasogastric, oralgastric, or gastric tube enteral feeding. Intended for single patient use.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/3 description: The image shows the words "TAB 5" on the top left corner and the words "510(K) SUMMARY" on the bottom left corner. The text is in bold and a sans-serif font. The background of the image is a gray and white pattern.
| Date of Submission | 10 September, 2008 | NOV 25 2008 |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Official Contact / Address<br>of Manufacturing facility | Andrew P. Zeltwanger<br>Manager, Regulatory Affairs<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>Phone: 724-387-7442<br>Fax: 724-387-7490 | |
| Proprietary Name | Enteral Only Extension Sets | |
| Common/Usual Name | Tubes, Gastrointestinal (and accessories) | |
| Classification Reference | 21 CFR 876.5980 | |
| Classification | Class II | |
| Appropriate Classification Panel | Gastroenterology / Urology | |
| Product Code | KNT | |
| Predicate Devices | Vygon Nutrisafe® Extension Tube (K991918)<br>Kangaroo Enteral Feeding Extension Set (K973409) | |
## Intended Use/Indications for Use
The Enteral Only Extension Set is intended for use as an extension set for nasogastric or gastric tube enteral feeding tubes, incorporating safety connectors which eliminate the risk of accidental connection of an I.V. system to the enteral system, or the enteral system to an I.V. system..
## Patient Population/Environment of Use
The Enteral Only Extension Set is a sterile disposable for single patient use only.
{1}------------------------------------------------
K082654 2 of 3
Enteral Only Extension Sets
# Substantial Equivalence
This traditional 510(k) submittal demonstrates that the Enteral Only Extension Set is substantially equivalent to the Vygon Nutrisafe® Extension Tube (K991918), and the Kangaroo Enteral Feeding Extension Set (K973409).
Design verification tests were performed on the Enteral Only Extension Set as a result of the risk analysis and the product requirements. All tests were verified to meet the required acceptance criteria. Respironics has determined that the differences between the Enteral Only Extension Set and the predicate enteral set have no impact on the safety and effectiveness of the Enteral Only Extension Set and that all hazards were successfully mitigated.
## Device Description
#### Physical Description and Usage
The Enteral Only Extension Set consists of flexible PVC tubing designed to connect existing feeding tubes (nasogastric, gastric, etc.) to various delivery systems including pumps and syringes. The basic set consists of tubing with an enteral connector (catheter tip) and either a luer lock connector or an oral syringe connector. Other variations include the basic sets with the addition of a stop cock or Y site, to allow the clinician to attach other equipment. One configuration (Model 95017-D) utilizes luer lock connections on either end of the extension set, to allow for the connection to existing enteral feeding tubes with luer fittings. Note that in this configuration set does not utilize an oral syringe connection, to minimize the possibility of oral feeding solutions to be delivered into an IV line.
Descriptions of the different models are shown in Table 11-1. A photograph of a basic Enteral Only Extension Set is shown in Figure 11-1. Engineering drawings for all models are included at the end of Tab 11 - Device Description.
| enfor<br>able<br>1 | 1<br>। |
|--------------------|--------|
|--------------------|--------|
| Model | Description |
|---------|----------------------------------------------------------------------|
| 95017-A | Extension Set w/Catheter Tip and Luer Lock Connector, 60" |
| 95017-B | Extension Set w/Catheter Tip, Stop Cock and Luer Lock Connector, 60" |
| 95017-C | Extension Set w/Catheter Tip, Y-site and Luer Lock Connector, 60" |
| 95017-D | Extension Set w/Male and Female Luer Lock Connectors, 60" |
@2008 Respironics Inc.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K082 654 313" in a stylized font. Below this text, the words "Enteral Only Extension Sets" are printed in a smaller, simpler font. The numbers and letters are bold and have a handwritten appearance.
| 95017-E | Extension Set w/Catheter Tip and Luer Lock Connector, 36" |
|---------|------------------------------------------------------------------------------|
| 1000364 | Extension Set w/ Catheter Tip, Oral Syringe Y-Site, Luer Lock Connector, 60" |
| 1017080 | Extension Set w/ Oral Syringe Connector and Catheter Tip, 60" |
| 1018514 | Extension Set w/ Oral Syringe Connector and Catheter Tip, 36" |
| 1018516 | Extension Set w/ Stop Cock, Oral Syringe Connector and Catheter Tip, 60" |
| 1018517 | Extension Set w/ Y-Site, Oral Syringe Connector and Catheter Tip, 60" |
Figure 11-1: Enteral Only Extension Set
Image /page/2/Picture/4 description: In the image, there are several wires that are bundled together. There is a tag attached to the wires that has some text on it. The wires are dark in color, and the tag is light in color. The background is dark and blurry.
#### Material Description
The Enteral Only Extension Sets are manufactured from polyvinyl chloride (PVC) tubing. The connectors and adaptors are manufactured of several plastics: PVC, acrylonitrile butadiene styrene (ABS), polyethylene (PE), polypropylene (PP), and polycarbonate (PC). All materials have been evaluated in accordance with ISO 10993-1 Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing. A biocompatibility assessment of the materials is included in Section 15 of this submittal.
(End of tab)
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### NOV 2 5 2008
Mr. Andrew P. Zeltwanger Manager, Regulatory Affairs Respironics, Inc. Sleep and Home Respiratory Group 1740 Golden Mile Highway MONROEVILLE PA 15146
K082654 Re:
> Trade/Device Name: Enteral Only Extension Set Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: September 10, 2008 Received: September 12, 2008
Dear Mr. Zeltwanger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Jorgu M. Whang
Joyce M. Whang, Ph.D. Acting Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:
Enteral Only Extension Set
Intended Use/Indications for Use:
The Enteral Only Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which reduce the risk of accidental connection of an I.V. system to the enteral system, or the enteral system to an I.V. system.
### Patient Population/Environment of Use:
The Enteral Only Extension Set is disposable and for single patient use only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hubert Reeves
of Reproductive. Abdominal a
Page
of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.