← Product Code [NDC](/productcode/NDC) · K082512

# INTELLIGENT DOSING SYSTEM FOR INSULIN-MY INSULIN DOSER (K082512)

_Dimensional Dosing Systems, Inc. · NDC · Nov 7, 2008 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082512

## Device Facts

- **Applicant:** Dimensional Dosing Systems, Inc.
- **Product Code:** [NDC](/productcode/NDC.md)
- **Decision Date:** Nov 7, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device, Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K082512 · Nov 7, 2008 | INTELLIGENT DOSING SYSTEM FOR INSULIN-MY INSULIN DOSER | Dimensional Dosing Systems, Inc. | Nursing home resident clinical records | A Phase IV study was conducted in a nursing home setting to evaluate the insulin titration algorithm's ability to stabilize blood glucose levels in residents converted from oral antidiabetic drugs (OADs) to insulin. | Phase IV study; Nursing home; Insulin titration; Blood glucose stabilization |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Phase IV Nursing Home Study; Phase IV clinical study | Nursing home residents maintained on two or more OADs | Not applicable for this study | Stabilization of blood glucose following conversion from OAD therapy |

## Indications for Use

My Insulin Doser/IDS allows a diabetic patient to calculate the best next dose of insulin to achieve a personal glucose target.

## Device Story

My Insulin Doser/IDS is a software-based dosing calculator for insulin. Patients input current blood glucose values and recent insulin doses; the device calculates the next insulin dose to reach a target glucose level. Used at home by patients; accessible via Palm, PC, or web platforms. The system provides warnings if values are out of range or if calculated doses deviate by >20% from the most recent dose. It enables remote patient management, allowing healthcare professionals to view patient data via internet access. The device aims to assist patients in managing insulin therapy and stabilizing blood glucose levels.

## Clinical Evidence

Phase IV study conducted in nursing home residents converted from oral anti-diabetic (OAD) therapy to an insulin replacement algorithm. The study employed the Dimensional Dosing Systems insulin titration algorithm to stabilize blood glucose post-conversion. Additionally, a labeling comprehension study was performed with 55 subjects (ages 18-80) to evaluate user understanding of the system, which was designed for a 6th-grade reading level.

## Technological Characteristics

Software-based dosing calculator; operates on Palm, PC, and web platforms. Utilizes a proprietary insulin titration algorithm. Features include user-entered glucose/insulin data, automated warnings for out-of-range values, and remote data access for healthcare providers. Software is classified as Major Level of Concern.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- TRxF Intelligent Dosing System ([K011571](/device/K011571.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Korsir

# 510(k) Executive Summary August 20, 2008

NOV - 7 2008

Company Name:

Dimensional Dosing Systems

Company Address

Telephone:

2465 Dogwood Drive Wexford, PA 15090 (724) 933-7874

Contact Person: John McMichael

Summary Preparation Date: September 4, 2007

Trade Name: My Insulin Doser/IDS

Common Name: Dosing Calculator

Classification Name: Calculator, Drug Dose

Predicate Device Identification: CFR21 868.1890 Dosing Calculator Product Code: NDC Device Class: II

| LEGALLY MARKETED EQUIVALENT DEVICE |                                |          |
|------------------------------------|--------------------------------|----------|
| Company                            | Product                        | 510(k) # |
| RxFiles                            | TRxF Intelligent Dosing System | K011571  |

### DESCRIPTION OF DEVICE

My Insulin Doser/IDS is a next dose calculator for any insulin that can be used by patients with a prescription from a licensed prescriber.

# INTENDED USE OF THE DEVICE

My Insulin Doser/IDS allows a diabetic patient to calculate the best next dose of insulin to achieve a personal glucose target.

