K082484 · Genesis Biosystems, Inc. · ONF · Sep 12, 2008 · General, Plastic Surgery
Device Facts
Record ID
K082484
Device Name
ACCELAWAVE SYSTEM
Applicant
Genesis Biosystems, Inc.
Product Code
ONF · General, Plastic Surgery
Decision Date
Sep 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Accelawave System is indicated for: The removal of unwanted hair and to effect stable long-term or permanent hair reduction. Permanent hair reduction is defined as a long-term stable reduction in the number of hairs re-growing after a treatment regimen. The treatment of moderate inflammatory acne vulgaris. The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles). The treatment of cutaneous lesions including warts, scars, and striae. The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas, and spider angiomas, poikiloderma of civatte, leg veins and venous malformations. The integrated thermal cooling is indicated for use in cooling the epidermis at the treatment site prior to, during, and after light treatment in general aesthetic dermatologic and plastic surgery procedures. Reduce pain during and/or with light treatment (via partial anesthesia from cooling). Reduce discomfort during and/or associated with light treatment. Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation. Allow the use of higher light fluences for light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions). Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).
Device Story
Accelawave System is an Intense Pulsed Light (IPL) device using xenon flash lamp technology. It emits light (420nm-1200nm) via a 15mm x 45mm applicator to the skin. Device includes integrated water-cooled heat transfer network for epidermal cooling. Operated by clinicians in aesthetic/dermatologic settings. Operator selects repetition rate (≤0.5Hz) and wavelength filters based on skin type and condition. Cooling applied before, during, and after light delivery to protect epidermis, reduce pain, and allow higher fluences. Output affects clinical decision-making by enabling targeted treatment of vascular/pigmented lesions and hair reduction. Benefits include reduced thermal injury risk and side effects like scarring or pigmentation changes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological comparison to legally marketed predicate devices and bench testing compliance with UL60601-1:1996 and European Medical Directive 93/42/EEC.
Technological Characteristics
Intense Pulsed Light (IPL) system using xenon flash lamp. Wavelength range: 420nm-1200nm. Treatment area: 15mm x 45mm. Repetition rate: ≤0.5Hz. Integrated water-cooled heat transfer network for epidermal cooling. Complies with UL60601-1:1996 and 93/42/EEC.
Indications for Use
Indicated for Fitzpatrick skin types I-VI for hair removal, acne vulgaris, benign pigmented epidermal lesions, cutaneous lesions (warts, scars, striae), and benign cutaneous vascular lesions. Integrated cooling indicated for epidermal cooling during aesthetic dermatologic/plastic surgery procedures to reduce pain, discomfort, and thermal injury risk.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
.
## 510(k) Summary of Safety and Effectiveness
Accelawave System
510(k) Number KO82484_
| Applicant: | Genesis Biosystems |
|-------------------------|-----------------------------------------------|
| Address: | 1500 Eagle Court Lewisville, TX 75057 |
| Contact Person: | Brian A. Buss, Director of Compliance |
| Telephone: | (972) 315-7888 |
| Fax: | (972) 315-7818 |
| Email: | bbuss@genesisbiosystems.com |
| Date prepared: | 04 / 23 / 2008 |
| Device Trade Name: | Accelawave System |
| Common Name: | Intense Pulsed Light (IPL) |
| Classification Name: | Laser Surgical Instrument 21 C.F.R § 878.4810 |
| Product Classification: | Class II device |
| Product Code: | ONF |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Genesis Biosystems. The word "Genesis" is in a larger, sans-serif font, with a small tree graphic above the "sis" portion of the word. Below "Genesis" is the word "Biosystems" in a smaller font.
