← Product Code [MHX](/productcode/MHX) · K082280

# SURESIGNS VS2 VITAL SIGNS MONITOR, SURESIGNS VM1 PATIENT MONITOR, SURESIGNS VS3 CITAL SIGNS MONITOR (K082280)

_Philips Medical Systems North America Co. · MHX · Oct 8, 2008 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082280

## Device Facts

- **Applicant:** Philips Medical Systems North America Co.
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Oct 8, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The SureSigns VS2 Vital Signs monitor and the SureSigns VM1 are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitors may be used in transport situations within a healthcare facility.

## Device Story

Philips SureSigns VM1 and VS2 are multi-parameter patient monitors; VS3 is a modified version with an optional wireless bridge. Devices monitor physiological parameters (e.g., blood pressure, cardiac, oximetry, temperature, gas analysis) in adults, pediatrics, and neonates. Used by healthcare professionals in clinical settings and during intra-facility transport. Inputs include patient-connected sensors/electrodes; devices process signals to provide real-time monitoring, recording, and alarms. Output displayed for clinician review to support clinical decision-making. Wireless bridge on VS3 enables data transmission to electronic medical record systems. Benefits include continuous patient status awareness and timely alarm notification.

## Clinical Evidence

Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Pass/Fail criteria were established against predicate device specifications; results demonstrated substantial equivalence.

## Technological Characteristics

Multi-parameter patient monitors. Materials: previously cleared medical-grade materials. Energy: external power supply and battery operation. Connectivity: optional wireless bridge interface for VS3. Parameters: blood pressure, cardiac, oximetry, temperature, gas analysis. Form factor: bedside/transport monitors.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- SureSigns VM Series Patient Monitors ([K052707](/device/K052707.md))
- SureSigns VS3 Patient Monitor ([K080495](/device/K080495.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K082280
pi/2

## OCT 0 8 2008

## 510K Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).

- The submitter of this pre-market notification is: 1.
Mary Kruitwagen Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States

Tel: 978-659-4932 Fax: 978-685-5624 Email: mary kruitwagen@philips.com

This summary was prepared on July 31, 2008.

- The names of the subject devices are the Philips SureSigns VM1 Patient Monitors and 2. SureSigns VS2 Vital Signs Monitor.
- ന The trade names of the devices are the SureSigns VM1 Patient Monitors (VM1) and the SureSigns VS2 Vital Signs Monitor.
- The common usual name is multi-parameter patient monitor 4.
- 5. The Classification names are as follows:

| Device Panel                            | Classification | ProCode | Description                                                                 | Applicable<br>Subject<br>Devices |
|-----------------------------------------|----------------|---------|-----------------------------------------------------------------------------|----------------------------------|
| Circulatory System<br>Devices           | 870.1025, II   | MHX     | Monitor, Physiological,<br>Patient (with arrhythmia<br>detection or alarms) | VS2, VM1                         |
|                                         | 870.1110, II   | DSJ     | Alarm, Blood Pressure                                                       | VS2                              |
|                                         | 870.1110, II   | DSK     | Computer, Blood Pressure                                                    | VS2                              |
|                                         | 870.1130, II   | DXN     | System, Measurement,<br>Blood Pressure, Non-<br>Invasive                    | VS2                              |
|                                         | 870.2300, II   | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate<br>Alarm                 | VS2, VM1                         |
|                                         | 870.2700, II   | DQA     | Oximeter                                                                    | VS2, VM1                         |
|                                         | 870.2900, I    | DSA     | Cable, Transducer and<br>Electrode, incl. Patient<br>connector              | VM1                              |
| General Hospital and<br>Personal Use    | 880.2910, II   | FLL     | Thermometer, Electronic,<br>Clinical                                        | VS2                              |
| Anesthesiology &<br>Respiratory Therapy | 868.1400, II   | CCK     | Analyzer, Gas,                                                              | VM1                              |

6. The modified devices are substantially equivalent to previously cleared Philips device, SureSigns VM Series Patient Monitors and SureSigns VS3 Patient Monitor marketed pursuant to K052707 and K080495.

{1}------------------------------------------------

- 7. The modifications are as follows:
	- . Introduction of the VM1 Patient Monitor (a subset of the predicate VM Series devices)
	- Introduction of the VS2 Vital Signs monitor (a subset of the predicate device VS3) .
	- . Modification of the predicate VS3 device to include the optional wireless bridge interface to transmit patient data records to an electronic medical record system.
- 8. The subject devices have the same intended use as the legally marketed predicate device. The SureSigns VS2 Vital Signs monitor and the SureSigns VM1 are for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitors may be used in transport situations within a healthcare facility. The addition of the wireless functionality to the predicate VS3 does not change the intended use or indications for use.
- 9. The subject devices have the same fundamental technological characteristics as the legally marketed predicate devices. The subject devices use the same design as the predict device. The composition of the VS2 and the VM1 materials are different than the predicate device, however are made of a material previously used in other medical devices. The chemical composition of the subject devices has changed but is the same as material used in other predicate devices. The energy source of the subject device the VS2 is an external power supply, while the VM1 will use a power supply that is the same as the predicate device. The VS2 and the VM1 both can run on battery power with batteries similar to the predicate devices.
- 10. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject devices with respect to the predicates. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device, the specifications of the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM1 Patient Monitors and SureSigns VS2 Vital Signs monitor and the modification to the SureSigns VS3 Vital Signs Monitor meet all reliability requirements and performance claims and supports a determination of substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

OCT 0 8 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Philips Medical Systems c/o Ms. Mary Kruitwagen Regulatory Affairs Specialist 3000 Minuteman Road Andover, MA 01810

Rc: K082280

Trade/Device Name: SureSigns VS2 (reference numbers: 863079,863080; 863081, 863082); SureSigns VM1 (reference number: 863078) SureSigns VS3 (reference number: 863069; 863070, 863071, 863072, 863073, 863074) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) Regulatory Class: II Product Code: MHX Dated: August 8, 2008 Received: August 11, 2008

Dear Ms. Kruitwagen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

{3}------------------------------------------------

Page 2 - Ms. Mary Kruitwagen

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

7 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

.

510 (k) Number (if known): __________________________________________________________________________________________________________________________________________________

Device Name: SureSigns VS2 (reference numbers: 863079, 863080, 863081, 863082) SureSigns VM1 (reference number: 863078) SureSigns VS3 (reference numbers: 863069, 863070, 863071, 863072, 863073, 863074)

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

| Prescription Use:            | YES | AND/OR over-the-counter Use: | NO |
|------------------------------|-----|------------------------------|----|
| (Part 21 CFFR 801 Subpart D) |     | (21 CFR 807 Subpart C)       |    |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH Office of Device Evaluation (ODE)

  

  

(Division Sign-Off)
Division of Cardiovascular Devices

| 510(k) Number | K082280 |
|---------------|---------|
| Page          | of ___  |

---

**Source:** [https://fda-staging.innolitics.com/device/K082280](https://fda-staging.innolitics.com/device/K082280)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
