K082326 · Zonare Medical Systems, Inc. · IYN · Oct 8, 2008 · Radiology
Device Facts
Record ID
K082326
Device Name
ZONARE ULTRA ULTRASOUND SYSTEM
Applicant
Zonare Medical Systems, Inc.
Product Code
IYN · Radiology
Decision Date
Oct 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal; Intra-operative (abdominal, thoracic, and vascular); Intra-operative Neuro; Pediatric; Small Organ (Thyroid, Breast, Testes, etc.); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-cranial, trans-esoph (cardiac/noncardiac), musculo-skel conventional & superficial 3D/4D. cardiac adult & pediatric and other applications as shown in section 4.3.
Device Story
ZONARE z.one Ultra is a portable diagnostic ultrasound system consisting of a handheld scanner with integrated display and controls, and a docking station. It accepts removable transducer modules to acquire ultrasound signals from the patient. The system digitizes and preprocesses these signals to generate B-mode, M-mode, Pulsed Wave Doppler (PWD), Continuous Wave Doppler (CWD), and Color Doppler images. Used in clinical settings by physicians for diagnostic imaging and fluid flow analysis. Output is displayed on the integrated screen to assist in clinical decision-making across various anatomical applications. The system enhances workflow and provides new clinical indications compared to previous versions.
Clinical Evidence
No clinical data required. Safety and effectiveness were verified through bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety standards.
Technological Characteristics
Portable ultrasound system with removable transducer modules. Supports B-mode, M-mode, PWD, CWD, and Color Doppler. Connectivity via docking station. Complies with 21 CFR 820 and ISO 13485:2003. Sterilization/disinfection per standard clinical protocols.
Indications for Use
Indicated for qualified physicians to perform ultrasound evaluation of fetal/OB, abdominal, intra-operative (abdominal, thoracic, vascular, neuro), pediatric, small organ (thyroid, breast, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, trans-urethral, trans-esophageal (cardiac/non-cardiac), musculoskeletal (conventional/superficial), cardiac (adult/pediatric), and peripheral vascular applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K08 a326
# OCT 0 8 2008
## Summary of Safety and Effectiveness
Prepared in accordance with 21 CFR Part 807.92(c)
| The assigned 510(k) number is: | ____________________________ |
|--------------------------------|------------------------------|
|--------------------------------|------------------------------|
**Applicant Information:**
| <b>Date Prepared:</b> | July 14, 2008 |
|---------------------------|------------------------------------------------------------------------------------|
| <b>Name:</b> | ZONARE Medical Systems, Inc.<br>420 North Bernardo Ave.<br>Mountain View, CA 94043 |
| <b>Contact Persons:</b> | Linda J. Moore<br>Director, Regulatory Affairs & Quality Assurance |
| <b>Telephone Numbers:</b> | 650-230-2724 |
| <b>Fax Number:</b> | 650-230-2818 |
| <b>Email</b> | lmoore@zonare.com |
**Device Information:**
| <b>Trade Name:</b> | ZONARE z.one Ultra Ultrasound System |
|---------------------|--------------------------------------|
| <b>Device Name:</b> | ZONARE Diagnostic Ultrasound System |
| | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
Marketed Device(s): The ZONARE z.one Ultrasound System (K022858), GE Voluson i(K053435), Philips iU22 (K042540), and Siemens Acuson S2000 (K072786) devices currently in commercial distribution.
Device Description: The z.one Ultra is a general purpose diagnostic ultrasound system. It consists of a portable scanner approximately 8 inches wide and 3 inches high that can be held by the user in one hand and includes buttons for controlling the system and a screen that display ultrasound mages and user interface. The portable scanner can be held by the user in one hand accommodates a removable transducer module. Signals received from the transducer module are digitized and preprocessed. The transducer module comes into contact with the patient and receive ultrasound energy.
