GRAFTON R II EDBM, DEMINERALIZED BONE MATRIX ALLOGRAFT, RESORBABLE CALCIUM SALT BONE VOID FILLER DEVICES
K082615 · Osteotech, Inc. · MQV · Oct 16, 2008 · Orthopedic
Device Facts
Record ID
K082615
Device Name
GRAFTON R II EDBM, DEMINERALIZED BONE MATRIX ALLOGRAFT, RESORBABLE CALCIUM SALT BONE VOID FILLER DEVICES
Applicant
Osteotech, Inc.
Product Code
MQV · Orthopedic
Decision Date
Oct 16, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GRAFTON®II eDBM is intended for use as a bone graft substitute and bone void filler in bony voids or gaps of the skeletal system (i.e., spine, pelvis and extremities) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. GRAFTON®II eDBM is resorbed/remodeled and is replaced by host bone during the healing process.
Device Story
GRAFTON®II eDBM is a human bone allograft consisting of demineralized bone matrix (DBM) and surface demineralized cortical bone chips contained within an absorbable mesh pouch. Designed for intraoperative handling, the device is provided ready-to-use in a malleable, flexible state that does not harden. It is used by surgeons to fill non-structural bony voids or gaps in the spine, pelvis, or extremities. The graft may be secured with sutures, staples, or wires. The device is osteoconductive and osteoinductive; it is resorbed and replaced by host bone during the healing process. Processing includes validated viral inactivation steps for HIV-1, hepatitis B, hepatitis C, CMV, and Polio virus. The device is intended for single patient use.
Clinical Evidence
Bench testing and in vivo animal studies were provided. Osteoinductivity is confirmed via a validated athymic rat assay using a five-point linear scale to score bone formation at 28 days. Animal study results indicate performance at least as well as, or better than, predicate devices and autograft. No human clinical trial data is presented as the primary basis for equivalence.
Technological Characteristics
Human bone allograft (DBM and cortical bone chips) in an absorbable mesh pouch. Osteoconductive and osteoinductive. Malleable/flexible form factor; does not harden. Provided ready-to-use. Viral inactivation validated for HIV-1, HBV, HCV, CMV, and Polio. No electronic components or software.
Indications for Use
Indicated for patients requiring bone graft substitutes or void fillers in the skeletal system (spine, pelvis, extremities) for non-structural bony defects caused by surgery or trauma.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
GRAFTON® I eDBM
Submission Summary (Full Text)
{0}------------------------------------------------
## September 5, 2008
- 1. Submission Applicant & Correspondent:
Name: Address: Phone No.: Contact Person:
Osteotech, Inc. 51 James Wav Eatontown, NJ 07724 (732) 542-2800 Chris Talbot
- 2. Name of Product:
| Trade/Proprietary/Model Name: | GRAFTON®II eDBM |
|-------------------------------|-------------------------------------|
| Common or Usual Name: | Demineralized Bone Matrix Allograft |
| Classification Name: | Resorbable Bone Void Filler |
- 3. Devices to Which New Product is Substantially Equivalent:
GRAFTON® I eDBM is substantially equivalent, for the purpose of this 510(k), to other devices that have received 510(k) clearance for similar indications for use.
- 4. Device Description:
GRAFTON®II eDBM is a human bone allograft product containing human demineralized bone matrix (DBM) and surface demineralized cortical bone chips sealed in an absorbable mesh pouch for intraoperative handling. It is intended for use in filling bony voids or gaps of the skeletal system not intrinsic to the stability of the bony structure. GRAFTON®II eDBM is provided ready-to-use in various package sizes by volume or dimension and is intended for single patient use.
GRAFTON®II eDBM is a demineralized bone product that is osteoconductive as well as osteoinductive in an athymic rat assay. It is prepared via a proprietary processing method of Osteotech, Inc. that has been shown to consistently produce DBM that is osteoinductive in an athymic rat assay. Product and process consistency are confirmed via ongoing testing of GRAFTON®II eDBM finished product for osteoinductivity in this validated athymic rat assay utilizing a five-point linear scale (0,1,2,3,4) to score bone formation at 28 days*. This bone forming activity exhibited by GRAFTON®II eDBM in the athymic rat surrogate assay should not be interpreted as a predictor of human clinical performance.
*Edwards, J.T., PhD, Diegmann, M.H., MS, Scarborough, N.L., PhD.: Osteoinduction of Human Demineralized Bone: Characterization in a Rat Morrogh, N.E., HD. Octobilduction in Hunlan, Malan, National, Nature, 1998, Volume 357.
Page XV - 1
{1}------------------------------------------------
## 5. Intended Use/Indications
GRAFTON®II eDBM is intended for use as a bone graft substitute and bone void filler in bony voids or gaps of the skeletal system (i.e., spine, pelvis and extremities) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. GRAFTON®II eDBM is resorbed/remodeled and is replaced by host bone during the healing process.
- 6. Technical Comparison
GRAFTON®II eDBM is substantially equivalent to one or more of the predicate devices with respect to materials in that it contains human demineralized bone matrix (DBM) and surface demineralized cortical bone chips. It is provided ready-to-use in various forms that can be manipulated by the user to contour the surgical site and may be secured to the graft site with sutures, staples or wires if the surgeon desires. GRAFTON®II eDBM is implanted in this malleable/flexible state and does not harden prior to or after implantation.
- 7. Performance Data
The results of in vivo studies in animals showed that GRAFTON®II eDBM performs at least as well as, if not better than, predicate devices and/or autograft. Additional relevant animal and clinical data exist that support the successful performance of demineralized bone matrix in GRAFTON®II eDBM.
- 8. Viral Inactivation
GRAFTON®II eDBM fibers and chips are produced by proprietary processing steps that have been validated or shown to inactivate viruses including: HIV-1: hepatitis B virus (duck hepatitis virus as model); hepatitis C virus (bovine diarrhea virus as model), CMV; and Polio virus. These processes are used to further reduce the risk of disease transmission via the use of this product beyond the protection provided by donor testing and screening procedures.
Page XV - 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", indicating the department's name and national affiliation.
OCT 1 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Osteotech, Inc. % Mr. Chris Talbot Director, Regulatory Affairs 51 James Way Eatontown, New Jersey 07724
Re: K082615
Trade/Device Name: GRAFTON® II eDBM Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: September 05, 2008 Received: September 09, 2008
Dear Mr. Talbot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Mr. Chris Talbot
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
VI. Indications for Use - Statement
5082615 510(k) Number (if known):
Device Name: GRAFTON® II e DBM
Indications for Use:
GRAFTON® II eDBM is intended for use as a bone graft substitute, and bone void filler in bony voids or gaps of the skeletal system (i.e., spine, pelvis and extremities) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. GRAFTON®II eDBM is resorbed/ remodeled and is replaced by host bone during the healing process.
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chillinave. ang h
**510(k) Number** k 982618
Page VI - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.