← Product Code [OTP](/productcode/OTP) · K082571

# AVAULTA SUPPORT SYSTEM (K082571)

_C.R. Bard, Inc. · OTP · Sep 30, 2008 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082571

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [OTP](/productcode/OTP.md)
- **Decision Date:** Sep 30, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.5980
- **Device Class:** Class 3
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

## Device Story

The Avaulta Support System is a sterile, single-use, permanent surgical implant designed for the correction of anterior or posterior vaginal wall prolapse. It consists of a soft knit section for compliant organ support and strong knit arms for tension-free fixation. The Avaulta Plus variant incorporates a porous, acellular, ultrathin crosslinked collagen sheet attached to a nonabsorbable monofilament polypropylene mesh; this collagen layer acts as a protective barrier between mucosal tissue and the mesh, featuring apertures to facilitate host tissue ingrowth and capillary vessel formation. The device is used by surgeons in a clinical/surgical setting to provide long-term reinforcement of pelvic floor structures.

## Clinical Evidence

Bench testing only. Testing was performed in accordance with the FDA Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999).

## Technological Characteristics

Implantable surgical mesh composed of nonabsorbable monofilament polypropylene. Avaulta Plus variant includes a porous, acellular, crosslinked collagen sheet. Features soft knit central section and strong knit arms. Sterile, single-use device.

## Regulatory Identification

Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).

## Predicate Devices

- Avaulta™ Solo Support System ([K063712](/device/K063712.md))
- Avaulta™ Plus Biosynthetic Support System ([K063712](/device/K063712.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Bard Urological Division C. R. Bard, Inc. 13183 Harland Dr. Covington, GA 30014

SEP 3 0 2008

Image /page/0/Picture/2 description: The image shows the text "K082571 1/2 BARD". The text is handwritten and appears to be a label or identifier. The word "BARD" is written in block letters.

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATIC

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

- Submitter Information: A.
Submitter's Name:

Address:

C. R. Bard, Inc. Bard Urological Division 13183 Harland Drive Covington, GA 30014

| Contact Person:                    | John C. Knorpp |
|------------------------------------|----------------|
| Contact Person's Telephone Number: | 678-342-4920   |
| Contact Person's Fax:              | 770-788-5513   |

- в. DEVICE NAME:

| Trade Name(s):        | Avaulta™ Solo Support System<br>Avaulta™ Plus Biosynthetic Support System |
|-----------------------|---------------------------------------------------------------------------|
| Common/Usual Name:    | Surgical Mesh                                                             |
| Classification Names: | 79 - Mesh, Surgical, Polymeric (OTP, PAI)                                 |
| CFR Reference:        | 21 CFR 878.3300                                                           |
| Classification Panel: | General and Plastic Surgery                                               |

- C. PREDICATE DEVICE NAME:
Trade Names: Avaulta™ Solo Support System Avaulta™ Plus Biosynthetic Support System K063712

- D. Device Description:
The Avaulta™ Support System includes a sterile, single use, permanent implant that provides long term reinforcement to support structures in the correction of anterior or posterior vaginal wall prolapse. The central soft knit section provides compliant organ support while the strong knit arms provide improved strength for tension free fixation of the implant.

The Avaulta™ Plus Biosynthetic Support System and Avaulta™ Solo support system both utilize a nonabsorbable monofilament, polypropylene mesh to provide long-term reinforcement for support structures. The Avaulta™ Plus Biosynthetic Support System adds a porous, acellular, ultrathin sheet of crosslinked collagen attached to the polypropylene mesh which serves to establish a protective barrier between mucosal tissue and

{1}------------------------------------------------

K08JΓ71
2/2

the polypropylene mesh and contains apertures uniformly sized to allow for ingrowth of host tissue and capillary vessels.

- E. INTENDED USE:
The Avaulta™ Support System is indicated for tissue reinforcement and longlasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

- TECHNOLOGICAL CHARACTERISTICS SUMMARY: F.
The subject Avaulta™ Support System has the same intended use, general design and fundamental scientific technology as the predicate device.

- PERFORMANCE DATA SUMMARY: ુ.
The appropriate testing to determine substantial equivalence was completed. This includes testing in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh (March 22, 1999).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 2 8 2012

C.R. Bard, Incorporated % John C. Knorpp, RAC Director, Regulatory Affairs Bard Urological Division 13183 Harland Drive COVINGTON GA 30014

Re: K082571 Trade/Device Name: Avaulta Solo™ Support System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP, PAI Dated: September 3, 2008 Received: September 5, 2008

Dear Mr. Knorpp:

This letter corrects our substantially equivalent letter of September 30, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Weeks

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

R. Bard, Inc., Bard Urological Division Avaulta™ Support System Premarket Notification [510(k)]

1.4 Indications for Use Statement

| 510(k) Number (if known): | K082571                 |
|---------------------------|-------------------------|
| Device Name:              | Avaulta™ Support System |

Indications for Use:

The Avaulta™ Support System is indicated for tissue reinforcement and long-lasting stabilization of fascial structures of the pelvic floor in vaginal wall prolapse where surgical treatment is intended either as mechanical support or bridging material for the fascial defect.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Milfre dyl. Forman

rivision of Anesthesiology, General Hospital ifection Control, Dental Devices

510(k) Number

(Recommended Format 11/13/2003)

---

**Source:** [https://fda-staging.innolitics.com/device/K082571](https://fda-staging.innolitics.com/device/K082571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).