PRIME ECG SYSTEM (WITH ENHANCED DIAGNOSTIC ALGORITHM)
Applicant
Heartscape Technologies, Ltd.
Product Code
DPS · Cardiovascular
Decision Date
Sep 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K082312 · Sep 12, 2008
PRIME ECG SYSTEM (WITH ENHANCED DIAGNOSTIC ALGORITHM)
Heartscape Technologies, Ltd.
IDE study (G990171) clinical data; FDA Troponin set (biomarker-confirmed clinical cases)
Retrospective clinical data from an IDE study (the 'FDA Troponin set') was used to validate the sensitivity and specificity of the enhanced diagnostic algorithm compared to 12-lead ECG and previous device versions.
IDE study (G990171) / FDA Troponin set; Retrospective analysis of clinical cases
Acute Coronary Syndrome patients; Sample Size: 225 (78 True Positive, 147 False Positive); Number of Sites: Multiple institutions
12-lead ECG
Sensitivity and specificity in the detection of Myocardial Infarction (MI)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Myocardial Infarction Detection
—
—
Sensitivity: 47%, Specificity: 88%
Alpha, Beta, and Charlie training sets from Acute Coronary Syndrome patients
—
FDA Troponin set (Gold Standard)
—
Indications for Use
The PRIME ECG® with enhanced Diagnostic Algorithm is intended to be used for the recording of electrocardiographic signals. The PRIME ECG® with enhanced Diagnostic Algorithm is indicated for the recording of electrocardiographic signals on the body surface
Device Story
PRIME ECG system records body surface potentials via a single-use 80-lead electrode vest (64 anterior, 16 posterior). System includes hardware, firmware, computer, display, and printer. It transforms ECG waveforms into color-coded maps on a simulated torso (yellow-to-red for positive, green for neutral, blue for negative). An automated diagnostic algorithm analyzes ECG parameters and morphology to suggest Normal, Abnormal, or Acute MI conditions. Used in clinical settings by healthcare providers; physicians review the algorithm's recommendation and explanation window to confirm or override findings. The device aims to improve sensitivity in detecting Myocardial Infarctions compared to standard 12-lead ECGs.
Clinical Evidence
Performance evaluated using a subset of IDE study (G990171) data, specifically an 'FDA Troponin' set (n=225: 78 positive, 147 negative) where MI diagnosis was confirmed by Troponin gold standard. Comparison of the enhanced algorithm against 12-lead ECG and previous PRIME ECG versions showed improved sensitivity (47% vs 22% for 12-lead) and specificity (88% vs 95% for 12-lead). Three additional training sets (Alpha, Beta, Charlie) were used for development, with the Troponin set reserved for final validation.
Technological Characteristics
System includes a 80-lead single-use electrode vest (screen-printed on plastic substrate with conductive gel), signal acquisition hardware, computer, flat panel display, and printer. Complies with EN60601 and AAMI 11 standards. Features patient isolation, defibrillation protection, and EMC compliance. Software includes automated analysis and color-mapping of ECG potentials.
Indications for Use
Indicated for recording body surface electrocardiographic signals in patients to assist in the detection of Myocardial Infarction (MI).
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
PRIME ECG TM System with Diagnostic Algorithm (K030104)
Submission Summary (Full Text)
{0}------------------------------------------------
#### V. 510(K) SUMMARY
# 510(k) SUMMARY
# Heartscape Technologies Ltd's PRIME ECG® with enhanced Diagnostic Algorithm
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Heartscape Technologies Ltd Unit 1, 6B Balloo Drive Bangor Co Down BT19 7QY United Kingdom
Phone: 011 44 2891 857500 Facsimile: 011 44 2891 857900
Contact Person: Paul Phillips
Date Prepared: 8th August 2008
#### Name of Device
PRIME ECG® with enhanced Diagnostic Algorithm
## Common or Usual Name
PRIME ECG®
## Classification Name/Product Code/CFR Reference
Electrocardiograph, classified under 21 CFR 870.2340
## Predicate Devices
- PRIME ECG The System: manufactured by Meridian Medical Technologies Ltd . (K012414)
- PRIME ECG TM System with Diagnostic Algorithm: formerly manufactured by . Meridian Medical Technologies Ltd, now manufactured by Heartscape Technologies Ltd. (K030104)
{1}------------------------------------------------
### Intended Use / Indications for Use
The PRIME ECG® with enhanced Diagnostic Algorithm is intended to be used for the recording of electrocardiographic signals.
The PRIME ECG® with enhanced Diagnostic Algorithm is indicated for the recording of electrocardiographic signals on the body surface
## Technological Characteristics
The device consists of the following components and accessories:
#### PRIME ECG Cart .
Data recording and analysis is contained within a durable medical device that is used to record, analyze and display electrocardiographic signals from the body surface. The system features a power supply with battery back-up, signal acquisitioning and processing hardware and firmware, a computer and a software, including automated analysis software, a flat panel color display and a color printer.
The PRIME ECG System incorporates computer processing that can present conventional ECG waveforms as color images displayed on a simulated torso. Each PRIME ECG recording captures body surface potentials for analysis. The PRIME ECG System translates the segments of the ECG waveform into ranges of color based on measured values. The range from vellow to red is assigned to positive values, with deep red assigned to the highest value. Green is assigned to neutral values and shades of blue are assigned to negative values with deep blue assigned to the highest negative value. This information allows physicians to identify areas of abnormality for interrogation.
