← Product Code [EBI](/productcode/EBI) · K082153

# MAJOR.REPAIR (K082153)

_Major Prodotti Dentari S.P.A. · EBI · Nov 25, 2008 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082153

## Device Facts

- **Applicant:** Major Prodotti Dentari S.P.A.
- **Product Code:** [EBI](/productcode/EBI.md)
- **Decision Date:** Nov 25, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K082153 · Nov 25, 2008 | MAJOR.REPAIR | Major Prodotti Dentari S.P.A. | Historical manufacturing records; General dental literature; Post-market clinical experience/monitoring | The sponsor uses 50 years of clinical history, extensive literature reviews, and internal manufacturing/adverse event tracking data to support the safety and effectiveness of the device, arguing that the material is well-established and does not require new clinical trials. | Historical clinical experience; Literature review; Post-market safety monitoring |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Historical Clinical Experience and Literature Review; Retrospective review of clinical history and literature; Follow-up/Duration: 50 years; Study Period: 1966-2008 | General dental patients receiving denture repairs/rebasing over 50 years; Sample Size: Over 200 tons of powder produced in the last 10 years; Number of Sites: Not applicable | Not applicable for this study | Clinical safety, absence of adverse effects |

## Indications for Use

Major.Repair is a cold-curing polymer for dental prosthesis. Poly-methylmethacrylate based. For reparing and rebasing dentures. Powder and liquid. It is used to: - repair and rebase prosthesis - - temporary prosthesis -

## Device Story

Major.repair is a cold-curing, poly-methylmethacrylate (PMMA) based polymer system consisting of powder and liquid components. Used by dental practitioners in clinical or laboratory settings to repair or rebase existing dentures and to fabricate temporary dental prostheses. The device functions via self-curing polymerization. The healthcare provider mixes the powder and liquid to create the material, which is then applied to the prosthesis. This process restores the function and fit of dental appliances, providing a benefit to patients requiring denture maintenance or temporary tooth replacement. The device is Cadmium-free and utilizes pigments approved for alimentary use.

## Clinical Evidence

No clinical data. The device relies on a 50-year history of clinical research, literature, and manufacturing experience for this class of dental materials. Bench testing performed by NIOM Laboratory (Norway) confirmed physical properties: flexural strength 66.4 MPa, flexural modulus 2217 MPa, water absorption 21.2 µg/mm3, water solubility 1.4 µg/mm3, and residual monomer 4.0%.

## Technological Characteristics

Cold-curing poly-methylmethacrylate (PMMA) polymer system. Composed of PMMA powder and methylmethacrylate liquid. Cadmium-free pigments. Complies with ISO 1567:1999. Physical properties: 66.4 MPa flexural strength, 2217 MPa flexural modulus. Manual mixing/application.

## Regulatory Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

## Predicate Devices

- Probase Cold by Ivoclar Vivadent AG ([K913655](/device/K913655.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K082153 510(K) SUMMARY

### NOV 2 5 2008

Image /page/0/Picture/2 description: The image shows a logo for Major Dental. The logo consists of a stylized letter "M" in white against a black background. Below the "M" are the words "MAJOR" and "DENTAL" stacked on top of each other. The text is in a simple, sans-serif font.

E.

MAJOR Prodotti Dentari S.p.A. Via Luigi Einaudi, 23, 10024 Moncalieri (TO) - ITAL Y Tel +39.011.6400.211 - Fax +39.011.6400.222 E-mail: info@majordental.com - Internet: www.majordental.com

#### SUBMITTER

| Name                            | MAJOR Prodotti Dentari S.p.A.                             |
|---------------------------------|-----------------------------------------------------------|
| Address                         | Via Luigi Einaudi, 23<br>I-10024 Moncalieri (TO)<br>ITALY |
| Facility registration<br>number | 4565                                                      |
| Contact person                  | Monica Funai                                              |
| Phone number                    | +39.011.6400.211                                          |
| Fax number                      | +39.011.6400.222                                          |
| e-mail                          | m.funai@majordental.com                                   |
| Document issued on              | July 25, 2008                                             |

