ZENO AI ECO DISC

K082257 · Wieland Dental + Technik GmbH & Co. KG · EIH · Sep 18, 2008 · Dental

Device Facts

Record IDK082257
Device NameZENO AI ECO DISC
ApplicantWieland Dental + Technik GmbH & Co. KG
Product CodeEIH · Dental
Decision DateSep 18, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

ZENO® Al cco Discs are milling blanks from which single copings and primary components for anteriors and premolars can be made. These are intended for use as ceramic frameworks for dental prosthetics ..

Device Story

ZENO Al eco Discs are pure aluminium oxide milling blanks used by dental technicians to fabricate ceramic frameworks for dental prosthetics. The process involves scanning a patient model, designing the restoration using CAD software, and milling the blank via CAM technology within the ZENO Tec system. Following hard sintering, the framework is veneered with ceramic. The device is used in a dental laboratory setting. The output is a custom-fitted dental coping or primary component, which aids the clinician in restoring patient dentition, providing a cost- and time-saving solution for high-aesthetic dental prosthetics.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Material: Pure aluminium oxide. Form factor: Milling blanks. Processing: CAD/CAM machining followed by hard sintering. Compatibility: ZENO Tec system. Intended for veneering with dental ceramics.

Indications for Use

Indicated for professional dental technicians to fabricate ceramic frameworks (single copings and primary components) for anterior and premolar dental prosthetics for individual patients.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082257 Image /page/0/Picture/1 description: The image shows the word "WIELAND" in a bold, sans-serif font. The letters are black and appear to have a slightly rough or textured edge, giving them a somewhat distressed look. The word is presented in all capital letters and is horizontally oriented. WIELAND Dental | Technik GmbH & Co. KG Schwenninger Straße 13 75179 Pforzheim, Germany Postlad: 10 20 40 /5120 Pforzheim, Germany Fon +49 72 31/ 37 05 - 0 Fax +49 77 317 35 79 59 > : : -- · SEP 1 8 2008 ### 7. 510 (k) Summary Wieland Dental + Technik GmbH & Co. KG Submitter of 510(k): Schwenninger Str. 13 D-75179 Pforzheim Germany Phone: +49-7231-3705-0 | Contact person: | Dr. Gerhard Polzer | |-----------------|----------------------------------| | Phone: | +49-7231-3705-219 | | Fax: | +49-7231-357959 | | e-mail: | gerhard.polzer@wieland-dental.de | Date of Summary: 2008-07-03 Trade name: ZENO Al eco Disc Classification name: Powder, Porcelain Product code: ElH C.D.R section: 872.6660 Classification: Class II Legally marketed equivalent device: inCoris AL 510(k) number: K062506 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "WIELAND" in a bold, sans-serif font. The letters are black and have a slightly rough or textured appearance, as if they were printed with a slightly worn-out ink or on a textured surface. The background is plain white, which makes the word stand out prominently. WIFLAND Dental + Technik Gmbll & Co. KG Schwenninger Straße 13 75179 Plorzheim, Germany Postfach 10 20 40 75120 Pforzheim, Germany Fort +49 /2 31/ 37 05 - 0 Fax +49 /2 31/ 35 79 59 # 510 (k) Summary # Device description ZENO Al eco Discs are milling blanks composed of pure aluminium oxide. They are intended to be used by professional dental technicians for making single copings and primary components for anteriors and premolars to apply them as ceramic frameworks for dental prosthetics for the sole use of particular patients. ZENO Al eco Discs can be machined in all machines of the ZENO Tec system. The manufacturing process of this ceramic framework consists of different steps. At first the model has to be scanned. In the next step, the restoration has to be designed virtually with the help of the CAD technology. Thereafter, the realization of this design has to be carried out by the CAM technology. In a final step after hard sintering of the ZENO Al eco, the framework can be veneered with a suitable veneering ceramic. ## Recommended application With the introduction of the ZENO AI eco Discs, Wieland Dental+Technik offers to the customer for the ZENO TEC System the possibility to produce high aesthetic single copings and primary components on the base of Aluminium oxide. This kind of restoration is cost- and time saving and the restorations have excellent properties. # Comparison with the predicate device - ZENO Al eco Discs are substantially equivalent to the dental device inCoris AL. Both devices based on Aluminium oxide and have similar indications for use and comparable physical, biological, and chemical properties. Due to the excellent material properties of the ZENO AI eco Discs, the restorations possess a high-level safety and effectiveness. ZENO AI eco Discs therefore are as safe, as effective, and performs as well as or better than the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, graphic representation of the department. SEP 1 8 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Dr. Gerhard Polzer Director, Regulatory Affairs Wieland Dental + Technik GmbH & Company KG Schwenninger Straße 13 75179 Pforzheim, Germany Re: K082257 Trade/Device Name: ZENO AI cco Disc Regulation Number: 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: August 4, 2008 Received: August 8, 2008 Dear Dr. Polzer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Dr. Polzcr Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "WIELAND" in large, bold, sans-serif font. The letters are white with a black speckled pattern, giving them a textured appearance. The word is presented horizontally and fills most of the frame. WILLAND Dental + Technik GmbH & Co. KG Schwenninger Straße 13 15179 Pforzheim, Germany Postfach 10 20 40 75120 Pforzheim, Germany Fon +49 72 31/ 37 05 - 0 Fax +49 72 311 35 79 59 #### 12.8. Indication for use statement Indications for Use 510(k) Number (if known): KO82257 Device Name: ZENO Al eco Disc Indications for Use: ZENO® Al cco Discs are milling blanks from which single copings and primary components for anteriors and premolars can be made. These are intended for use as ceramic frameworks for dental prosthetics .. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) MSR (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number. K01 Page 1 of 1 (Posted November 13, 2003)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...