K082567 · Howmedica Osteonics Corp. · HRY · Oct 3, 2008 · Orthopedic
Device Facts
Record ID
K082567
Device Name
STRYKER COMPARTMENTAL KNEE SYSTEM LINE EXTENSION
Applicant
Howmedica Osteonics Corp.
Product Code
HRY · Orthopedic
Decision Date
Oct 3, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker® Compartmental Knee System Line Extension consists of sterile, single-usc devices intended for replacement of the patellofemoral joint and/or the condyle region(s) of the femoral joint as needed. The Stryker® Compartmental Knee System Line Extension includes Triathlon® PKR femoral condyle regions for cither the right or left knee.
Device Story
System of sterile, single-use orthopedic implants for patellofemoral and/or unicondylar knee arthroplasty; includes Triathlon® PKR femoral condyle components. Used by orthopedic surgeons in clinical settings to replace damaged joint surfaces. Implants are fixed using bone cement. Device allows surgeons to combine components from existing Howmedica Osteonics knee systems to address specific patient anatomy and pathology; provides mechanical restoration of joint function; reduces pain; improves mobility in patients with degenerative or post-traumatic joint disease.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and intended use compatibility with existing marketed systems.
Technological Characteristics
Metal/polymer semi-constrained cemented knee prosthesis. Components sourced from existing Howmedica Osteonics knee systems. Single-use; requires bone cement for fixation.
Indications for Use
Indicated for cemented patellofemoral and/or unicondylar arthroplasty in patients with conditions unaddressable by a single device. Applicable for degenerative arthritis of distal femur/patella; history of patellar dislocation or fracture; failed previous surgeries (arthroscopy, tibial tubercle elevation, lateral release); moderately disabling osteo- or post-traumatic arthritis; revision of previous unsuccessful procedures; alternative to tibial osteotomy for unicompartmental osteoarthritis; or poor bone stock/deficiencies of femoral condyle/tibial plateau.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Submission Summary (Full Text)
{0}------------------------------------------------
K082567
# OCT 3 - 2008
### Summary of Safety and Effectiveness Stryker® Compartmental Knee System Line Extension
| Proprietary Name: | Stryker® Compartmental Knee System Line Extension |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Knee Prosthesis |
| Classification Name and Reference: | Knee joint femortotibial polymer/metal semi constrained cemented prosthesis. 21 CFR §888.3530 |
| | Knee joint patellofemoral polymer/metal semi constrained cemented prosthesis. 21 CFR §888.3540 |
| | Knee joint femorotibial metal/polymer non-constrained cemented prosthesis. 21 CFR §888.3520 |
| | Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis. 21 CFR §888.3530 |
| Device Product Code: | 87 NPJ, 87 KRR, 87 HSX, 87 HRY |
| For Information Contact: | Francisco Haro, Sr. Regulatory Affairs Specialist Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 Phone: (201) 831-5493 Fax: (201) 831-6038 |
| Date Summary Prepared: | September 3, 2008 |
#### Description:
The Stryker® Compartmental Knee System consists of sterile, single-use components intended for replacement of the femoral side of the patellofemoral joint and/or the condyle region(s) of the femoral joint as needed. The system includes patellofemoral, femoral, and tibial components from currently marketed Howmedica Ostconics' knee systems for patellofemoral and unicondylar arthroplasty. The system allows the physician to choose the most appropriate option to treat the patient with patellofemoral arthroplasty and/or unicondylar arthroplasty as needed.
{1}------------------------------------------------
#### Intended Use:
The Stryker® Compartmental Knee System Line Extension consists of sterile, single-usc devices intended for replacement of the patellofemoral joint and/or the condyle region(s) of the femoral joint as needed. The Stryker® Compartmental Knee System Line Extension includes Triathlon® PKR femoral condyle regions for cither the right or left knee.
#### Indications for Use:
The Stryker® Compartmental Knee System is intended to be used in cemented patellofemoral and/or unicondylar arthroplasty in patients where conditions exist that cannot be addressed by a single device to treat the femorotibial or patellofemoral regions of the knee. The indications for the different components of the Stryker® Compartmental Knee System include conditions when the patellofemoral and/or condylar region(s) have been affected by one or more of the following conditions:
- . Degenerative arthritis in the distal femur and patella.
- Patients with a history of patellar dislocation or patella fracture, .
- . Patients with failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists,
- . Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- . Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of a unicompartmental knec prosthesis,
- As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis, . or
- Where bone stock is of poor quality or inadequate for other reconstructive techniques . as indicated by deficiencies of the femoral condvle/tibial plateau.
These components are single use only and are intended for implantation with bone cement only.
{2}------------------------------------------------
## Substantial Equivalence:
The device is substantially equivalent to its predicates for patellofemoral arthroplasty and femorotibial arthroplasty in regards to intended use, design, materials, and operational principles. The analysis demonstrated that the components from these systems are compatible when used for patellofemoral and/or femorotibial replacement.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three intertwined strands and a wing-like structure above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
Public Health Service
OCT 3 - 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Francisco Haro Senior Regulatory Affairs Specialist Howmedica Osteonics Corporation 325 Corporate Drive Mahwah, New Jersey 07430
Re: K082567
Trade/Device Name: Stryker® Compartmental Knee System Line Extension Regulation Number: 21 CFR 888.3530 Regulation Name: Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Codes: IIRY, NPJ, HSX, KRR Dated: September 3, 2008 Received: September 4, 2008
Dear Mr. Francisco Haro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal l'ood, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the gencral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mikkelson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K082567 510(k) Number (if known):
Device Name: Stryker® Compartmental Knec System Line Extension
Indications for Use;
The Stryker® Compartmental Knee System line extension is intended for use in cemented patellofemoral and/or unicondylar arthroplasty in patients where conditions exist that cannot be addressed by a single device to treat the femorotibial or patellofemoral regions of the knee. The indications for the different components of the Stryker® Compartmental Knec System line extension include conditions when the patellofemoral and/or condylar region(s) have been affected by one or more of the following conditions:
- . Degenerative arthritis in the distal femur and patella,
- . Patients with a history of patellar dislocation or patella fracture,
- . Patients with failed previous surgery (arthroscopy, tibial tubercle elevation, lateral release) where pain, deformity or dysfunction persists,
- Moderately disabling joint disease of the knee resulting from painful osteo- or post . traumatic arthritis,
- . Revision of previous unsuccessful surgical procedures, either involving, or not involving, previous use of a unicompartmental knee prosthesis,
- As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis, . or
- . Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the femoral condyle/tibial plateau.
These components are single use only and are intended for implantation with bone cement.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Mark Mill
Concurrence of CDRH, Office of Device Evaluation (ODE)
**(Division Sign-Off)**
Division of General, Restorative,
**(Division Sign-Off)**
Division of Anesthesiology, General Hospital
Division of General, Restorative, and Neurological Devices
Infection Control, Dental Devices
510(k) Number:
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.