K082421 · Emageon, Inc. · DQK · Dec 17, 2008 · Cardiovascular
Device Facts
Record ID
K082421
Device Name
HEARTSUITE HEMODYNAMICS 7.10
Applicant
Emageon, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Dec 17, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
HeartSuite Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and CO2, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output. Some systems are built and designed to measure End Tidal CO2. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/110V. The system is intended for use in hospital cardiac catheterization laboratories. The Monitoring System (comprising the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC) is a medical product of the Class IIa (RL 93/42/EWG, Appendix IX).
Device Story
HeartSuite Hemodynamics is a diagnostic monitoring system for cardiac catheterization labs. It consists of a Patient Data Module (PDM) and a Hemo Monitor PC connected via serial interface. The PDM acquires and calculates vital parameters (ECG, SpO2, invasive BP, temperature, NIBP, CO2) and performs evaluations (arrhythmias, ST-segments, cardiac output). Data is transmitted to the PC for visualization. The system aids clinicians in real-time patient monitoring and hemodynamic assessment during procedures. It uses an isolated power supply and battery-backed PDM for patient safety. The device provides continuous physiologic data to support clinical decision-making during invasive cardiac procedures.
Clinical Evidence
No clinical performance data. Bench testing only. Validation included physiologic simulators to test ECG aberrations, R-wave detection, pressure/valve analysis, hemodynamic calculations, and End Tidal CO2 verification. Software integrity, data entry restrictions, command responses, and report generation were validated. Final product testing included patient isolation and leakage current tests.
Technological Characteristics
System comprises a Patient Data Module (PDM) and Hemo Monitor PC. PDM uses internal battery with external 12V/RS232 power; PC uses 230V/110V mains. Connectivity via serial interface. Isolated power and data transmission. Software performs signal processing for ECG, pressure, and hemodynamic calculations.
Indications for Use
Indicated for monitoring and recording physiologic data (ECG, SpO2, invasive BP, temperature, NIBP, CO2) and evaluating resting ECG, arrhythmias, ST-segments, and cardiac output in patients undergoing cardiac catheterization procedures in hospital settings.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
| 510(k) Summary | K082421 | DEC 1 7 2008 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Date Prepared: | June 27, 2008 | |
| Contact Person: | Inger Hanson<br>Executive Director, Quality Systems and Regulatory Affairs,<br>Management Representative, CPO<br>Emageon, Inc.<br>Phone: 262-369-3379<br>Fax: 262-367-0728 | |
Device Trade Name: HeartSuite Hemodynamics
Common Name: Programmable Diagnostic Computer
Classification Name: 21 CFR 870.1425, Cardiovascular
Predicate Substantially Equivalent Devices: GE Mac-Lab (K061741)
Device Description: HeartSuite Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.
The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and CO2, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output. Some systems are built and designed to measure End Tidal CO2.
The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface.
All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC.
The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/110V.
The system is intended for use in hospital cardiac catheterization laboratories.
The Monitoring System (comprising the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC) is a medical product of the Class IIa (RL 93/42/EWG, Appendix IX).
l
{1}------------------------------------------------
Intended Use: HeartSuite Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.
The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature. NIBP and CO2, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output. Some systems are built and designed to measure End Tidal CO2.
The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface.
All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC.
The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/110V.
The system is intended for use in hospital cardiac catheterization laboratories.
The Monitoring System (comprising the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC) is a medical product of the Class IIa (RL 93/42/EWG, Appendix IX).
Performance Data: In order to ensure that the HeartSuite Hemodynamics software performs safely and efficiently, the HeartSuite Hemodynamics software was tested using physiologic simulators. Various types of ECG aberrations, rates, amplitudes, and deviations were used for validation of R-Wave detection for the rate meter and timing measurements related to pressure analysis. Multiple pressures were simulated under various conditions to validate the pressure and valve analyses. Multiple oxygen saturations were entered to validate hemodynamic calculations. End tidal CO2 was validated through Schiller and verified through tests run using SimMan. Multiple erroneous and/or incongruous entries were made in data entry locations to validate data entry restrictions. Multiple cases were created and exported to validate data integrity. All command buttons were tested for their appropriate responses. Multiple reports were generated to validate the report generation functions. The software responded appropriately in the tests described. After assembly is completed each unit undergoes final products testing. Patient isolation and leakage current tests are to be performed on each unit prior to packaging for shipment.
In addition to functional testing, a risk analysis, requirements review and design review was performed.
{2}------------------------------------------------
No clinical performance data has been used to support the substantial equivalence claim.
Substantial equivalence summary: The Emageon HeartSuite Hemodynamics system is a comparable type and substantially equivalent to a legally marketed predicate device. The intended use of the Emageon HeartSuite Hemodynamics system is the same as that of the predicate device Mac-Lab, marketed by GE Healthcare. No new safety or effectiveness issues are raised with the Emageon HeartSuite Hemodynamics system. The subject device has substantially equivalent technological characteristics, features, specifications, materials, modes of operation and intended uses as a legally marketed predicate device.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 7 2008
Emageon, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 53313
Re: K082421
> Trade/Device Name: HeartSuite Hemodynamics Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (Two) Product Code: DOK Dated: November 25, 2008 Received: November 28, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
## Page 2 - Mr. Mark Job
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) prematket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
07 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): k082421
Device Name: HeartSuite Hemodynamics
Indications for Use:
HeartSuite Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.
The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and CO2, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output. Some systems are built and designed to measure End Tidal CO2.
The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface.
All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC.
The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/110V.
The system is intended for use in hospital cardiac catheterization laboratories.
The Monitoring System (comprising the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC) is a medical product of the Class IIa (RL 93/42/EWG, Appendix IX).
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|-------------------------------------------------|--------|------------------------------------------------|
|-------------------------------------------------|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
for B Zuckerman
| (Division Sign/off) | 12/16/08 |
|------------------------------------|----------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K082421 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.