K082614 · Laborie Medical Technologies, Inc. · GWF · Mar 6, 2009 · Neurology
Device Facts
Record ID
K082614
Device Name
EVOX ELECTRO DIAGNOSTIC DEVICE
Applicant
Laborie Medical Technologies, Inc.
Product Code
GWF · Neurology
Decision Date
Mar 6, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1870
Device Class
Class 2
Indications for Use
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies, electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
Device Story
EVOX Electro Diagnostic Device performs neurological/physiological assessments for urinary and fecal incontinence. System inputs include electrical signals from patients via bar, ring, needle, cup, or St. Mark's electrodes. Hardware utilizes Laborie 94-based platform or compatible microprocessor to perform stimulation and signal acquisition. Device delivers monophasic, current-controlled stimulation (max 100mA, 270V). Software processes captured signals to calculate latency, peak, duration, conduction velocity, and area. Output provided to clinicians for diagnostic decision-making. Used in clinical settings by healthcare professionals. Benefits include objective quantification of nerve and muscle function to aid in incontinence diagnosis.
Clinical Evidence
Bench testing only. Safety and efficacy confirmed via electrical safety testing and electromagnetic compatibility testing to recognized standards. No clinical trial data provided.
Technological Characteristics
Monophasic, current-controlled stimulator (max 100mA, 270V). 2 electrode inputs (bar, ring, needle, cup, St. Mark). FIR low-cut filters (10-4000 Hz). Common mode rejection 100dB. Input impedance >1,000 Mohms. Complies with IEC 60601-1 and IEC 60601-2-10. Software-based analysis and report generation.
Indications for Use
Indicated for patients requiring diagnostic evaluation of urinary and fecal incontinence, and other neurological and physiological assessments involving sacral reflexes, PNTML, EMG, and cerebral pudendal SEP.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{0}------------------------------------------------
KOR2614
### 510(k) Summary (per 21 CFR 807.92)
MAR = \$ 2009
### I. Applicant
Laborie Medical Technologies, Inc. 6415 Northwest Drive, Unit 10 Mississauga Ontario Canada L4V 1X1
Contact Person: Barbara Mornet. Regulatory Affairs Deputy Tel: (802) 857-1300 Fax: (802) 878-1122 Email: bmornet@laborie.com
July 30, 2008 Date Prepared:
#### II. Device Name
Laborie EVOX Electro Diagnostic Device Proprietary Name: Common/ Usual Name: Classification Name: Stimulator, Electrical, Evoked Response Regulation Number: 882.1870 and 882.1550 GWF, JXE Product Codes: Classification: 2 Classification Panel: Neurology
### III. Predicate Device
The Laborie Evo Electro Diagnostic Device is substantially equivalent to the Dantec Kevpoint K944547 and the Cadwell K962455 in its intended use and in its technical characteristics as well as the safety and effectiveness of the device as a diagnostic tool.
#### IV. Intended Use of the Device
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies. electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
#### V. Description of the Device
The EVOX Electro-Diagnostic Device utilizes Laborie Urodynamic equipment 510(k) exempt under section § 876.1620 or with another microprocessor that is compatible with the required capability.
The Laborie EVOX Electro Diagnostic Device will help diagnose and perform evaluation tests for urinary and fecal incontinence as well as other neurological and physiological assessment, including:
Pudendal Nerve Terminal Motor Latency with St Marks Electrodes Anal Sphincter CN EMG
Cerebral Pudendal Somatosensory Evoked Potential
{1}------------------------------------------------
### 510(k) Summary
(per 21 CFR 807.92)
Sacral Reflexes
Calculations will include: latency, peak, duration, conduction velocity and area.
The equipment includes the Evox Electro-diagnostic Software and the hardware required to perform the stated tests which include:
- Laborie 94-based Hardware with Integrated Electrodiagnostic Stimulation .
- . Ring/Bar Stimulation Electrodes
- Needle/Cup/Patch EMG/ECG Electrodes .
- St. Mark's Electrode and Cable .
- EMG/ECG Cabling .