The docking station (aka carts) provides holders for the portable scanner, and transducer modules, as well as battery chargers and other accessories. The modification for this submission include new features and functionality that will enhance user convenience and work flow as well as provide clinicians with new indications of use.
510k Submittal ZONARE Medical Systems
{1}------------------------------------------------
Indications For Use: The device is intended for use by a qualified physician for utrasound evaluation of Fetal/OB; Abdominal; Intra-operative (abdominal, thoracic, and vascular); Intra-operative Neuro; Pediatric; Small Organ (Thyroid, Breast, Testes, etc.); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-cranial, trans-esoph (cardiac/noncardiac), musculo-skel conventional & superficial 3D/4D. cardiac adult & pediatric and other applications as shown in section 4.3.
Comparison with Predicate Device: With respect to features and applications, the ZONARE z.one Ultra is comparable and substantially equivalent to the currently marketed ZONARE z.one and the Voluson i, Siemens Acuson S2000, and Philips iU22 in terms of portability, features and functionality. Additionally, they have the same important safety and effectiveness features, as well as design, materials, and construction.
Non-clinical tests: The device has been evaluated according to the applicable medical device safety standards for acoustic output, biocompatibility, cleaning, and disinfection effectiveness as well as thermal, electrical, and mechanical safety.
### Clinical Tests: Non Required
Conclusion: ZONARE designs and develops their products according to 21 CFR 820, ISO 13485:2003 quality systems. The device conforms to applicable medical device safety standards and compliance for safety and effectiveness is verified through defined evaluation and market surveillance. Conforming to the required worldwide ultrasound standards, enables ZONARE to state that the ZONARE z.one Ultra Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to devices current cleared for market.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread.
OCT 0 8 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zonare Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K082326
Trade/Device Name: ZONARE System with Doppler & Harmonic Imaging Modes Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: September 20, 2008 Received: September 22, 2008
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the ZONARE System with Doppler & Harmonic Imaging Modes, as described in your premarket notification:
Transducer Model Number
C5-2 Curvilinear C6-2 Curvilinear C9-3 Curvilinear C8-3r (3D/4D) Curvilinear P4-1 Phased (Sector) Array P4-1c Phased (Sector) Array P10-4 Phased (Sector) Array E9-4 Endo-Cavity E9-4 3D Endo-Cavity L10-5 Linear
{3}------------------------------------------------
L8-3 Linear L14-5SP Linear L14-5W Linear E14-5 EndoScopic P8-3T Tran-Esophageal AUX CW2 AUX CW4 AUX CW5
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.
Sincerely yours,
for Joyce M. Whang, Ph.D.
Acting Director, Division of Reproductive,
Abdominal, and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
#### 4.3 Indications for use
This device is intended for use by a qualified physician for ultrasound evaluation of the following: Fetal, Abdominal, Intraoperative, Pediatric, Small organ/parts (breasttestes, thyroid, etc), Transrectal, Transcranial, Trans-esoph, Trans-urethral, OB/GYM, Cardiac, Pelvic, Neonatio. Vascular, 3D/4D, Tissue elasticity, Musculoskeletal (cardias, Superficial Musculoskeletal, and Peripheral Vascular applications.
