Static Vessel Analyzer (SVA) with VesselMap2 is intended to capture, display, store, and manipulate images of the eye, especially the retina area, as well as surrounding areas, to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. Specifically, the VesselMap2 software is intended to be used for semiautomated measurement and calculation of the retinal artery/vein diameter ratio.
Device Story
Static Vessel Analyzer (SVA) captures, displays, stores, and manipulates retinal images; utilizes fundus camera and digital image sensor. SVA software comprises VisualIS (imaging management) and VesselMap2 (analysis add-on). User selects veins and arteries on images; software estimates vessel diameters and calculates arterial-venous ratio. Used in clinical settings by healthcare providers to aid in diagnosis or monitoring of eye diseases. Semi-automated measurement improves repeatability and reduces reader-to-reader bias compared to manual methods. Output assists clinicians in assessing retinal vascular health.
Clinical Evidence
Method comparison study (bench testing) compared VesselMap2 semi-automated analysis to manual methods. Results demonstrated agreement between methods, smaller intra- and intergrader variability for the semi-automated method, and elimination of systematic reader-to-reader bias observed in manual analysis. Appropriate intravisit and intervisit reproducibility confirmed.
Technological Characteristics
Fundus imaging system; includes digital image sensor. Software components: VisualIS (image management) and VesselMap2 (semi-automated analysis). Connectivity: standalone or integrated with fundus camera. Software algorithm: semi-automated vessel diameter estimation and ratio calculation.
Indications for Use
Indicated for patients requiring diagnostic imaging or monitoring of eye and retinal diseases. Used for semiautomated measurement and calculation of retinal artery/vein diameter ratios.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
# JUN 1 6 2009
## 510(k) SUMMARY
## Imedos GmbH's Static Vessel Analvzer
· 510(k) Number: K082196
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
IMEDOS GmbH Am Nasstal 4 D-07751 Jena GERMANY Tel .: +49-3641-6396-0 Fax .: +49-3641-6396-12
Date Prepared: June 12, 2009
Contact Person: Dr. Hans Stromever Stromeyer & Partners Consulting 317 Shore Breeze Lane League City. TX 77573 Phone: +1.281.467.3424 Email: hans@stromeyerconsulting.com
Trade/Device Name: Static Vessel Analyzer (SVA) Classification Name: Opthalmic camera Regulation Number: 21 CFR 886.1120 Product Code: HKI, Class II
Classification Name: Medical image communications device. Regulation Number: 21 CFR 892.2010 Product Code: NFF, Class I 510(k) exempt
Classification Name: Medical image storage device Regulation Number: 21 CFR 892.2020 Product Code: NFG, Class I 510(k) exempt
#### Indications for Use:
Static Vessel Analyzer (SVA) with VesselMap2 is intended to capture, display, store, and manipulate images of the eye, especially the retina area, as well as surrounding areas, to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. Specifically, the VesselMap2 software is intended to be used for semiautomated measurement and calculation of the retinal artery/vein diameter ratio.
\\\DC - 033404/000001 - 2909663 v1
{1}------------------------------------------------
## Technological Characteristics:
The IMEDOS SVA unit is designed as a complete Fundus Imaging system or the software can be sold separately as a standalone product. The SVA software consist of two components - VisualIS, an imaging software for capture, display, storage and manipulation of images of the eye and VesselMap2, an add-on for enhanced analysis of retina images.
VisualIS captures the images which are provided by the fundus camera and the connected digital image sensor. Together with an image set the data of the patient are recorded and stored. The complete examination can be stored and opened for follow-up examination purposes.
The add-on VesselMap2 offers a semi-automated measurement for Arterial -- Venous Ratio of retinal vessels. The software allows the user to select veins and arteries. Vessel diameters are estimated by the software and the ratio of the analyzed arteries to veins is calculated. The image is not altered in any way during this calculation.
The manual and semi-automatic calculation were compared and verified on representative retina images and the results show improved repeatability of the semi-automated calculations based on the uncertainty and precision of manual vessel diameter determination based on the user's subjective perception of vessel edges.
## Performance Data
Imedos completed a comprehensive method comparison study assessing the VesselMap software function compared to an accepted, manual method for determining arterial-venous ratios. This study assessed agreement, reproducibility, and repeatability. The completed study demonstrated:
- 1. Agreement between manual and semi-automatic (VesselMap2) retinal vessel analysis, without any indication of statistically significant differences between methods;
- 2. Smaller intragrader variability) and intergrader variability (reproducibility) for the semi-automatic method (VesselMap2) than for the manual method;
- 3. In the manual method, a systematic bias between the two graders exists, which is not the case in the semi-automatic method. Thus, the comparison study demonstrated that the VesselMap software does not introduce reader-to-reader bias;
- 4. Appropriate intravisit and intervisit reproducibility for the semi-automatic method. The visit-tovisit variability (whether for manual or semi-automated analysis) was shown to be much higher than the variability in results associated with the device.
## Substantial Equivalence
The Imedos SVA System has the same intended use, principles of operation and technological characteristics as the Carl Zeiss FF450plus VISUPAC System (K011877). The minor differences between the Carl Zeiss FF450plus VISUPAC System and the Imedos SVA system do not raise new questions of safety or effectiveness. Performance data demonstrate that the SVA System is as safe and effective its predicates. Thus, the SVA System is substantially equivalent.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem, with three thick, curved lines forming the body and wings of the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Imedos GmbH c/o Jonathan S. Kahan Hogan & Hartson, L.L.P. 555 Thirteenth Street, NW Washington, DC 20004
JUN 1 6 2009
Re: K082196
Trade Name: Static Vessel Analyzer (SVA) Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI, NFF, NFG Dated: May 22, 2009 Received: May 26, 2009
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number: K082196
Device Name: Static Vessel Analyzer (SVA)
Indications For Use:
Static Vessel Analyzer (SVA) with VesselMap2 is intended to capture, display, store, and manipulate images of the eye, especially the retina area, as well as surrounding areas, to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. Specifically, the VesselMap2 software is intended to be used for semiautomated measurement and calculation of the retinal artery/vein diameter ratio.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off) vision of Ophthalmic and Ear, ose and Throat Devices
510(k) Number K082196
\\\DC - 033404/000001 - 2909663 v1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.