← Product Code [NKB](/productcode/NKB) · K082592

# LDR SPINE EASYSPINE SYSTEM (K082592)

_Ldr Spine USA · NKB · Oct 8, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082592

## Device Facts

- **Applicant:** Ldr Spine USA
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Oct 8, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The LDR Easyspine System is a posterior, noncervical pedicle and non-pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: - Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis - Trauma (i.e. fracture or dislocation) - Spinal stenosis - Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis) - Tumor - Pseudoarthrosis - Failed previous fusion

## Device Story

Easyspine System is a posterior, non-cervical spinal fixation system; components include sacral and pedicle screws, transverse connectors, hooks, and rods of varying rigidities. Used for mono- and multi-segmental stabilization of thoracic, lumbar, and sacral vertebrae to promote fusion. Implantation performed by surgeons in an operating room setting using specialized instrumentation for distraction, compression, or reduction of the spine. System provides mechanical immobilization to treat instabilities or deformities; benefits include structural support during the fusion process.

## Clinical Evidence

Bench testing only. Mechanical test results demonstrate that the modified Easyspine Transverse Connector design is substantially equivalent to the predicate design.

## Technological Characteristics

Posterior, non-cervical pedicle and non-pedicle spinal fixation system. Components: sacral/pedicle screws, transverse connectors, hooks, rods. Materials: metallic spinal fixation components. Principle: mechanical immobilization and stabilization. Single-use. Manual instrumentation used for implantation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- LDR Spine Easyspine System ([K043094](/device/K043094.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for LDR, with the letters in bold black font. To the left of the letters is a circular design made up of small dots. Below the letters is the phrase "a passion for innovation" in a smaller, lighter font.

## 510(k) Summary

LDR Spine USA Easyspine System

### 1. Owner's Name & Address

LDR Spine USA 4030 West Braker Lane, Suite 360 Austin, TX 78759

(512) 344-3333 Phone: (512) 344-3350 Fax:

### 2. Contact Person

Noah Bartsch MS, RAC, Manager, Clinical, Regulatory & Quality Affairs LDR Spine USA 4030 West Braker Lane, Suite 360 Austin, TX 78759

| Phone: | (512) 344-3319           |
|--------|--------------------------|
| Fax:   | (512) 344-3350           |
| Email: | noahbartsch@ldrspine.com |

## 3. Date 510(k) Summary Prepared

September 5, 2008

- 4. Trade Name LDR Spine Easyspine® System Common Name: Posterior Thoracolumbar Spinal System Classification: KWP: Spinal Interlaminal Fixation Orthosis - Class II per 888.3050 MNI and MNH: Pedicle Screw Spinal System - Class !! per 888.3070 NKB: Pedicle Screw Spinal System - Class III per 888.3070

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for LDR. To the left of the letters "LDR" is a circular design that is made up of many small dots. The letters "LDR" are in a bold, sans-serif font.

a passion for innova

# 5. Leqally Marketed Equivalent Predicate Devices:

LDR Spine Easyspine System (K043094)

## 6. Device Description

Easyspine® implants are single use devices for mono-and multi-segmental stabilization of the lumbar and thoracic vertebrae to promote fusion. Easyspine® consists of sacral and pedicle screws, transverse connectors, hooks, and rods of different rigidities. Specialized associated instrumentation is designed for implantation of these devices and for the distraction, compression or reduction of the lumbar and thoracic spine.

## 7. Intended Use of the Device

The LDR Easyspine System is a posterior, noncervical pedicle and non-pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

- . Deqenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis .
- Trauma (i.e. fracture or dislocation) .
- Spinal stenosis .
- . Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- . Tumor
- Pseudoarthrosis t
- Failed previous fusion .

## 8. Non-Clinical Performance Data

Mechanical test results demonstrate that the modified Easyspine Transverse Connector design is substantially equivalent to the predicate Easyspine Transverse Connector design, and that the fundamental scientific technology of the device and of the Easyspine System is not changed.

pap 2 of 2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 0 8 2008

LDR Spine USA % Mr. Noah Bartsch Manager - Clinical, Regulatory & Quality Affairs 4030 West Braker Lane, Suite 360 Austin, TX 78759

Re: K082592 Trade/Device Name: Easyspine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, KWP, MNH, MNI Dated: September 5, 2008 Received: September 8, 2008

Dear Mr. Bartsch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Noah Bartsch

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark N. Mulkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): 5087597

Device Name: LDR Spine Easyspine® System

Indications for Use:

The LDR Easyspine System is a posterior, noncervical pedicle and non-pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

- Degenerative disc disease (defined as back pain of discogenic origin . with degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis .
- Trauma (i.e. fracture or dislocation) �
- . Spinal stenosis
- Deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis) .
- . Tumor
- Pseudoarthrosis .
- Failed previous fusion .

Prescription Use ___ X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) A. Millous

(Division Sign-Off) bir islan of General, Restorative, and Neurological Devices

510(k) **Number** K082592

---

**Source:** [https://fda-staging.innolitics.com/device/K082592](https://fda-staging.innolitics.com/device/K082592)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).