# NAVIGATION PROBES (K082404)

_We Sense, Ltd. · GZL · Feb 25, 2009 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082404

## Device Facts

- **Applicant:** We Sense, Ltd.
- **Product Code:** [GZL](/productcode/GZL.md)
- **Decision Date:** Feb 25, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1330
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Navigation probes are intended to be used in the human brain for the recording of Neuro potentials of single brain cells and for electrical stimulation of small areas in the brain. We Sense Navigation Probes electrodes are intended to be used in operation room during neurosurgery for temporary (less than two hours) recording and stimulation of defined small area in the brain, and for recording electrical signals from target brain cells. These electrodes can be used only by professional staff and not sell over the counter

## Device Story

Navigation Probes are depth electrodes featuring a micro tip and a macro tip. Micro tip records single cell potentials or cell clusters and performs microstimulation; macro tip records local field potentials and performs stimulation. Used in operating rooms during neurosurgery by professional staff. Device provides electrical signal input to external recording/stimulation equipment; output allows clinicians to identify target brain areas for surgical guidance. Benefits include precise localization and stimulation of neural tissue during neurosurgical procedures.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to the predicate device.

## Technological Characteristics

Depth electrode with micro and macro tips. Intended for temporary (<2 hours) intracranial use. Technical specifications and materials are stated to be similar to the predicate FHC microTargeting Electrode (K033173).

## Regulatory Identification

A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

## Predicate Devices

- FHC microTargeting Electrodes ([K033173](/device/K033173.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "rwe sense", a company that specializes in brain sensing technology. The logo features a stylized graphic of a brain with a leaf-like shape surrounding it. The text "rwe sense" is written in a bold, sans-serif font, with the tagline "Pioneers in Brain Sensing Technology" underneath.

BETT BE RECT TELET ense com Web www we COLLERS CON

Image /page/0/Picture/2 description: The image shows a handwritten word that appears to be "Coryou." The letters are connected in a cursive style, and the writing is somewhat stylized. The word is written in black ink on a white background.

# 510(k) Summary

FEB 2 5 2009

As Required by 21 CFR 807.87(k)510 (k) Summary

## 1. Subscribers Name & Address

WE SENSE LTD 2,Eksal Street , Nazareth 16000, Israel Tel: (972) 4 656-4222 Fax: (972) 4 656-4282 Official Correspondent: Mr. Nabil Jadaon

## 2. Trade Name

Navigation Probes

### 3. Predicate Device Identification

| Common Name     | Product Code | Class | Regulation Number |
|-----------------|--------------|-------|-------------------|
| Depth electrode | 84 GZL       | II    | 21 CFR 882.1330   |

### 4. Predicate Device Identification

Legally marketed devices to which equivalence is being 510(k) # claimed K033173 FHC microTargeting Electrodes

### 6. Device Description

NavigationProbes are equipped with a Micro tip and a Macro tip. The micro tip is used to for recording cell potentials of single cells or of cell clusters and for MicroStimulation.

The Macro tip is used for stimulation and recording of local field potentials.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for "we sense", a company that specializes in brain sensing technology. The logo features a stylized leaf design to the left of the company name. Below the company name is the tagline "Pioneers in Brain Sensing Technology".

7405 P.O.B. 50736 Israel Tell 9722 282 Email:Info@we sense com Web ww

# 7. Intended Use:

The Navigation probes are intended to be used in the human brain for the recording of Neuro potentials of single brain cells and for electrical stimulation of small areas in the brain.

# 8. Summary of technological characteristics of Device and Predicate Device:

The functionality for the Navigation Probes are equivalent to its predicate device the FHC Inc, microTargeting Electrode (K033173) in safety and effectiveness. The fundamental technical characteristics are similar to those of the predicate device and are listed on the comparison charts provided in this 510 k submission.

Nabil Jadaon General Manager

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains an image of an eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the top of the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

We Sense Ltd. % Nabil Jadaon 2 Eksal St. P.O. Box 50736 Nazareth Israel 16000

FEB 2 5 2009

Re: K082404

Trade/Device Name: Navigation Probes Regulation Number: 21 CFR 882.1330 Regulation Name: Depth electrode Regulatory Class: II Product Code: GZL Dated: January 21, 2009 Received: February 2, 2009

Dear Mr. Jadaon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Sincerely yours,

F-n Mark Dec. 1, 2005
Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indication for Use

510(k) Number K 082404

Device Name: Navigation Probes

Indication For Use:

We Sense Navigation Probes electrodes are intended to be used in operation room during neurosurgery for temporary (less than two hours) recording and stimulation of defined small area in the brain, and for recording electrical signals from target brain cells. These electrodes can be used only by professional staff and not sell over the counter

Prescription Use _0111V (21 CFR Part 801 Subpart D)

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

And/Or

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Dra

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 1682404

Page 1 of 2

---

**Source:** [https://fda-staging.innolitics.com/device/K082404](https://fda-staging.innolitics.com/device/K082404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com