#### PRIME ECG Algorithm .
The PRIME Diagnostic Algorithm is an adjunct to the color displays of the underlying ECG waveforms to assist the physician in coming to a conclusion on the patient's cardiac status. The Diagnostic Algorithm uses multiple parameters of ECG potentials and ECG morphology to make a recommendation on whether the patient has a Normal, Abnormal or Acute MI condition. An explanation window describes to the physician the basis of the Diagnostic Algorithm conclusion. Thus, the physician is at liberty to interrogate the recommendation and agree to or override it.
#### PRIME ECG Single-Use Patient Vest .
The PRIME ECG data recording and analysis system is attached to a single-patient electrode array (vest) that is placed on the patient in two parts consisting of a 64 lead
{2}------------------------------------------------
anterior and a 16 lead posterior segment. Electrodes and signal conduction pathways are screen-printed onto a clear plastic substrate. The vest is secured to the patient with a pre-applied conductive adhesive gel. The vest is a single use device and cannot be re-used.
The device complies with the relevant parts of EN60601 and AAMI 11. In overview:
- The device has features to isolate the patient from mains AC electricity supply, even . in the event of fault conditions.
- The device is meant to be removed from contact with the patient before defibrillation. . However, it has features to protect itself and users in the event of defibrillation of the patient whilst the device is still attached to the patient.
- The device conforms to the required standards for ElectroMagnetic Compatibility . ("EMC"), both in its proof against external sources of interference and in limiting its own electromagnetic radiations.
### Performance Data
The purpose of this 510(k) submission is to change the labeling of the device to incorporate the effects of enhancements to the Diagnostic Algorithm. These enhancements have resulted in an improved sensitivity in the detection of Myocardial Infarctions (MIs) over both the predecessor version (as cleared under K030104) and also the ubiquitous 12-lead ECG type of device.
The Company's IDE study (G990171) gathered data from a number of institutions to compare the sensitivity and specificity of both the 12-lead and PRIME ECG in the detection of MIs. This data was used as the basis for the 510(k) submission for K012414, demonstrating equivalence between the 12 lead as a predicate and PRIME ECG.
The IDE study had allowed the institutions to follow their normal protocol in the treatment of suspect MI. As a result there were a number of methods of confirmation of the diagnosis, no one methodology had been mandated. A sub-set of the IDE data was generated, the so called "FDA Troponin" set, where the MI diagnosis had been confirmed by Troponin as a Gold Standard. This sub-set was used to confirm that the Diagnostic Aleorithm in the 510(k) submission for K030104 was at least as effective as the 12 lead in the detection of MI. Since the submission for K030104, the Company has added further enhancements to the Diagnostic Algorithm that improve its sensitivity to MI, whilst not significantly degrading its specificity. A table of comparison is below:
{3}------------------------------------------------
| Test | True<br>Positive<br>(n=78) | False<br>Positive<br>(n=147) | Sensitivity | Specificity | Ratio of PRIME<br>Sensitivity to<br>12-lead ECG |
|---------------------------------------------------|----------------------------|------------------------------|-------------|-------------|-------------------------------------------------|
| 12-lead ECG | 17 | 7 | 22% | 95% | not applicable |
| PRIME ECG, physician<br>interpretation (K012414) | 26 | 10 | 33% | 93% | 1.5 |
| PRIME Algorithm<br>(K030104) | 29 | 14 | 37% | 90% | 1.7 |
| Enhanced PRIME<br>Algorithm<br>(this application) | 37 | 17 | 47% | 88% | 2.1 |
Table 1: Comparative Device Performance among patients with Biomarker confirmed MI
The Company has created 3 training sets of cases (called Alpha, Beta and Charlie) for the development of the Diagnostic Algorithm. These cases have been drawn from Acute Coronary Syndrome patients attending a UK hospital and have no overlap with the FDA Troponin set. The Alpha, Beta and Charlie sets are used for the testing/training of possible enhancements to the Diagnostic Algorithm. The Troponin set is reserved solely for the final validation of MI detection performance.
#### Substantial Equivalence
The version of PRIME ECG, with enhanced Diagnostic Algorithm, that is the subject of this submission is as safe and effective as the predicate versions of PRIME ECG. The device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the PRIME ECG with enhance Diagnostic Algorithm and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the PRIME ECG with enhanced Diagnostic Algorithm is as safe and effective as the predicate PRIME ECG with Diagnostic Algorithm and the PRIME ECG. Thus, the PRIME ECG with enhanced Diagnostic Algorithm is substantially equivalent.
#### Labeling Change
The main purpose of this 510(k) application is the desire to have a change to the labeling of the device that informs the user of the improved comparative performance of the device over the 12-lead EKG (the predicate of the original PRIME ECG) in the detection of ST Elevation Myocardial Infarction.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three overlapping wings, representing the department's commitment to health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
.
SEP 1 2 2008
Heartscape Technologies Ltd c/o Mr. Paul Phillips Director Unit 1, 6B Balloo Drive Bangor Co Down BT19 7QY United Kingdom
Re: K082312
Trade/Device Name: PRIME ECG ® Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: August 8, 2008 Received: August 13, 2008
Dear Mr. Phillips:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Paul Phillips
Please he advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,
Sincerely yours,
Duma R. Va. Vamer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
.
.
. 1
Enclosure
- 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 1
{6}------------------------------------------------
#### IV. INDICATIONS FOR USE STATEMENT
Indications for Use Statement
510(k) Number (if known): K082 31 2 ---------
Device Name: PRIME ECG®
Indications for Use:
The PRIME ECG® with enhanced Diagnostic Algorithm system is intended for the recording of electrocardiographic signals from the body surface
メ Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Mines
vision Sign-Off) ivision of Cardiovascular Devices
K082312 510(k) Number
Page ***_ of _***
Page 15 of 57
510k_prime1_enhanced da_application-080808.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.