Q-STRESS AND HEARTSTRIDE, MODELS QSTRESS/Q40 AND H4/H4S
Applicant
Cardiac Science Corporation
Product Code
DPS · Cardiovascular
Decision Date
Aug 26, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Indications for Use
The device is intended to acquire, process, record, archive, analyze, and output electrocardiographic data during physiologic stress testing. The device may interface with external devices, including a treadmill or ergometer for dynamic exercise evaluation, non-invasive blood pressure equipment, and computer communications equipment. The device is intended for use in a clinical setting by trained personnel under the supervision of a licensed physician. The device is intended for use in adult populations, typically symptomatic. The device is not intended to be used as a vital signs physiological monitor.
Device Story
Diagnostic stress testing system (Q-Stress/HeartStride) acquires ECG data via patient cable and preamplifier; processes signals for ST analysis and ventricular ectopic beat detection; controls treadmill or ergometer; generates/archives stress reports. Used in clinical settings by trained personnel under physician supervision. HeartStride available as software-only product for customer-supplied PC hardware. Output includes ECG displays and printed reports. Facilitates dynamic exercise evaluation; aids clinical decision-making by providing diagnostic cardiac data. Benefits patients through standardized stress testing and cardiac monitoring.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
System consists of patient cable, preamplifier, PC, display, mouse, printer, keyboard, and isolation power supply. Supports treadmill/ergometer and non-invasive blood pressure interfaces. Software-based ECG processing and analysis. HeartStride supports software-only deployment on customer-provided hardware meeting manufacturer specifications.
Indications for Use
Indicated for adult populations, typically symptomatic, undergoing physiologic stress testing.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Cardiac Science Corporation Bothell, WA Special 510K July 30, 2008
# AUG 2 6 2008
## 11. 510(k) Summary
K082173
#### Name of Device A.
Trade name: Q-Stress and HeartStride Common name: Stress system Classification name: Electrocardiograph
#### B. Predicate Devices
| Device Name | Premarket Notification |
|-------------|------------------------|
| Q-Stress | K001492 |
#### C. Device description
Q-Stress and HeartStride are diagnostic devices capable of ECG monitoring; ST analysis and ventricular ectopic beat detection; generation, review, and storage of stress reports; and treadmill or ergometer control. These devices consist of a patient cable, preamplifier, PC, display, mouse, printer, keyboard, and isolation power supply. Approved serial devices such as non-invasive blood pressure measurement may be supported by these devices. Electrocardiographic data is obtained by the preamplifier and sent to the CPU for processing. The user may generate reports for display or may opt to print results via a printer.
The differences between the Q-Stress and HeartStride are as follows:
- 1. System Specifications/ Firmware
- a. HeartStride system specifications are the same as the Q-Stress. HeartStride could also be available as a software only product. The PC, monitor, mouse, and printer may be supplied by the customer, however, they must meet Cardiac Science Corporation (CSC) specifications.
- b. Q-Stress reports can be printed using chart recorder or laser printer. However, HeartStride reports can only be printed using laser printer.
1 3
Page of 40
Confidential
{1}------------------------------------------------
- c. Remote keypad and external hard drive are available with Q-Stress, but not with HeartStride.
- 2. Software
- a. HeartStride has a different toolbar and provides a testing wizard.
- b. Different options are available with the Q-Stress and the HeartStride device.
Q-Stress:
- Network
- Q-Exchange (includes Q-Exchange+)
- Full Disclosure -
- Freeze Frame
- Risk Scoring
- Re-analysis
- Ramped Protocol Generator
- Custom Reports
- SpO2
- Interface to TMX425
HeartStride:
- Network
- -Q-Exchange (does not include Q-Exchange+)
- QRS sync and Analog outs
- Interface to TMX425
#### D. Intended use
The device is intended to acquire, process, record, archive, analyze, and output electrocardiographic data during physiologic stress testing. The device may interface with external devices, including a treadmill or ergometer for dynamic exercise evaluation, non-invasive blood pressure equipment, and computer communications equipment. The device is intended for use in a clinical setting by trained personnel under the supervision of a licensed physician. The device is intended for use in adult populations, typically symptomatic. The device is not intended to be used as a vital signs physiological monitor.
Confidential
{2}------------------------------------------------
### E. Summary
The intended use, indication for use, and principle of operation are substantially equivalent to the predicate devices. We believe that the modifications to the device do not raise any new questions of safety and/or effectiveness.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 6 2008
Cardiac Sciences Corporation c/o Ms. Beverly Magrane Senior Manager RA/RC 3303 Monte Villa Parkway Bothell, WA 98201-8969
Re: K082173
Trade/Device Name: Q-Stress and HeartStride Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: July 30, 2008 Received: August 1, 2008
Dear Ms. Magrane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);
{4}------------------------------------------------
Page 2 - Ms. Beverly Magrane
labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillaree at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
Cardiac Science Corporation Bothell, WA Special 510K July 30, 2008
# l. Indications for Use Statement
K082173
510(k) Number: TBD
Device Name: Q-Stress and HeartStride
Indications for Use:
The device is intended to acquire, process, record, archive, analyze, and output electrocardiographic data during physiologic stress testing. The device may interface with external devices, including a treadmill or ergometer for dynamic exercise evaluation, non-invasive blood pressure equipment, and computer communications equipment. The device is intended for use in a clinical setting by trained personnel under the supervision of a licensed physician. The device is intended for adult populations, typically symptomatic. The device is not intended to be used as a vital signs physiological monitor.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use X (Per 21 CFR 801.109)
| | Over-The Counter Use |
|-----------------|----------------------|
| - Real Property | |
ion of Cardiovascular Devices
Confidential
KORZI73 510(k) Number
OR
Page 3 of 40
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.