#### Diagnostic Ultrasound Indication for Use
#### System: ZONARE System with Doppler & Harmonic Imaging Modes
#### Transducer: ZONARE System (Union of all Transducer Types)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|------------------------|------------------------------------------------|----|---|------|----------------|------------------|--------------------|-----------|
| General | Specific | B¹ | M | PWD² | CWD<br>CWD Aux | Color<br>Doppler | Combined<br>Modes⁴ | Other⁵, ⁸ |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P⁵ |
| | Abdominal | P | P | P | P | P | P | P⁵ |
| | Intra-operative<br>(Specify)⁶ | P | P | P | | P | P | P⁵ |
| | Intra-operative (Neuro) | P | | P | | P | P | P⁵ |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P⁵ |
| | Pediatric Aux | | | | N | | | |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.) | P | P | P | | P | P | P⁵<br>N⁸ |
| | Neonatal Cephalic | P | P | P | P | P | P | P⁵ |
| General<br>Application | Adult Cephalic | P | P | P | P | P | P | P⁵ |
| | Trans-rectal | P | P | P | | P | P | P⁵ |
| | Trans-vaginal | P | P | P | | P | P | P⁵ |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | N | N | N | N | N | N | N⁵ |
| | Musculo-skel.<br>(Conventional) | N | N | N | | N | N | N⁵, ⁸ |
| | Musculo-skel.<br>(Superficial) | N | N | N | | N | N | N⁵, ⁸ |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | N | N | N | | N | N | |
| | Cardiac Adult | P | P | P | P | P | P | P⁵ |
| | Cardia Adult Aux | | | | N | | | |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P⁵ |
| | Cardiac Pediatric Aux | | | | N | | | |
| | Trans-esoph. (Cardiac) | N | N | N | N | N | N | N⁵ |
| | Other (Specify)<br>(3D/4D) | N | N | N | N | N | N | |
| | Peripheral Vessel | P | P | P | N | P | P | P⁵<br>N⁸ |
| Peripheral vascular | Peripheral Vessel Aux | | | | N | | | |
| | Other (Specify)<br>(3D/4D) | N | N | N | | N | N | |
N = new indication; P = previously cleared by FDA 510k # 022852
Includes B-Mode and Harmonic imaging (HI)
Includes PWD-Mode imaging and High Pulse Repetition Rate PWD-Mode (HPRF)
Includes Color Doppler (CD) Dina
) * new indication; P = previously cleared by FDA S 10k # (22858)
Taclides PWD-Mode imaging (10)
1 moludes WD-Mode imaging add Highe Reputition Rate PWD-M-Repler (PD)
1 sent (CD, DPD, PD, or BD)
Color M-Mode (CM)
Abdominal includes renal, GYN/Pelvic
Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K082326
{5}------------------------------------------------
### Transducer: Curvilinear Transducers C5-2*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other7,8 |
|-------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|----------|
| General | Ophthalmic | | | | | | | |
| Ophthalmic | Fetal | P | P | P | | P | P | P5 |
| | Abdominal6 | P | P | P | | P | P | P5 |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P5 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General<br>applications | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | P | P | P | | P | P | P5 |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by FDA 510k # 022858 E added under Appendix E *market designation Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD-Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
్ Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Hale Leem
ivision Sidn-Off) Division of Reproductive, Abdomin Radiologic 510(k) Number
{6}------------------------------------------------
#### Transducer: Curvilinear Transducers C6-2*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-------------------------|---------------------------------------------------|-------------------|---|------|-----|-------------------|--------------------|--------|
| General | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other8 |
| Ophthalmic | Ophthalmic | | | | | | | |
| General<br>applications | Fetal | P | P | P | | P | P | P7 |
| | Abdominal6 | P | P | P | | P | P | P7 |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P7 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| Cardiac | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | P | P | P | | P | P | P5 |
N = new indication; P=previously cleared by FDA 510k # 022858 E= added under Appendix E *market designation Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
్ Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITEBELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helmut Werner
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{7}------------------------------------------------
#### Transducer: Curvilinear Transducers C9-3*
| Clinical Application | Specific<br>(Track I & III) | Mode of Operation | | | | | | |
|-------------------------|---------------------------------------------------|-------------------|---|------|-----|-------------------|--------------------|--------|
| General | | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other3 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | E | E | E | | E | E | E3 |
| | Abdominal6 | E | E | E | | E | E | E3 |
| | Intra-operative<br>(Abdominal) | N | N | N | N | N | N | N5 |
| | Intra-operative<br>(Vascular) | N | N | N | N | N | N | N5 |
| | Laparoscopic | | | | | | | |
| | Pediatric | E | E | E | | E | E | E3 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General<br>applications | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | N | N | N | N | N | N | N5 |
| | Musculo-skel.<br>(Superficial) | N | N | N | N | N | N | N5 |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Peripheral Vascular | E | E | E | | E | E | E3 |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P=previously cleared by FDA 510k # 022858 E= added under Appendix E * market designation,
Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD) S Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Helens
(Division Sign-Off) vision of Reproductive, Abdominal an Radiological De
510(k) Number
(k) Number K082346
{8}------------------------------------------------
#### Transducer: Curvilinear Transducers C8-3r* (3D/4D)
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Specific<br>(Track I & III) | B¹ | M | PWD² | CWD | Color<br>Doppler³ | Combined<br>Modes⁴ | Other⁵ |
|-------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|--------|
| General | Ophthalmic | | | | | | | |
| Ophthalmic | Fetal | N | N | N | | N | N | N⁵ |
| | Abdominal⁶ | N | N | N | | N | N | N⁵ |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | N | N⁵ |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General<br>applications | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | N | N | N | | N | N | N⁵ |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | N | N | N | | N | N | N⁵ |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by FDA 510k # 022858 E= added under Appendix E *market designation Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
1 Includes Color Doopler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
' Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITEBELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Halutleem
(Division Sign-Off) Division of Reproductive, Abdominal ar Radiologica 510(k) Number
{9}------------------------------------------------
### Transducer: Phased (Sector) Array Transducers P4-1*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|--------|
| General | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other8 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| | Abdominal | P | P | P | P | P | P | P |
| | Intra-operative<br>(Specify)7 | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P5 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| General<br>application | Neonatal Cephalic | P | P | P | P | P | P | P5 |
| | Adult Cephalic/ trans<br>cranial | P | P | P | P | P | P | P5 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | P | P | P5 |
| | Cardiac Pediatric | P | P | P | P | P | P | P5 |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| Peripheral<br>vascular | Peripheral Vascular | P | P | P | P | P | P | P5 |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by the FDA510k # 022858 E=added under Appendix E *market designation
Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD-Mode imaging and High Pulse Repetition Rate PWD-Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
1 Includes B+M , B+CM+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
5 Color M-Mode (CM)
.
* Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
· Freehand tissue elasticity
(PLEASE DO NOT WRITEBELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
uctive, Abdominal an
Radiological Devices
510(k) Number K082326
{10}------------------------------------------------
#### Transducer: Phased (Sector) Array Transducers P4-1c*
| Clinical Application | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other5,<br>8 |
|------------------------|---------------------------------------------------|---------------------|---|------|-----|-------------------|--------------------|--------------|
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P5 |
| | Abdominal6 | P | P | P | P | P | P | P5 |
| | Intra-operative<br>(Specify)7 | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P5 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | P | P | P | P | P | P | P5 |
| General<br>application | Adult Cephalic/ trans<br>cranial | P | P | P | P | P | P | P5 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P5 |
| | Cardiac Pediatric | P | P | P | P | P | P | P5 |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| | Peripheral | Peripheral Vascular | P | P | P | P | P | P |
| vascular | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P=previously cleared by the FDA510k # 022858 E= added under Appendix E *market designation
Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
5 Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
1 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene Reiner
livision Radiol 510/k) Number
{11}------------------------------------------------
#### Transducer: Phased (Sector) Array Transducers P10-4
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|--------|
| General | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other5 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P5 |
| | Abdominal | P | P | P | P | P | P | P5 |
| | Intra-operative<br>(Specify)7 | N | N | N | N | N | N | N5 |
| | Intra-operative<br>(Neuro) | N | N | N | | N | N | N5 |
| General<br>application | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P5 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | P | P | P | P | P | P | P5 |
| | Adult Cephalic/ trans<br>cranial | P | P | P | P | P | P | P5 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | P | P | P5 |
| | Cardiac Pediatric | P | P | P | P | P | P | P5 |
| | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| Peripheral<br>vascular | Peripheral Vascular | P | P | P | P | P | P | P5 |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by the FDA510k # 022858 E= added under Appendix E*market designation
Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
S Color M-Mode (CM)
� Abdominal includes renal, GYN/Pelvic
Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
· Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helena Reuss
(Division Sign-Off)
oductive. Abdominal a Radio 510(k)
{12}------------------------------------------------
#### Transducer: Endo-Cavity Transducers E9-4*
... ... ... ..