| Legally Marketed<br>Predicate Device: | The Accelawave System is substantially equivalent in terms of<br>technological characteristics, performance, intended use,<br>indications for use and operator interface to; |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Chromogenex (K053324) |
| | • Sciton (K032460) |
| | • Lumenis (K020839) |
| System Description: | The Accelawave is an Intense Pulsed Light-based medical device<br>utilizing xenon flash lamp technology to illuminate the dermis to<br>offer light based therapies as listed in the indications of use. The<br>Accelawave emits *light at 420nm to 1200nm via a 15mm x 45mm<br>treatment area at a repetition rate equal to or less than 0.5Hz<br>(operator selective).<br>*Light is delivered to the skin surface while a water-cooled heat<br>transfer network cools the contacting surfaces and the epidermis. |
| Performance<br>Standards: | The device complies with the European Medical Directive<br>93/42/EEC concerning medical devices, and will comply with<br>voluntary standards UL60601-1:1996 when marketed in the U.S. |
Indications for use: Indications for Use for Fitzpatrick skin types (I to VI)
- The removal of unwanted hair and to effect stable long-term or *permanent hair . reduction. *Permanent hair reduction is defined as a long-term stable reduction in the number of hairs re-growing after a treatment regimen.
- The treatment of moderate inflammatory acne vulgaris .
- o The treatment of benign pigmented epidermal lesions including dyschromia. hyperpigmentation, melasma, ephelides (freckles)
- The treatment of cutaneous lesions including warts, scars, and striae. 0
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Genesis Biosystems. The word "Genesis" is in a large, bold font, and the word "Biosystems" is in a smaller font below it. There is a small graphic above the word "Genesis" that looks like a cluster of cells.
- The treatment of benion cutaneous vascular lesions, including port wine stains. . hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas, and spider angiomas, poikiloderma of civatte, leg veins and venous malformations.
The integrated thermal cooling is indicated for use in cooling the epidermis at the treatment site prior to, during, and after treatment in general aesthetic dermatologic and plastic surgery procedures.
- Reduce pain during and/or with light treatment (via partial anesthesia from cooling) .
- Reduce discomfort during and/or associated with light treatment .
- Minimize thermal injury, including thermal necrosis, to non-target skin and skin . structures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper - and/or hypopigmentation
- Allow the use of higher light fluences for light or laser treatments (such as for hair . removal and the treatment of vascular or pigmented lesions)
- Reduce potential side effects of light or laser treatments (such as for hair removal and . the treatment of vascular or pigmented lesions)
- The Accelawave System is substantially equivalent to its predicate Conclusion: devices, based upon predicate similarities. The skin is treated in the same manner as the predicate devices, Intense Pulsed Light and dermal cooling.
| Similarities/<br>Differences | Lumenis Vasculite | Sciton BBL | Chromalight |
|--------------------------------------|------------------------------------------------------|---------------------------------|--------------------------|
| Output Spectrum<br>Characteristics | Similar | Similar | Similar |
| Output Pulse<br>Characteristics | Similar,<br>Programmable | Similar,<br>Programmable | Similar,<br>Programmable |
| Treatment<br>Parameters | Similar,<br>Programmable | Similar,<br>Programmable | Similar |
| Skin<br>Cooling/Epidermal<br>Cooling | Cold Gel vs. Cold<br>Contact Cooling<br>(difference) | Similar Cold<br>Contact Cooling | none |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Genesis Biosystems. The word "Genesis" is on the top line, and the word "Biosystems" is on the second line. There is a graphic above the word "Genesis" that looks like a tree.
- Side Effects: In extreme cases, effects from treatment can include excessively red patches in the shape of the applicator head and blistering. If this occurs the tissue should be cooled and cared for as would normally be the case with burns to the skin, i.e. do not burst any blister formation, keep clean and covered until healed.
