BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PHOENIX INDUCIBLE MACROLIDE RESISTANCE TEST

K082538 · Becton, Dickinson & CO · LON · Nov 20, 2008 · Microbiology

Device Facts

Record IDK082538
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PHOENIX INDUCIBLE MACROLIDE RESISTANCE TEST
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateNov 20, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) of gram positive bacteria from pure culture belonging to the genera Staphylococcus, Enterococcus, other gram positive cocci and gram positive bacilli. The BD Phoenix™ Automated Microbiology System is also intended for the quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram positive bacterial isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

Device Story

System uses colorimetric oxidation-reduction, growth-based AST in molded polystyrene trays. Input: pure culture bacterial colonies suspended in broth, equated to 0.5 McFarland. Instrument incubates at 35°C; monitors color change from blue to pink/colorless via automated optical readings every 20 minutes. Software-driven 'EXPERT' system interprets growth patterns against CLSI-derived rules to determine MIC and resistance phenotypes. Output: ID, MIC values, and categorical interpretations (SIR). Used in clinical microbiology labs; operated by technicians. Benefits: rapid (4-16 hours) detection of inducible clindamycin resistance (iMLSb) in Staphylococcus, distinguishing it from macrolide-only efflux resistance, aiding antibiotic selection.

Clinical Evidence

Performance evaluated using clinical, stock, and challenge isolates at multiple US sites. Comparison against CLSI reference broth microdilution method. For the Phoenix Inducible Macrolide Resistance Test, Categorical Agreement (CA) was 97.6% (n=295). Reproducibility studies showed >90% intra-site and >95% inter-site reproducibility for Gram-positive isolates.