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|--------|
| General<br>(Track I Only) | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other8 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | P | P | P5 |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify)7 | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General<br>application | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | P | P | P5 |
| | Trans-vaginal | P | P | P | | P | P | P5 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | | | | | | | |
| | | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by FDA 510k # 02258 E = added under Appendix E *market designation E9-4 Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
5 Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
1 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITEBELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Herbert Hoover
(Division Sign-Off) oductive. Abdominal a Division of I Radiolo 510(k) Number
{13}------------------------------------------------
#### Transducer: Endo-Cavity Transducers* E9-4 3D
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------------------------|-------------------|---|------|-----|-------------------|--------------------|--------|
| General<br>(Track I Only) | Specific<br>(Track I & III) | B¹ | M | PWD² | CWD | Color<br>Doppler³ | Combined<br>Modes⁴ | Other⁸ |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | N | N | N | | N | N | N⁵ |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify)⁷ | | | | | | | |
| | Intra-operative<br>(Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| General<br>application | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | | N | N | N⁵ |
| | Trans-vaginal | N | N | N | | N | N | N⁵ |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | | | | | | | |
| | Musculo-skel.<br>(Superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify)<br>(3D/4D) | N | N | N | | N | N | N⁵ |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously cleared by FDA 510k # 022858 E= added under Appendix E *market designation
1 Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, or BD)
S Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene Newman
ion Sign-Off) productive. Abdomina Radiologica 510(k) Number
{14}------------------------------------------------
#### Transducer: Linear Transducers L10-5
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|-----------|
| General<br>(Track I Only) | Specific<br>(Track I & III) | B1 | M | PWD2 | CWD | Color<br>Doppler3 | Combined<br>Modes4 | Other5, 8 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | P | P | P5 |
| | Abdominal | P | P | P | | P | P | P5 |
| | Intra-operative<br>(Specify)7 | P | P | P | | P | P | P5 |
| | Intra-operative<br>(Neuro) | P | | P | | P | P | P5 |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P5 |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | P | P | P | | P | P | P5<br>N8 |
| General<br>application | Neonatal Cephalic | P | P | P | | P | P | P5 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | N | N | N | | N | N | N5, 8 |
| | Musculo-skel.<br>(Superficial) | N | N | N | | N | N | N5, 8 |
| | Intra-luminal | | | | | | | |
| | Other (Specify)7<br>(3D/4D) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | P | P | P | | P | P | P5<br>N8 |
| | Other (Specify)<br>3D/4D | | | | | | | |
N = new indication; P=previously cleared by the FDA 510k # (22858 E=added under Appendix E * market designation L10-5, 1 Includes B-Mode and Harmonic imaging (HI)
2 Includes PWDMode imaging and High Pulse Repetition Rate PWD-Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
S Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Hult Lewis
ductive, Abdominal ar Radiologi 510(k) Number
{15}------------------------------------------------
#### Transducer: Linear Transducers L8-3*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|----------|
| General<br>(Track I Only) | Specific<br>(Track I & III) | B¹ | M | PWD² | CWD | Color<br>Doppler³ | Combined<br>Modes⁴ | Other'8 |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | E | E | E | | E | E | E5 |
