← Product Code [NAY](/productcode/NAY) · K082497

# INTUITIVE SURGICAL ENDOWRIST ONE HOT SHEARS INSTRUMENT (K082497)

_Intuitive Surgical, Inc. · NAY · May 7, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082497

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/productcode/NAY.md)
- **Decision Date:** May 7, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Intuitive Surgical EndoWrist One Hot Shears Instrument is intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery.

## Device Story

EndoWrist One Hot Shears is a single-use disposable endoscopic instrument for use with da Vinci and da Vinci S Surgical Systems. It performs tissue cutting and coagulation via electrosurgical energy. The device features an integrated insulating Tip Cover. It interfaces with the surgical system identically to reusable EndoWrist instruments; the system tracks and decrements the instrument's single working life, preventing reuse. Used by surgeons in an OR setting, the device provides mechanical control of the instrument tip through the robotic system. The output is the physical manipulation of tissue for surgical procedures, benefiting patients by enabling minimally invasive surgery. The device is supplied sterile.

## Clinical Evidence

Bench testing only. Design analysis and functional testing confirmed that the subject device meets design input requirements and is substantially equivalent to the predicate.

## Technological Characteristics

Single-use disposable endoscopic instrument; electrosurgical cutting/coagulation; compatible with da Vinci/da Vinci S systems (IS1200/IS2000); integrated insulating Tip Cover; sterile; identical mechanical interface and system-tracked working life (1 life) as predicate.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Intuitive Surgical EndoWrist Monopolar Curved Scissors Instrument ([K050005](/device/K050005.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

082497

## Section IV

# MAY - 7 2009

### 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### 510(k) Number:

| Date              | August 28, 2008                                                                                                                         |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Submitter         | Intuitive Surgical, Inc.                                                                                                                |
|                   | 950 Kifer Road                                                                                                                          |
|                   | Sunnyvale, CA 94086                                                                                                                     |
| ER Number         | 2955842                                                                                                                                 |
| Contact           | James Farnworth                                                                                                                         |
|                   | Director, Regulatory Affairs                                                                                                            |
|                   | Telephone: (408) 523 - 8687                                                                                                             |
|                   | Fax: (408) 523 - 1390                                                                                                                   |
|                   | E-mail: james.farnworth@intusurg.com                                                                                                    |
| Subject Device    | Trade Name(s):                                                                                                                          |
|                   | Intuitive Surgical® EndoWrist® One Hot Shears Instrument                                                                                |
|                   | Classification Name:                                                                                                                    |
|                   | System, Surgical, Computer Controlled Instrument (21 CFR 876.1500)                                                                      |
|                   | Common Name:                                                                                                                            |
|                   | Endoscopic Instrument Control System,                                                                                                   |
|                   | Endoscopic Instruments and Accessories                                                                                                  |
|                   | Device Class:                                                                                                                           |
|                   | Intuitive Surgical da Vinci Surgical System and Endoscopic Instruments: Class II, NAY, System, Surgical, Computer Controlled Instrument |
| Predicate Devices | Intuitive Surgical EndoWrist Monopolar Curved Scissors Instrument, K050005                                                              |
| Device            | This 510(k) is being submitted for an endoscopic instrument, the                                                                        |

P. 1/3

{1}------------------------------------------------

#### Description

EndoWrist One Hot Shears Instrument, to be used with the Intuitive Surgical da Vinci and da Vinci S Surgical Systems. The Intuitive Surgical EndoWrist® instrument family currently consists of resposable endoscopic tools including the predicate EndoWrist® Monopolar Curved Scissors Instrument. The subject Hot Shears instrument has the same intended use as the Monopolar Curved Scissors but is a single-use disposable endoscopic tool.

This Intuitive Surgical instrument will be compatible with the previously cleared da Vinci and da Vinci S Endoscopic Instrument Control Systems Models IS1200 and IS2000.

Indications for The Intuitive Surgical EndoWrist One Hot Shears Instrument is Use intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery.

## Comparison to Predicate Device

The EndoWrist One Hot Shears Instrument is identical in shape, function, activation, and use as the predicate reusable Monopolar Curved Scissors instrument. Specific design and material changes were made in order to create a functionally equivalent instrument intended only for a single surgical procedure. Additionally, the trade name has been updated from EndoWrist to EndoWrist One to hight the single-use characteristic of the new instrument.

The overall scissor shape and size of the Hot Shears are essentially identical to that of the Monopolar Curved Scissors. The proximal housing and chassis have the same physical dimensions as the predicate instrument, but are provided with a different color scheme and labeling to aid in identification.

The predicate Monopolar Curved Scissors instrument uses a sterile removable and disposable Tip Cover accessory which insulates the clevis portion of the distal instrument tip. This Tip Cover is provided as an accessory sciparate from the instrument. The subject Hot Shears incorporates a Tip Cover with the same function and insulating material directly into its assembly and the user receives the sterile instrument with the tip cover already in place on the distal tip.

The single-use Hot Shears Instrument is compatible for use with all electrosurgical generators already cleared for use with the reusable 

{2}------------------------------------------------

Monopolar Curved Scissors instrument.

The Hot Shears instrument interfaces with the da Vinci and da Vinci S surgical systems in an identical manner to the reusable instruments. Both reusable EndoWrist instruments and the single use Hot Shears are assigned a specific number of working lives, and the lives are decremented in exactly the same manner by the da Vinci Systems. The Hot Shears are assigned only one life, and the systems will not recognize the instrument after that one life is expired, ensuring that it cannot be used again with the da Vinci This is identical to the manner in which a reusable systems. instrument expires.

The EndoWrist One Hot Shears is provided to the customer sterile and packaged in a sealed tray and outer carton. The predicate Monopolar Curved Scissors is provided non-sterile, and the user is responsible for cleaning and sterilizing the instrument prior to each use.

Labeling updates for the Hot Shears Instrument consist of new instrument housing markings, new tray and box labeling and a new Instructions For Use.

| Technological<br>Characteristics | The technological characteristics of the subject devices are identical<br>to the predicate devices.                                                                                                                                                                                                  |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Data              | Design analysis and comparison, as well as bench testing, have<br>been conducted to confirm that basic functional characteristics of<br>the subject device are substantially equivalent to the predicate<br>device cited, and that design output meets the design input<br>requirements.             |
| Conclusion                       | Based upon the technical information, Indications for Use and<br>performance information provided in this pre-market notification,<br>the EndoWrist One Hot Shears Instrument described herein has<br>been shown to be substantially equivalent to the current legally<br>marketed predicate device. |

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.

Public Health Service

MAY - 7 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Intuitive Surgical, Inc. % Ms. James Farnworth Director, Regulatory Affairs 950 Kifer Road Sunnyvale, California 94086

Re: K082497

Trade/Device Name: Intuitive Surgical® EndoWrist® One Hot Shears Instrument Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NAY Dated: March 26, 2009 Received: March 27, 2009

Dear Ms. Farnworth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at

{4}------------------------------------------------

Page 2 - Ms. James Farnworth

(240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Daniel Keane

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known): K082497

Intuitive Surgical® EndoWrist® One Hot Shears Instrument Device Name:

Indications For Use:

The Intuitive Surgical EndoWrist One Hot Shears Instrument is intended to be used with the da Vinci and da Vinci S Surgical Systems for endoscopic cutting and coagulation of tissue during surgery,

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neike Oden for mkm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K082497

EndoWrist® One Hot Shears Instrument

---

**Source:** [https://fda-staging.innolitics.com/device/K082497](https://fda-staging.innolitics.com/device/K082497)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).