K082338 · Kyung Won Medical Co., Ltd. · MQV · Oct 16, 2008 · Orthopedic
Device Facts
Record ID
K082338
Device Name
POLYBONE GRANULE & BLOCK
Applicant
Kyung Won Medical Co., Ltd.
Product Code
MQV · Orthopedic
Decision Date
Oct 16, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PolyBone® Granule & Block is indicated for filling bone voids or defects that are not intrinsic to the stability of the bony structure. PolyBone® Granule & Block is to be packed into bony voids or gaps of the skeletal system (such as extremities, spine and the pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PolyBone® Granule & Block is a bone graft substitute that resorbs and it replaced with bone during the healing process.
Device Story
PolyBone® Granule & Block is a synthetic, resorbable bone void filler composed of 99.99% beta-tricalcium phosphate and 0.01% polyphosphate. It functions as an osteoconductive scaffold to support bone regeneration. The material features multidirectional interconnected porosity (75-85%) with pore sizes ranging from 200-500 μm. During the healing process, bone cells resorb the polyphosphate, and the scaffold is gradually replaced by new bone. The device is provided in various shapes and sizes for surgical implantation into bony voids or gaps in the skeletal system, such as the extremities, spine, or pelvis. It is intended for use by surgeons to fill defects that do not require intrinsic structural stability.
Clinical Evidence
No clinical data. Biocompatibility testing was performed in accordance with ISO 10993-1:2003, including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, subchronic toxicity, genotoxicity, and implantation studies. Results demonstrated the device is biologically safe.
Technological Characteristics
Synthetic resorbable bone void filler; 99.99% beta-tricalcium phosphate, 0.01% polyphosphate. Porous scaffold structure; 75-85% porosity; 200-500 μm pore size. Available in granule and block forms. Biocompatibility tested per ISO 10993-1:2003.
Indications for Use
Indicated for filling non-structural bone voids or defects in the skeletal system, including extremities, spine, and pelvis. Applicable to surgically created osseous defects or those resulting from traumatic injury. Not for use where structural stability is required.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Kasios TCP
Vitoss Scaffold
Submission Summary (Full Text)
{0}------------------------------------------------
K082338 (pg 1 of 2)
(yungwon Medical Co., Ltd.
Confidential
### 510(k) Summary
### OCT 1 6 2008
#### Date: 8-11-2008
- 1. Company making the submission
| | Submitter |
|---------|--------------------------------------------------------------------------------------------------|
| Name | Kyungwon Medical Co.,Ltd |
| Address | Suite 601 World Meridian Venture Center, 60-24<br>Gasan-Dong, Geumcheon-Gu, Seoul, Korea 153-801 |
| Phone | +82-2-2113-8693~8 |
| Fax | +82-2-2113-8699 |
| Contact | Shim, JaeBum Ph.D. Technical Director |
| Website | http://www.kyungwonmedical.co.kr |
- 2. Device:
Trade Name: PolyBone® Granule & Block Common Name: Bone Void Filler Classification Name: Resorbable Calcium Salt Bone Void Filler
- 3. Predicate Device: Kasios TCP, Vitoss Scaffold.
#### 4. Description:
PolyBone® Granule & Block is a synthetic resorbable calcium phosphate bone void filler consisted of 99.99% beta-tricalcium phosphate, and 0.01% polyphosphate. It is an osteoconductive material which provides a porous scaffold upon which bone formation can occur. The multidirectional interconnected porosity ranges from 75~85% with a 200~500 um pore size.
Mechanism of PolyBone® Granule & Block's bone regeneration works by bone cells resorbing polyphosphate, which is a bone regenerating palpation material. Indication for use of PolyBone® Granule & Block is a filler of bone void regions.
The device is available in a variety of shape and sizes.
{1}------------------------------------------------
KO82338
Kyungwon Medical Co., Ltd.
Confidential
- 5. Indication for use:
PolyBone® Granule & Block is indicated only for filling bone voids or defects that are not intrinsic to the stability of the bony structure. PolyBone® Granule & Block is to be gently packed into bony voids or gaps of the skeletal system (such as extremities, spine and the pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PolyBone® Granule & Block is a bone graft substitute that resorbs and is replaced with bone during the healing process.
#### 6. Review:
PolyBone® Granule & Block has the similar technological characteristics to the predicate device; components, indication for use, chemical and performance properties.
Components Similarities All devices are packed with a container
Indication for Use Similarities All devices have the same indication for use
Chemical Similarities All devices are made up from ß-Tricalcium phosphate family
Performance Properties Similarities
All devices are for filling bone voids or defects that are not intrinsic to the stability of the bony structure.
#### Biocompatibility
The biocompatibility of PolyBone® Granule & Block has been performed by ISO10993-1:2003; cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, subchronic toxicity, genotoxicity and implantation. The testing results show PolyBone® Granule & Block to be biologically safe.
#### 7. Conclusion
Based on the information provided in this premarket notification Kyungwon Medical Co.,Ltd. Concludes that PolyBone® Granule & Block is safe and effective and substantially equivalent to predicate device as described herein.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
OCT 1 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kyungwon Medical Co., Limited % Kodent, Inc. Jung Bae Bang US Agent 13340 East Firestone Boulevard, Suite J Santa Fe Springs, California 90670
Re: K082338
Trade/Device Name: PolyBone® Granule & Block Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: August 11, 2008 Received: August 15, 2008
Dear Jung Bae Bang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Jung Bae Bang
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N. Millison
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use
6082338 510(k) Number (if known)
# Device Name: PolyBone® Granule & Block
## Indications for Use:
PolyBone® Granule & Block is indicated for filling bone voids or defects that are not intrinsic to the stability of the bony structure. PolyBone® Granule & Block is to be packed into bony voids or gaps of the skeletal system (such as extremities, spine and the pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PolyBone® Granule & Block is a bone graft substitute that resorbs and it replaced with bone during the healing process.
Prescription Use X (21CRF 801 Subpart D)
Over-The-Counter Use AND/OR (21CRF 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | 14082338 |
|---------------|----------|
|---------------|----------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.