K082480 · Contec Medical System Co., Ltd. · KNG · Feb 25, 2009 · Obstetrics/Gynecology
Device Facts
Record ID
K082480
Device Name
CONTEC POCKET FETAL DOPPLER
Applicant
Contec Medical System Co., Ltd.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Feb 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
Contect Pocket Fetal Doppler (SONOLINE A, SONOLINE B, Baby Sound A and Baby Sound B) are hand-held, battery powered audio Doppler devices used for detecting fetal heart beats. They are all available for user replaceable batteries. The user interface includes power button, mode button, volume control, single speaker, headphone jack and LCD display for heart rate, battery and working mode, probe type. SONOLINE A and SONOLINE B includes four interchangeable probes (2MHz normal probe,2MHz water proof probe,3MHz normal probe,3MHz water proof probe), while Baby Sound A and Baby Sound B are integrated with 2MHz probe.
Device Story
Hand-held, battery-powered audio Doppler device; detects fetal heart beats using ultrasonic transducers (2MHz or 3MHz). Input: ultrasonic signals from fetus; processing: Doppler shift analysis; output: audible heart sounds via speaker/headphone jack and visual heart rate/mode/battery status on LCD. Used in clinical or home settings by clinicians or patients. Healthcare providers use audible and visual outputs to monitor fetal status; facilitates assessment of fetal well-being. Models include Sonoline A/B (interchangeable probes) and Baby Sound A/B (integrated probe).
Clinical Evidence
Bench testing only. No clinical data provided. Validation included electrical safety and electromagnetic compatibility (EMC) testing to demonstrate compliance with consensus standards.
Indicated for fetal heart beat detection in pregnant patients. Used for diagnostic ultrasound fluid flow analysis via continuous wave (CW) Doppler mode.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
KORZYSTN
# Section III 510(k) Summary
FEB 2 5 2009
As Required by CFR 807.92
Sponsor:
Contec Medical Systems Co., LTD No.2-1 Hengshan Road, Economic and Technical Development Zone Qinhuangdao, Hebei, 066000, People's Republic of China
Mr. Li Xueyong Quality Manager Tel: Fax: Email:
Correspondent:
Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, China
Ms. Diana Hong / Mr. Lee Fu Tel: 021-64264467 Fax: (760)466-5084 Email: diana.hong@mid-link.net
Proposed Device:
Device Trade/Proprietary Name: Contec Pocket Fetal Doppler Model: SONOLINE A/ SONOLINE B/Baby Sound A/Baby Sound B Device Common Name: Fetal ultrasonic monitor and accessories Device Classification Name: monitor, ultrasonic, fetal Product Code: KNG Regulation Number: 884.2660 Device Class: II
7
{1}------------------------------------------------
# Intended Use:
Predicate Device
Contect Pocket Fetal Doppler (SONOLINE A, SONOLINE B, Baby Sound A and Baby Sound B) are hand-held, battery powered audio Doppler devices used for detecting fetal heart beats. They are all available for user replaceable batteries. The user interface includes power button, mode button, volume control, single speaker, headphone jack and LCD display for heart rate, battery and working mode, probe type.
SONOLINE A and SONOLINE B includes four interchangeable probes (2MHz normal probe,2MHz water proof probe,3MHz normal probe, 3MHz proof resistance probe), while Baby Sound A and Baby Sound B are integrated with 2MHz probe.
LifeDop Doppler Ultrasound System K-number: K024197
Product Code: KNG
## Intended Use:
The LifeDop is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats and for blood flow detection in veins and arteries. the product includes four interchangeable probes(OB Late Term, OB Early Term, Vascular pencil probe, Vascular flat face probe) and user replaceable batteries. the user interface includes an on/off button, play/record button, volume control, single 2-1/4" speaker, headphone jack and LCD display for heart rate, battery and waveform information.
# Manufactured by:
Summit Doppler Systerms, Inc. Add:5350 Vivian St. Suite A. Arvada, CO 80002-1957 Tel:(303)423-7572 Fax:(303)431-5994
# Device Description
Contect Pocket Fetal Doppler includes four models, SONOLINE A. SONOLINE B, Baby Sound A and Baby Sound B. They are handheld, battery powered audio Doppler devices used for detecting fetal heart beats. They are all available for user replaceable batteries.