If blisters form, they are usually intra-epidermal in nature and heal without scarring. Inappropriate management of blisters during the healing stage will increase the chance of scarring.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Genesis Biosystems, Inc. % Underwriters Laboratories, Inc. Mr. Casey Conry Senior Project Engineer 1285 Walt Whitman Road Melville, New York 11747
AUG 2 6 2011
Re: K082484
Trade/Device Name: Accelawave System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: ONF
Dated: August 26, 2008 Received: August 28, 2008
Dear Mr. Conry:
This letter corrects our substantially equivalent letter of September 12, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other
{5}------------------------------------------------
Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark A Millikan
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Genesis Biosystems. The word "Genesis" is written in a bold, sans-serif font, with a graphic above the "sis" portion of the word. Below "Genesis" is the word "Biosystems", also in a bold, sans-serif font, but slightly smaller than the word above it. The logo is black and white.
| | Indications for Use | |
|----------------------------------------------------------|---------------------|--------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K082484 | |
| Device Name: | Accelawave System | (Division Sign-Off)<br>Division of General, Restorative,<br>and Neurological Devices |
| Indications for Use for Fitzpatrick skin types (I to VI) | 510(k) Number | K082484 |
Indications for Use for Fitzpatrick skin types (I to VI)
- The removal of unwanted hair and to effect stable long-term or "permanent hair ● reduction.
*Permanent hair reduction is defined as a long-term stable reduction in the number of hairs re-growing after a treatment regimen.
- . The treatment of moderate inflammatory acne vulgaris
- . The treatment of benign pigmented epidermal lesions including dyschromia. hyperpigmentation, melasma, ephelides (freckles)
- The treatment of cutaneous lesions including warts, scars, and striae. .
- The treatment of benign cutaneous vascular lesions, including port wine stains, . hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas, and spider angiomas, poikiloderma of civatte, leg veins and venous malformations.
The integrated thermal cooling is indicated for use in cooling the epidermis at the treatment site prior to, during, and after light treatment in general aesthetic dermatologic and plastic surgery procedures.
- Reduce pain during light treatment (via partial anaesthesia from cooling) ●
- Reduce discomfort during and/or associated with light treatment �
- . Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper - and/or hypo pigmentation
- Allow the use of higher light or laser fluences for light treatments (such as for hair . removal and the treatment of vascular or pigmented lesions)
- Reduce potential side effects of light treatments (such as for hair removal and the . treatment of vascular or pigmented lesions)
Section #5 - 1
PAGE 1 OF 2
MNM
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K082484
awave Device Name: ACC
Indications For Use:
## Indications and Filters vs. Skin Types
| CONDITIONS | FILTER SETTINGS and WAVELENGTH RANGE | I | II | III | IV | V | VI |
|--------------------------------|--------------------------------------|----------|----------|----------|----------|----------|----------|
| Hair (course) | | 640-1200 | 640-1200 | 640-1200 | 640-1200 | 690-1200 | 690-1200 |
| Hair (fine) | | 640-1200 | 640-1200 | 640-1200 | 640-1200 | 690-1200 | 690-1200 |
| Acne Vulgaris | | 420-1200 | 420-1200 | 420-1200 | 420-1200 | 510-1200 | N/A |
| Pigmented<br>Epidermal Lesions | | | | | | | |
| a) Dyschromia | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| b) Hyperpigmentation | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| c) Melasma | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| d) Ephelides | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| Cutaneous Lesions | | | | | | | |
| a) Warts | | 420-1200 | 420-1200 | 420-1200 | 420-1200 | 420-1200 | N/A |
| b) Scars | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| c) Striae | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| Cutaneous Vascular<br>Lesions | | | | | | | |
| a) Port Wine Stain (Child) | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| b) Port Wine Stain (Adult) | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| c) Hemangiomas | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| d) Telangiectasias | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| e) Rosacea | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| f) Angiomas | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| g) Spider Angiomas | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| h) Poikiloderma | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| i) Venous Malformations | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| Leg Veins | | | | | | | |
| a) Small | | 510-1200 | 510-1200 | 510-1200 | 560-1200 | 560-1200 | N/A |
| b) Medium | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
| c) Large | | 560-1200 | 560-1200 | 560-1200 | 560-1200 | 560-1200 | N/A |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Ogle formxn
(Division Sign-Off)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 2
mnm
510(k) Number K082484
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.