{2}------------------------------------------------
### 510(k) Summary
(per 21 CFR 807.92)
,
## Summary of the Technical Characteristics
| Components<br>and or Features | Laborie EVO | Predicate Device- Dantec<br>KeypointK944547 | Cadwell Sierra/Cadwell<br>6200A K962455 |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Safety | Complies with<br>IEC 60601-1<br>standards and<br>IEC 60601-2-10<br>standards. | | Designed to comply with<br>requirements of UL 544;<br>Isolated patient connection<br>IEC 601-1: Type BF |
| Electrode<br>inputs | 2 Electrode<br>Inputs: Bar,<br>Ring, Needle,<br>Cup or St. Mark<br>Electrodes. | 5-pin Din connectors | 4 Buffered electrode<br>inputs with separate active<br>and reference 1.5 mm<br>touch proof connectors or<br>5 pin DIN connector |
| Isolated<br>Ground<br>Connections | There is main<br>power supply<br>double/reinforced<br>isolation between<br>the UDS system<br>and<br>Live/Neutral/<br>Earth, which act<br>as another<br>isolation barrier. | | 2 Connections |
| Isolation Mode<br>Rejection | No isolation<br>amplifier is used.<br>For IMRR<br>instrumentation<br>amplifier is used. | | > 150dB |
| Common Mode<br>Rejection | 100dB | > 100 dB | 90dB |
| Sensitivities | 2,100<br>MicroV/div | .5μV to +/- 20,00<br>µV/div | 2,5,10,20,50,100,200,500<br>Micro V/div; 1,2,5,10,20<br>m V/div. |
| Noise | 6 microV peak to<br>peak | | 2 micro V peak to peak<br>(10Hz to 10kHz) |
| Input<br>Impedance | > 1,000 Mohms<br>(common mode) | | > 1,000 Mohms (common mode) |
| Notch Filter | 50-60Hz | 50-60 Hz | 50-60Hz |
| Low-cut Filters | FIR filter<br>Selectable at 10,<br>50, 100, 300, 500,<br>1000, 2000, 3000,<br>4000 Hz | Filter selectable at<br>20, 50,<br>100,200,300,500,1000,<br>2000,3000,5000,<br>10000 Hz | 1-2 pole filter<br>Selectable at 0.04, 0.1, 1,<br>3, 10, 30, 100, 500 Hz |
| High-cut filters | N/A | Filter selectable at 0,<br>0.1,0.2,0.5,1,2,5,10,<br>20,50,100,200,500,<br>1000,2000,3000 Hz | 2-pole filter Selectable at<br>30, 50,100,200,300,500 Hz;<br>1, 1.5,2,3,5,10,15kHz |
| Common<br>recording<br>reference input | 1 input | 1 input | 1 input |
| Temperature<br>probe input | N/A | 20 to 45°C | |
| Stimulation<br>Signal | Monophasic<br>Current controlled<br>for a maximum of<br>up to 100mA and<br>270 V | Biphasic and monophasic | |
| Signal Capture-<br>EMG | +/- 1mV range | +/-5µV to +/- 20,000 µV range available +/-1<br>to 50mV used for test | |
| Signal Capture-<br>EEG | +/- 20µV range | +/-5µV to +/- 20,000 µV range available +/-5<br>µV to 50 µV used for test | |
| Multiple<br>Sample<br>Acquisition for<br>Accuracy | Yes | Yes | Yes |
| Event Marking | Yes | Yes | Yes |
| Software Data<br>Analysis | Yes | Yes | Yes |
| Report<br>Generation &<br>Data Storage | Yes | Yes | Yes |
| Patient Info | Yes | Yes | Yes |
{3}------------------------------------------------
# 510(k) Summary
(per 21 CFR 807.92) ,
{4}------------------------------------------------
### 510(k) Summary (per 21 CFR 807.92)
### VI. Testing
Bench studies have confirmed the efficacy of the EVOX Electro Diagnostic Device. Safety testing included electrical safety testing and electromagnetic compatibility testing to recognized standards.
### VII. Safety & Effectiveness
The Laborie EVOX Electro-Diagnostic Device is substantially equivalent to the Dantec Keypoint and the Cadwell Sierra.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the bird figure.
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Laborie Medical Technologies, Inc. % Ms. Barbara Mornet Regulatory Affairs Deputy 400 Avenue D, Suite 10 Williston, VT 05495-7828
MAR - 6 2009
Re: K082614
Trade/Device Name: Laborie EVOX Electro Diagnostic Device Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: II Product Code: GWF, JXE Dated: January 23, 2009 Received: January 26, 2009
Dear Ms. Mornet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va may, annols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device to such additional controls. Existing major regulations affecting your device can may or cally of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r tease oc acrised that I be roundevice complies with other requirements of the Act that I Drives Intacted and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Ms. Barbara Mornet
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark McMillan
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K082614
Device Name: Laborie EVOX Electro Diagnostic Device
Indications For Use:
The EVOX Electro Diagnostic Device is intended for the testing of sacral reflexes, pudendal nerve terminal motor latency (PNTML) studies, electromyography (EMG), and cerebral pudendal somatosensory evoked potential (SEP).
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K08 2614 |
|---------------|----------|
|---------------|----------|
Page 1 of ____
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.