SONOLINE A and SONOLINE B include four interchangeable probes (2MHz normal probe, 2MHz water proof probe, 3MHz normal probe and 3MHz proof resistance probe), while Baby Sound A and Baby Sound B are integrated with 2MHz probe.
8
{2}------------------------------------------------
Testing Laboratory testing was conducted to validate and verify that Contec Pocket Fetal Doppler met all design specifications, including electrical safety, EMC, specification. Results of these tests demonstrate compliance to the requirements of all consensus standards Substantially Equivalent
The proposed device, Contec Pocket Fetal Doppler, is substantially equivalent (SE) to the predicate device LifeDop Doppler Ultrasound System (K024197).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the left side of the eagle. The eagle symbol is composed of three curved lines that suggest the shape of an eagle in flight.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Contec Medical Systems Co., Ltd. % Ms. Diana Hong General Manager Shanghai Mid-Link Business Consultant Co., Ltd. Suite 8D, No. 19, Lane 999, Zhongshan No.2 Road Shanghai, 200030 CHINA Re: K082480 Trade/Device Name: Contec Pocket Fetal Doppler
FEB 2 5 2009
Models: Sonoline A/B and Baby Sound A/B Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: February 17, 2009 Received: February 17, 2009
## Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Intersule octimation of prices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include the Evilual Connents for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Contec Pocket Fetal Doppler - Models: Sonoline A/B and Baby Sound A/B, as described in your premarket notification:
## Transducer Model Number
2MHz Normal Probe CMS-150-T0 2MHz CW Water Proof CMS-150-T1 3MHz CW Normal Probe CMS-150-T5 3MHz CW Water Proof Probe CMS-150-T2
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.
Sincerely yours,
Lori McWhany
An Jahine M. Morris Acting Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
# Exhibit D Indication for Use Form
510(k) Number: K082480 Device Name: Contec Pocket Fetal Doppler
## Indications for Use:
Contect Pocket Fetal Doppler (SONOLINE A, SONOLINE B, Baby Sound A and Baby Sound B) are hand-held, battery powered audio Doppler devices used for detecting fetal heart beats. They are all available for user replaceable batteries. The user interface includes power button, mode button, volume control, single speaker, headphone jack and LCD display for heart rate, battery and working mode, probe type.
SONOLINE A and SONOLINE B includes four interchangeable probes (2MHz normal probe,2MHz water proof probe,3MHz normal probe,3MHz water proof probe), while Baby Sound A and Baby Sound B are integrated with 2MHz probe.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arun Th Thang
(Division Sign-Off) Division of Reproductive, Abdominal an Radiological Devices 510(k) Number
Page 1 of 7
{6}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
### Device Name: SONOLINE A and SONOLINE B fetal Doppler
## Main unit fetal system with either 2MHz CW or 3.0MHz CW
#### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | N | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletalConventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The system consists of main unit with either 2MHz CW or 3,2 MHz CW for fetal applications. Only one transducer can be used with the main unit at one time
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Page 2 of
Tony M. Whang
---
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K082480
{7}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
# Device Name: 2MHz normal probe CMS-150-T0
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | N | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The above is a 2.0MHz CW transducer for fetal heart detection
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Page 3 of 7
Lomi M. Wz
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
# Device Name: 2MHz CW water proof probe CMS-150-T1
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
|------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletalConventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The above is a 2.0MHz CW transducer for fetal heart detection.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Page 4 of
Vonghi M. Whang
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{9}------------------------------------------------
# Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
## Device Name: 3MHz CW normal probe CMS-150-T5
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The above is a 3.0MHz CW transducer for fetal heart detection.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Donna M. Whitten
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page 5 of 7
{10}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
## Device Name: 3MHz CW water proof probe CMS-150-T2
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The above is a 3.0MHz CW transducer for fetal heart detection
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Page 6 of 7
Arguello Whan
(Division Sign-Off) Division of Reproductive, Abdominat and Radiological Devices 510(k) Number
{11}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
## Device Name: Baby Sound A and Baby Sound B
#### Main unit fetal system ingreted with 2MHz CW transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B/M) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurologica | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletalConventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Comments:
The system consists of main unit is integrated with 2MHz CW transducer for fetal applications.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Page
Tony M. Wang
(Division Sigh-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.