← Product Code [HRS](/productcode/HRS) · K082625

# SYNTHES (USA) 3.5MM LCP PERIARTICULAR PROXIMAL HUMERUS PLATES (K082625)

_Synthes (Usa) · HRS · Dec 17, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082625

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Dec 17, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Synthes 3.5mm LCP Periarticular Proximal Humerus Plates are indicated for fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.

## Device Story

Synthes 3.5mm LCP Periarticular Proximal Humerus Plates are low-profile, pre-contoured metallic bone fixation plates. Used for surgical treatment of proximal humerus fractures, fracture dislocations, osteotomies, and non-unions; specifically designed for osteopenic bone. Plates are implanted by surgeons in an operating room setting to stabilize bone segments, facilitating healing. Available in stainless steel or titanium; provided sterile or non-sterile.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Low-profile, pre-contoured bone fixation plates. Materials: stainless steel or titanium. Locking compression plate (LCP) technology. Sterile or non-sterile.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes LCP Proximal Humerus Plates
- Synthes Small Fragment Dynamic Compression Locking Plates

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized figure inside. A horizontal line runs beneath the word and logo. The registered trademark symbol is located to the right of the word.

. . . . .

.

KOR2625

: 

:

1 : 18

11 - 1 - 1 - 1 - 1 - 1 - 1 -

1

| 3.0 | 510(k) Summary              | Page 1 of 1                                                                                                                                                                                                                                                                        |
|-----|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|     | Sponsor:                    | Synthes (USA)<br>Karl J. Nittinger<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>(610) 719-6941<br>DEC 17 2008                                                                                                                                                               |
|     | Device Name:                | Synthes (USA) 3.5mm LCP Periarticular Proximal Humerus Plates                                                                                                                                                                                                                      |
|     | Classification:             | Class II, §888.3030 - Single / multiple component metallic bone<br>fixation appliance and accessories.                                                                                                                                                                             |
|     | Predicate Device:           | Synthes LCP Proximal Humerus Plates<br>Synthes Small Fragment Dynamic Compression Locking Plates                                                                                                                                                                                   |
|     | Device Description:         | The Synthes 3.5mm LCP Periarticular Proximal Humerus Plates are<br>low-profile, pre-contoured bone fixation plates intended for the<br>treatment of fractures of the humerus. The plates are offered sterile<br>and non-sterile and are available in stainless steel and titanium. |
|     | Intended Use:               | Synthes 3.5mm LCP Periarticular Proximal Humerus Plates are<br>indicated for fractures, fracture dislocations, osteotomies, and non-<br>unions of the proximal humerus, particularly in osteopenic bone.                                                                           |
|     | Substantial<br>Equivalence: | Information presented supports substantial equivalence.                                                                                                                                                                                                                            |

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular fashion around the caduceus symbol.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# DEC 1 7 2008

Synthes (USA) % Mr. Karl J. Nittinger Sr. Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K082625

Trade/Device Name: Synthes (USA) 3.5 mm LCP Periarticular Proximal Humerus Plates Regulation Number: 21 CFR 888.3030

Regulation Name: Single/multiple component metallic bone fixation appliances and accessories

Regulatory Class: II

Product Code: HRS Dated: December 4, 2008 Received: December 5, 2008

Dear Mr. Nittinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

# Page 2 - Mr. Karl J. Nittinger

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N. Mulhearn

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A horizontal line runs underneath the word "SYNTHES" and the logo.

### Indications for Use

510(k) Number (if known):

Synthes (USA) 3.5mm LCP Periarticular Proximal Humerus Plates Device Name:

Indications for Use:

Synthes 3.5mm Periarticular Proximal Humerus Plates are indicated for fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.

Prescription Use X (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark H. Milken

(Division Sign-Off) (Division Signeral, Restorative, and Neurological Devices

**510(k) Number**

2.0

---

**Source:** [https://fda-staging.innolitics.com/device/K082625](https://fda-staging.innolitics.com/device/K082625)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com