{1}------------------------------------------------

# TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE AS COMPARED TO PREDICATE DEVICE [

## Predicate Product Comparison Table:

| Manufacturer          | Dimensional Dosing<br>Systems                                                                                                                                                                               | Dimensional Dosing Systems                                                                                                                                                                           |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name            | "My Insulin Doser/IDS"                                                                                                                                                                                      | TRXF Intelligent Dosing System<br>(IDS)                                                                                                                                                              |
| K Number              |                                                                                                                                                                                                             | K011571                                                                                                                                                                                              |
| Intended Use          | My Insulin Doser/IDS allows<br>a diabetic patient to calculate<br>the best next dose of insulin<br>to achieve a personal glucose<br>target.                                                                 | Software based drug dosing<br>calculator designed for use by<br>the physician to calculate next<br>dose of any drug to achieve a<br>desired target                                                   |
| Device<br>Description | My Insulin Doser/IDS a next<br>dose calculator for any insulin<br>that can be user by patients<br>with a prescription from a<br>licensed prescriber.                                                        | The IDS™ is a "next" dose<br>calculator for any drug that can<br>be used by physicians<br>to<br>calculate the next dose<br>for<br>patients.                                                          |
| Target<br>population  | Diabetic Patient                                                                                                                                                                                            | Healthcare professional                                                                                                                                                                              |
| Where used            | Home                                                                                                                                                                                                        | Healthcare facility                                                                                                                                                                                  |
| Human<br>Factors      | Patient enters their own<br>glucose values and amounts<br>of insulin taken.<br>Warnings are presented<br>when values are out of range<br>and/or insulin doses are < or<br>> 20% of the most recent<br>dose. | Physicians enter patient's<br>glucose values and amounts of<br>insulin. Warnings are<br>presented when values are out<br>of range and/or insulin doses<br>are < or > 20% of the most<br>recent dose. |
| Performance           | Performance testing was<br>performed and all tests<br>showed satisfactory results.                                                                                                                          | Performance testing was<br>performed and all tests showed<br>satisfactory results.                                                                                                                   |
| Software<br>based     | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                                                                  |
| Dose<br>calculation   | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                                                                  |

{2}------------------------------------------------

### Discussion of Similarities and Differences with the Predicate Product:

The Intelligent Dosing System (IDS) for Physicians was chosen as a substantially equivalent device. My Insulin Doser/IDS has the following similarities to those of the predicate device:

- Uses the same algorithm .
- Uses the same operating principles .
- Uses a similar graphical user interface. �
- Available via the Palm, PC or Website. .
- User enters same variables. .

The only modifications made are:

- . Single Patient only
- . Insulin only

#### Conclusion:

My Insulin Doser/IDS is identical to the predicate device in functionality except for a few features that allow the device to be a single patient dosing calculator devised for insulin dosing only. The predicate device, the Intelligent Dosing System IDS) was developed as a dosing calculator for the healthcare professional and was approved for all drugs, including insulin. The IDS for Insulin takes the same software and technology and is a dosing calculator developed for a single patient and single drug. This device allows the patient to participate in insulin dosing, but has the added benefit of a remote patient management that allows the healthcare professional to view patient data anytime from anywhere with internet access

My Insulin Doser/IDS uses the same fundamental technology features as the Intelligent Dosing System and delivers the same level of effectiveness. Therefore, it is concluded that there is no significant difference in the basic function, safety and effectiveness between the Intelligent Dosing System (Predicate Device) and My Insulin Doser/IDS.

#### NON-CLINICAL PERFORMANCE DATA AND CONCLUSIONS FROM NON-CLINICAL TESTS

Labeling Comprehension study for 55 patient users to implement and understand My Insuiln Doser system. The system was designed around a 6th grade reading comprehension level. Subjects ranged in age from 18-80 years old.

Subjects were given My Insulin Doser/IDS users manual to read as well as given a demonstration of the software on any of the 3 platforms (PALM, PC and WEB),

{3}------------------------------------------------

## CLINICAL PERFORMANCE DATA AND CONCLUSIONS FROM CLINICAL TESTS

The study was a Phase IV in which nursing home residents who were maintained on two or more OADs were converted to an insulin replacement algorithm. The insulin titration algorithm developed by Dimensional Dosing Systems, Inc. was employed to help stabilize blood glucose following conversion from OAD therapy.

| SOFTWARE INFORMATION                         |          |
|----------------------------------------------|----------|
| Software Documentation                       | Concern  |
| Level of Concern                             | Major    |
| Software Description                         | Provided |
| Device Hazard Analysis                       | Provided |
| Software Requirements Specifications         | Provided |
| Architecture Design Chart                    | Provided |
| Software Design Specifications               | Provided |
| Traceability Analysis                        | Provided |
| Software Development Environment Description | Provided |
| Verification and Validation Documents        | Provided |
| Revision Level History                       | Provided |
| Unresolved Anomalies (Bugs or Defects)       | Provided |

# SOFTWARE INFORMATION

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in a simple, black line drawing, and the text is also in black. The overall design is clean and professional.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# NOV - 7 2008

Dimensional Dosing Systems C/O Mr. E.J. Smith Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114

Rc: K082512

Trade/Device Name: My Insulin Doser/IDS Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: August 29, 2008 Received: August 29, 2008

Dear Mr.Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sxythe y. Michino ms
FOR DR. CHIU LIN

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(K) Number:

Device Name: My Insulin Doser/IDS

Intended Use Statement:

My Insulin Doser/IDS allows a diabetic patient to calculate the best next dose of insulin to achieve a personal glucose target.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Centering Di non

Page of

flo

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Ancolnoon - O.
Infection Control, Dental Devices

K # 8 2512____________________________________________________________________________________________________________________________________________________________________ 510(k) Number: _

---

**Source:** [https://fda-staging.innolitics.com/device/K082512](https://fda-staging.innolitics.com/device/K082512)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).