Technological Characteristics

Colorimetric oxidation-reduction, growth-based AST. Molded polystyrene trays with 136 micro-wells. Cation-adjusted Mueller-Hinton broth with 0.01% Tween 80. Automated optical reading system. Software-driven 'EXPERT' system for rule-based interpretation. Connectivity: Automated instrument integration.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative and Gram-positive bacterial isolates (Staphylococcus, Enterococcus, Streptococcus) from pure culture. Specifically includes detection of inducible macrolide-lincosamide-streptogramin B (iMLSb) resistance in Staphylococcus species.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k082538 B. Purpose for Submission: Addition of the Phoenix Inducible Macrolide resistance (iMLSb) test to the BD Phoenix™ Automated Microbiology System C. Measurand: Inducible Macrolide resistance (iMLSb) test (Clindamycin/Erythromycin) D. Type of Test: Antimicrobial Susceptibility Test (AST) (Qualitative) colorimetric oxidation-reduction, growth-based E. Applicant: Becton, Dickinson &amp; Company F. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System – Phoenix Inducible Macrolide resistance (iMLSb) in *Staphylococcus* species G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System 2. Classification: Class II 3. Product code: LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation {1} 4. Panel: Microbiology H. Intended Use: 1. Intended use(s): The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) of gram positive bacteria from pure culture belonging to the genera Staphylococcus, Enterococcus, other gram positive cocci and gram positive bacilli. The BD Phoenix™ Automated Microbiology System is also intended for the quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram positive bacterial isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. 2. Indication(s) for use: This premarket notification is for the addition of the Phoenix Inducible Macrolide resistance (iMLSb) Test in Staphylococcus species to Gram-positive ID/AST or AST only Phoenix panels. The Phoenix Inducible Macrolide resistance Test is used to detect inducible macrolide-lincosamide-streptogramin B resistance in Staphylococcus species. 3. Special conditions for use statement(s): Prescription use Perform an alternate method when a positive result with Coagulase Negative Staphylococcus (CNS) specifically, S. cohnii ssp cohnii, S. gallinarum, and S. hominis and iMLSb is identified. 4. Special instrument requirements: Not applicable I. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation 2 {2} of Mueller-Hinton broth containing 0.01% Tween 80. After adding the indicator solution to the AST inoculum the color turns to blue. After inoculation and incubation, the color changes to pink then to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The AST has a final inoculum of 5 x 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the CLSI documentation. Readings are taken every 20 minutes with an AST result available between 4-16 hours. This is only an autoread result; no manual readings are possible with this system. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Sensititre® Dtest 2. Predicate K number(s): k073653 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria. | An in vitro diagnostic product for clinical susceptibility testing of gram negative and gram positive organisms. | | Inoculum | Prepared from colonies using the direct inoculation method | Prepared from colonies using the direct inoculation method | | Result Reported | Report results as minimum inhibitory concentration (MIC) and | Report results as minimum inhibitory concentration (MIC) and | {3} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | categorical interpretation (SIR) | categorical interpretation (SIR) | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Incubation | Rapid (<16 hours) | 18-24 hours | | Reading method | Automated | manually or automatically | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S18) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." # L. Test Principle: The Phoenix Inducible Macrolide resistance (iMLSb) Test is used to detect inducible macrolide-lincosamide-streptogramin B (MLSb) resistance in Staphylococcus species. MLSb resistance, usually encoded by ermA or ermC genes, may be either constitutive (always expressed) or inducible after exposure to a macrolide antibiotic (e.g. erythromycin, clarithromycin, etc.). The Phoenix Inducible Macrolide Resistance Test is based on the same principle as the CLSI-recommended Disk Approximation test (D-Test) for the detection of inducible clindamycin resistance. When the Phoenix Inducible Macrolide Resistance Test result is positive, the categorical interpretation of clindamycin on the same Phoenix panel will be reported as resistant and accompanied by a separate BDXpert message. Staphylococcus isolates resistant to both erythromycin and clindamycin on initial testing will be reported as the MLSb phenotype to distinguish them from isolates that are resistant to macrolides alone by efflux mechanism. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: a. Precision/Reproducibility: Reproducibility testing was performed on 10 Staphylococcus isolates. Acceptable reproducibility was demonstrated with only category agreement (Negative, Positive) since that is all that is detected. b. Linearity/assay reportable range: Not applicable {4} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The CLSI recommended Quality Control (QC) isolates, S. aureus ATCC BAA-976 and S. aureus ATCC BAA-977 were tested on every test occasion with the reference method and the BD Phoenix™. The BD Phoenix™ was tested a sufficient number of times to demonstrate that the system can produce QC results in the recommended ranges. Quality Control Table | ORGANISM | Conc ug/mL (Ery/Cli) | Reference | BD Phoenix™ | | --- | --- | --- | --- | | | | | | | S. aureus BAA-976 | Neg | 63 | 68 | | Expected Range: Neg | Pos | 0 | 0 | | | | | | | S. aureus BAA-977 | Neg | 2 | 1 | | Expected Range: Pos | Pos | 60 | 68 | | | | | | | | | | | Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The CLSI recommended disk diffusion reference panel was prepared according to the CLSI recommendation. The disk diffusion reference panel was set up on a blood agar plate, inoculated with a direct colony suspension {5} and incubated in ambient air at 35°C for 16 – 18 hours as recommended by CLSI. Clinical testing was performed on 295 Staphylococcus species at three external sites which included fresh and stock clinical isolates, a set of challenge organisms, and CDC study isolates. The Phoenix Inducible Macrolide resistance (iMLSb) Test is used to detect induction of clindamycin resistance by erythromycin. The test algorithm compares the growth rate observed in the test well to a predetermined threshold to determine the expression of the inducible resistance phenotype in the test isolate. The comparison resulted in the following performance evaluations as reflected below. There are 3 maj errors with the CNS group, specifically S. cohnii ssp cohnii, S. gallinarum, and S. hominis. Summary Table | | Total | CA | CA% | # Neg | # Pos | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | | S. aureus | 214 | 210 | 98.1 | 94 | 120 | 1 | 3 | | S. epidermidis | 39 | 39 | 100 | 23 | 16 | 0 | 0 | | Coagulase-negative Staphylococcus (CNS) | 42 | 39 | 92.9 | 21 | 21 | 3 | 0 | CA - Category Agreement maj-major discrepancies vmj-very major discrepancies Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: {6} Not applicable 5. Expected values/Reference range: Staphylococcus species – Positive (inducible clindamycin resistance) Negative (no inducible clindamycin resistance) N. Proposed Labeling: The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The information submitted in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...