| | Abdominal6 | E | E | E | | E | P | E5 |
| | Intra-operative<br>(Specify)7 | E | E | E | | E | E | E5 |
| | Intra-operative<br>(Neuro) | E | | E | | E | E | E5 |
| | Laparoscopic | | | | | | | |
| | Pediatric | E | E | E | | E | E | E5 |
| General<br>application | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | E | E | E | | E | E | E5<br>N8 |
| | Neonatal Cephalic | E | E | E | | E | E | E5 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | N | N | N | | N | N | N5, 8 |
| | Musculo-skel.<br>(Superficial) | N | N | N | | N | N | N5, 8 |
| | Intra-luminal | | | | | | | |
| | Other (Specify)7<br>3D/4D | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | E | E | E | | E | E | E5<br>N8 |
| | Other (Specify)<br>3D/4D | | | | | | | |
N = new indication; P=previously cleared by the FDA 510k # 022858E= added under Appendix E *market designation Includes B-Mode and Harmonic imaging (HI)
2 Includes PWD Mode imaging and High Pulse Repetition Rate PWD Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
4 Includes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
S Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
7 Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
(Division Sign-Off)
8 Freehand tissue elasticity
Examples may include A-mode, Amplitude Doppler, 3-D imaging, Tissue Motion Doppler, Color velocity imaging
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene Reuer
Division of Reproductive. Abdominal a
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K082326
{16}------------------------------------------------
### Transducer: Linear Transducers L14-5SP*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------|----|---|------|-----|-------------------|--------------------|----------|
| General<br>(Track I Only) | Specific<br>(Track I & III) | B¹ | M | PWD² | CWD | Color<br>Doppler³ | Combined<br>Modes⁴ | Other⁸ |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | E | E | E | | E | E | E⁵ |
| | Abdominal⁶ | E | E | E | | E | E | E⁵ |
| | Intra-operative<br>(Specify)⁷ | E | E | E | | E | E | E⁵ |
| | Intra-operative<br>(Neuro) | E | | E | | E | E | E⁵ |
| | Laparoscopic | | | | | | | |
| | Pediatric | E | E | E | | E | E | E⁵ |
| | Small Organ<br>(Thyroid, Breast,<br>Testes, etc.) | E | E | E | | E | E | E⁵<br>N⁸ |
| General<br>application | Neonatal Cephalic | E | E | E | | E | E | E⁵ |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-<br>Card.) | | | | | | | |
| | Musculo-skel.<br>(Conventional) | N | N | N | | N | N | N⁵, ⁸ |
| | Musculo-skel.<br>(Superficial) | N | N | N | | N | N | N⁵, ⁸ |
| | Intra-luminal | | | | | | | |
| | Other (Specify)⁹<br>3D/4D | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Trans-esoph.<br>(Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vascular | Peripheral Vascular | E | E | E | | E | E | E⁵<br>N⁸ |
| | Other (Specify)<br>3D/4D | | | | | | | |
N = new indication; P=previously cleared by the FDA 510k # 022858 E= added under Appendix E *market designation Includes B-Mode and Harmonic imaging (HI)
Includes PWD.Mode imaging and High Pulse Repetition Rate PWD.Mode (HPRF)
3 Includes Color Doppler (CD), Directional Power Doppler (DPD), and Power Doppler (PD))
And Posted Bridge (CD), Directional Power Doppler (DPD), and Power Doppler (PD)
1 meludes B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD)
2 Colors B+M, B+M+CM, B+CD+M+CM, B+CD+PWD where CD could represent (CD, DPD, PD, or BD)
3 Color M Med
5 Color M-Mode (CM)
6 Abdominal includes renal, GYN/Pelvic
Intra operative include abdominal, thoracic (cardiac) and vascular (PV)
8 Freehand tissue elasticity
Examples may include A-mode, Amplitude Doppler, 3-D imaging, Tissue Motion Doppler, Color velocity imaging
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{17}------------------------------------------------
#### Transducer: Linear Transducers L14-5W*
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.