← Product Code [FRO](/productcode/FRO) · K082296

# COPA AMD (K082296)

_Covidien Lp, Formerly Registered AS Tyco Healthcar · FRO · Feb 26, 2009 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082296

## Device Facts

- **Applicant:** Covidien Lp, Formerly Registered AS Tyco Healthcar
- **Product Code:** [FRO](/productcode/FRO.md)
- **Decision Date:** Feb 26, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

COPA AMD dressings are indicated for use in management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, lacerations, 1st and 2nd degree burns, dermatologic disorders, other wounds inflicted by trauma and, as a secondary dressing or cover dressing for packed wounds.

## Device Story

COPA AMD is a hydrophilic polyurethane foam dressing impregnated with Polyhexamethylene Biguanide Hydrochloride (PHMB). PHMB acts as an antimicrobial agent to protect the dressing from bacterial penetration and colonization. The device is applied topically to various wound types to manage exudate and provide an antimicrobial barrier. It is intended for use by healthcare professionals or patients in clinical or home settings. The device functions as a passive wound dressing; it does not require active operation or electronic processing. The primary benefit is the management of wound moisture and the reduction of bacterial colonization within the dressing.

## Clinical Evidence

Bench testing only; in-vitro testing was submitted to support the antimicrobial efficacy of the PHMB-impregnated foam.

## Technological Characteristics

Hydrophilic polyurethane foam impregnated with Polyhexamethylene Biguanide Hydrochloride (PHMB). Sterilization method: Gamma irradiation.

## Predicate Devices

- COPA AMD antimicrobial wound dressings ([K082296](/device/K082296.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "KENDALL" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The word is centered and takes up most of the frame. The font is simple and easy to read.

1

KENDALL 15 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 • 5009 261-80

Special 510(k) Premarket Notification COPA AMD Antimicrobial Wound Dressing

# Section B – 510(K) Summary

| Date Summary Was Prepared: | August 8, 2008 |
|----------------------------|----------------|
|----------------------------|----------------|

| Submitter's Information: | Kendall<br>a Division of Tyco Healthcare Group LP<br>15 Hampshire Street<br>Mansfield, MA 02048<br>Phone: 508-261-8000<br>Fax: 508-261-6644 | FEB 2 6 2009 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact:                 | James Welsh<br>VP, Regulatory Affairs<br>Kendall<br>a Division of Tyco Healthcare Group LP<br>Telephone: 508-261-8532<br>Fax: 508-261-8461  |              |
| Device Trade Name:       | COPA AMD Antimicrobial Wound Dressing                                                                                                       |              |
| Device Common            |                                                                                                                                             |              |

Name:

Wound Dressing, Antimicrobial

Classification Panel: General and Plastic Surgery

### Legally Marketed Devices To Which Substantial Equivalence Is Claimed:

The COPA AMD antimicrobial wound dressing is substantially equivalent to the existing COPA AMD antimicrobial wound dressings in intended use, materials, physical characteristics, and performance characteristics. The modification attributed to the predicate device is a change in sterilization method from ETO to Gamma irradiation

K08 2296
page 1 of 2

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K082296", and the text at the bottom reads "page 2 of 2". The handwriting is in black ink and appears to be cursive.

15 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 · (508) 261-8000

Special 510(k) Premarket Notification COPA AMD Antimicrobial Wound Dressing

## Section B -- 510(K) Summary (continued)

#### Device Description:

COPA AMD is a hydrophilic polyurethane foam that is impregnated with Polyhexamethylene Biguanide Hydrochloride (PHMB), an antimicrobial agent that protects the dressing from bacterial penetration and colonization.

#### Intended Use:

Kem

COPA AMD dressings are indicated for use in management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, lacerations, 1st and 200 degree burns, dermatologic disorders, other wounds inflicted by trauma and, as a secondary dressing or cover dressing for packed wounds.

Performance Data: Performance data submitted in support of this 510k included in-vitro testing.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an image of an eagle.

Public Health Service ·

Covidien LP, Formerly Registered as Tyco Healthcare % Mr. James Welsh Vice President, Regulatory Affairs 15 Hampshire Street Mansfield, Massachusetts 02048

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Re: K082296

Trade/Device Name: COPA AMD antimicrobial wound dressing Regulatory Class: Unclassified Product Code: FRO Dated: February 12, 2009 Received: February 13, 2009

FEB 2 6 2009

Dear Mr. Welsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 - Mr. James Welsh

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

L. Arjona. fo

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

510(k) number:

K082296

Device Name:

COPA AMD antimicrobial wound dressing

Indications for Use:

COPA AMD dressings are indicated for use in management of post-surgical incisions, pressure sores, venous stasis ulcers, diabetic ulcers, donor sites, abrasions, lacerations, 1*-and-200degree burns, dermatologic disorders, other wounds inflicted by trauma and, as a secondary dressing or cover dressing for packed wounds.

Prescription Use (21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Please Do Not Write Below This Line - Continue On Another Page If Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Krone 2/26/2004

(Division Size-Oils Division of General, Restorative, and Neurological Devices

**510(k) Number** K062296

---

**Source:** [https://fda-staging.innolitics.com/device/K082296](https://fda-staging.innolitics.com/device/K082296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com