For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status. This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values; fat%, fat weight, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
Device Story
Bodystat 1500 is an impedance plethysmograph using bio-electrical impedance analysis (BIA). Device applies 50 kHz electrical current via electrodes on subject; measures voltage drop to determine tissue impedance. Impedance correlates with fluid content; water low in fat tissue (high impedance), high in lean tissue (high impedance). Device uses proprietary regression equations to convert impedance and subject demographics (age, gender, weight, height, waist/hip circumference) into body composition estimates: fat%, fat weight, lean weight, total body water, basal metabolic rate, BMI, and waist/hip ratio. Used for health assessment; optional Body Manager and Wellness software programs track changes over time or provide educational graphical reports. Intended for use on healthy adults.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on hardware and software modifications to a previously cleared predicate device using the same operating principle.
Technological Characteristics
Impedance plethysmograph; BIA technology; 50 kHz signal source; electrodes for body contact; calculates body composition via proprietary regression equations; includes optional software for data management and reporting.
Indications for Use
Indicated for non-invasive bio-impedance analysis (BIA) to determine body composition status in normal healthy human adults.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Bodystat 1500 Body Composition Monitoring Unit (K951097)
Submission Summary (Full Text)
{0}------------------------------------------------
# PAGE 1 OF 2
## NOV 2 5 2008
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
510(k) Number : K082253
#### Submitter: A.
Bodystat Ltd P O Box 50 Douglas Isle of Man IM99 1DQ BRITISH ISLES
> Phone: +44-1624-629571 444-1624-611544 Fax:
Contact: IJ Meeuwsen
Date Prepared: October 318 , 2008
## B. Device Names:
Classification name Impedance Plethysmograph Common/usual name Body Composition Monitor Proprietary name Bodystat 1500 Body Composition Monitoring Unit
#### C. Predicate Device:
The modified BODYSTAT®1500 Body Composition Monitoring Unit is substantially equivalent to previously cleared device, K951097, Bodystat 1500 Body Composition Monitoring Unit.
The BODYSTAT® 1500 Body Composition Monitoring Unit device is a hardware and software modification of the previously cleared device.
#### D. Device Description:
The Bodystat 1500 Body Composition Monitoring Unit is an impedance plethysmograph that uses bio-electrical impedance analysis (BIA) to measure the impedance of the flow of an electrical current through the body. The impedance of tissue is proportional to the amount of fluid in the tissue; water is low in fat tissues, thus fat tissue has a high impedance, and high in lean tissues, thus lean tissue has a high impedance.
In practice, a small constant current is passed between electrodes spanning the body and the voltage drop between electrodes provides a measure of impedance. Prediction
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K082253 PAGE 1 OF 2" in a handwritten style. The text is arranged in two lines, with the first line containing the alphanumeric code "K082253". The second line indicates that it is page 1 of a total of 2 pages. The handwriting appears to be clear and legible.
equations, previously generated by correlating impedance measures against an independent estimate of TBW (total body water), may be used subsequently to convert a measured impedance to a corresponding estimate of TB W. Lean body mass is then calculated from this estimate using an assumed hydration fraction for lean tissue; Bodystat uses a proprietary regression equation for this calculation. Fat mass is calculated as the difference between body weight and lean body mass.
#### E. Intended Use:
For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status.
This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values; fat%, fat weight, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
#### OPTIONAL ACCESSORIES: Bodystat (1) BODY MANAGER and (2) WELLNESS SOFTWARE PROGRAMS.
The Body Manager Program is ideal when a client returns to be tested in order to track their changes in body composition over a period of time.
The Wellness Program is ideal for first time assessments on subjects. The program's graphical presentations provide information on each of the specific measurements in an educational format.
#### F. Comparison with the Predicate Device:
The modified BODYSTAT®1500 Body Composition Monitoring Unit is a hardware and software modification of the previously cleared Bodystat 1500. The modified Bodystat 1500 and the previously cleared Bodystat 1500 have the same intended use and use the same operating principle.
Based on the data and information presented here, the modified BODYSTAT® 1500 Body Composition Monitoring Unit is substantially equivalent to the previously cleared Bodystat 1500 device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, arranged in a cascading pattern.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 25 2008
Mr. I. J. Meeuwsen President BODYSTAT® LIMITED P.O. Box 50 Douglas, Isle of Man IM99 1DQ BRITISH ISLES
K082253 Re:
> Trade/Device Name: BODYSTAT® 1500 Body Composition Monitoring Unit Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: November 14, 2008 Received: November 19, 2008
Dear Mr. Meeuwsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Joppu M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## ATTACHMENT 4
## Indications for Use
510(k) Number : K082253
## Device Name: BODYSTAT®1500 Body Composition Monitoring Unit
Indications for Use:
For the purposes of performing a non-invasive bio-impedance analysis (BIA) measurements on normal healthy human adults to determine their Body Composition status.
This device measures bio-electrical impedance when a 50 kHz signal is applied at specific locations on the subject. From this and the subject's age, gender, weight, height, waist and hip circumferences (optional), the device calculates the following estimated values: fat%, fat weight, lean%, lean weight, total body water%, total body water, dry lean weight, basal metabolic rate, basal metabolic rate/body weight ratio, average requirement (EAR), and actual body mass index and waist/hip ratio.
OPTIONAL ACCESSORIES: Bodystat (1) BODY MANAGER and (2) WELLNESS SOFTWARE PROGRAMS.
The Body Manager Program is ideal when a client returns to be tested in order to track their changes in body composition over a period of time.
The Wellness Program is ideal for first time assessments on subjects. The program's graphical presentations provide information on each of the specific measurements in an educational format.
Prescription Use (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
Helent lemen
Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.