← Product Code [LLZ](/productcode/LLZ) · K082318

# CENTRICITY PACS IW (K082318)

_Ge Healthcare Dynamic Imaging Solutions · LLZ · Aug 25, 2008 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082318

## Device Facts

- **Applicant:** Ge Healthcare Dynamic Imaging Solutions
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 25, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Centricity PACS IW ™ by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

## Device Story

Centricity PACS IW is an internet-based software system for archiving, transferring, displaying, and processing medical images; receives data from DICOM-compliant modalities (CT, MR, US, NM, CR, DR, DM, DX, XA, PET). Deployed over standard TCP/IP networks on Intel Pentium-based hardware running Windows OS. Operated by trained healthcare professionals (radiologists, physicians, nurses, technicians) in clinical environments. System facilitates image management and review; does not generate original images. Provides tools for enhanced PET-CT user preferences. Supports clinical decision-making by enabling efficient access and visualization of diagnostic studies. Benefits include streamlined workflow and centralized image availability for diagnostic interpretation.

## Clinical Evidence

Reader evaluation conducted with 11 board-certified radiologists comparing enhanced PET-CT user preferences against predicate device using actual clinical cases. Study confirmed performance and safety for intended use.

## Technological Characteristics

Software-based PACS; operates on Intel Pentium-based hardware; utilizes TCP/IP networking; DICOM compliant. Supports various imaging modalities. Does not contact patients; no life-sustaining functions. Requires human intervention for image interpretation.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- IntegradWeb™ PACS System ([K031311](/device/K031311.md))
- IntegradWeb™ MRP/MIP™ ([K042313](/device/K042313.md))
- IntegradWeb™ PACS System ([K072986](/device/K072986.md))
- Xeleris 2 Processing and Review Workstation ([K051673](/device/K051673.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0823/8

# GE Healthcare Intearated IT Solutions

Image /page/0/Picture/2 description: The image shows a black and white drawing of a circular shape. The drawing is made up of many small dots, which create a textured effect. The dots are arranged in a way that suggests a swirling or spiraling pattern within the circle. The overall impression is that of a stylized or abstract representation of a circular form.

#### 510(k) Summary Section 5:

AUG 2 5 2008

- GE Healthcare Dynamic Imaging Solutions Submitter: 40 Boroline Road Allendale, NJ 07401 Jillian M. Reed Contact Person: Consultant Reed Technical Associates, LLC Date Prepared: August 12, 2008 Classification Name: Picture Archiving Communications System Proprietary Name: Centricity PACS IW ™ PACS System #K031311 IntegradWeb™ PACS System by Dynamic Imaging, Inc. Predicate Devices: #K042313 IntegradWeb™ MRP/MIP™ by Dynamic Imaging, Inc. #K072986 IntegradWeb™ PACS System by GE Healthcare Dynamic Imaging Solutions #K051673 Xeleris 2 Processing and Review Workstation by GE Medical Systems
### Device Description:

Centricity PACS IW ™ PACS System is an Internet bases software picture archiving and communications system that provides users with capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images (including digital mammograms). Centricity PACS System includes features to access and manage medical imaging studies from cat-scan (CT), magnetic radiography (MR), ultrasound (US), nuclear medicine (NM), computerized radiography (CR), digital radiography (DR), digital mammography (DM), digital x-ray (DX), x-ray angiography (XA), PET scan (PT), and other imaging modalities. Centricity PACS IW ™ PACS System is designed to be deployed over conventional TCP/IP networking infrastructure available in most healthcare organizations and utilizes commercially available computer platforms (Intel Pentium-based) and operating systems (Microsoft Windows 2000, Windows NT, and Windows 98). The system does not produce any original medical images. All images located on Centricity PACS IW ™ PACS System have been received from DICOM compliant modalities and/or systems.

### Intended Use:

Centricity FACS IW ™ PACS System by Dynamic Imaging, Inc. is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

### CONFIDENTIAL

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black and white abstract pattern. The pattern appears to be a circular design with swirling lines and dots. The overall impression is one of complexity and movement. The image is somewhat grainy and lacks sharp details.

12

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

#### Technical Characteristics:

This device is a medical device image software that is used with computer hardware in a picture archiving and communications system user environment. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention, interprets images and information being displayed and printed.

#### Substantial Equivalence:

Testing performed has shown that the Centricity PACS IW ™ PACS System incorporating enhanced PET-CT user preferences is substantially equivalent to the above referenced predicate devices.

### Discussion of Non-Clinical Testing Performed:

Thorough system verification and validation testing was performed to ensure the safe and effective use of the Centricity PACS IW ™ PACS System with enhanced PET-CT user preferences.

### Discussion of Clinical Testing Performed:

A reader evaluation, consisting of 11 board certified Radiologist was conducted where they evaluated the new enhanced PET-CT user preferences to the predicate device on actual clinical cases.

#### Conclusions:

The information provided in this special premarket notification submission has shown that the Centricity PACS IN ™ PACS System with enhanced PET-CT user preferences is substantially equivalent to the predicate device and is safe and effective for its intended use.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers, positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular fashion around the eagle symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## AUG 2 5 2008

GE HealthCare Dynamic Imaging System % Ms. Jillian M. Reed Consultant Reed Technical Associates, LLC 25 Walnut Street MONROE CT 06468

Rc: K082318

Trade/Device Name: Centricity PACS IWTM PACS System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 12, 2008 Received: August 13, 2008

Dear Ms. Reed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Nancy Brogdon

Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K082318

Device Name: Centricity PACS IW ™ PACS System

Indications for Use:

Centricity PACS IW ™ by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.

Prescription Use XXX (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

fonguithon

(Division S Division of Reproductive, Abdominal a Radiological Devices 510(k) Number

Page 1of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K082318](https://fda-staging.innolitics.com/device/K082318)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