#### NAME OF DEVICE

| Proprietary name    | major.repair                                                                         |
|---------------------|--------------------------------------------------------------------------------------|
| Common name         | Cold cure polymer for dental prosthesis                                              |
| Classification name | Denture relining, reparing, or rebasing resin<br>(21 CFR 872.3760, Product Code EBI) |
| Predicate Devices   | Probase Cold by Ivoclar Vivadent AG .........K913655                                 |

#### DESCRIPTION OF THE DEVICE

Major.repair is a cold-curing polymer for dental prosthesis. For reparing and rebasing dentures. Powder and liquid.

It can be used to:

- repair and rebase prosthesis .
- temporary prosthesis .

{1}------------------------------------------------

Major.repair is a self-curing polymer composed of a poly-methylmethacrylate powder and a liquid consisting of methylmethacrylate and other ingredients solution.

Major.repair is substantially equivalent to predicate denture base system presently on the USA market and safety and effectiveness are well documented in the dental literature.

All pigments used are approved for alimentary or similar use and are Cadmium free.

Polymerized material technical data (test performed by NIOM Laboratory – Norway) are the following:

- . Flexural strength: 66.4 MPa
- Flexural modulus: 2217 MPa .
- Water absorption: 21,2 µg/mm3 �
- . Water solubility: 1.4 ug/mm3
- Residual monomer: 4,0 % .

Major.repair is inherently safe when used according to the instructions for use. It is for use only by dental practitioners; it is not intended for OTC use.

## TECHNOLOGICAL CHARACTERISTICS (compared to the predicate device)

Major.repair has the same technological characteristics as the predicate devices since all the devices are complying with ISO 1567:1999 and have:

- . Same intended use
- . Same polymers composition
- Same working technique .

#### NONCLINICAL TESTING AND CONCLUSIONS

Taking in account that:

- the products and their formulation and use are widely described and totally accepted in dentistry (McKabe, J., Applied Dental Materials, 7th, Blackwell Scientif. Publ .; Phillips, J., Skinner's Science of Dental Materials, Saunders Co .; Graig and Oth., Dental Materials, Properties and Manipulation, 6th, Mosby Publ.; O' Brien, W. J., Dental Materials, Properties and Selection, Quintessence Books);
- a large amount of literature has assessed the clinical liability of these product types and their formulation, dedicating special attention to the denture base products (about 12 articles registered on MedLine last year);
- these product types and their formulation are universal in the last fifty years manufacturing and clinical experience, counting for almost the totality of the not pre-formed temporary crowns and bridges in the world;
- the composition and philosophy of the evaluated products are exactly within what is described in most general literature:
- our company manufactures and sells these products, under similar formulations and processes, since 1966, without any record of human incompatibility or

{2}------------------------------------------------

clinical evidence of adverse effects, having totaled, directly or through connected companies or transferred technologies, more than 200 tons of powders in the last ten years;

we can assess these statements:

- the materials with which the products are manufactured, own a 50 years history । of clinical research, testing and literature, which is largely sufficient not to request any further clinical research or testing;
- the products can then be considered clinically tested to be safe; -
- the clinical monitoring is effected independently by the general literature and it is largely sufficient not to request any further research or testing.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with stylized wings.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

#### NOV 2 5 2008

Ms. Monica Funai Ouality and Regulatory Affairs MAJOR Prodotti Dentari S.p.A. Via Luigi Einaudi, 23 I-10024 Moncalieri (Torino) ITALY

Re: K082153

Trade/Device Name: Major.Repair Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: November 4, 2008 Received: November 6, 2008

Dear Ms. Funai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Ms. Funai

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sultie Y. Maheen Oms

Chiu S. Lin. Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Major Dental. The logo is a black square with a white letter "M" in the center. Below the square, the words "MAJOR DENTAL" are written in a sans-serif font. The word "MAJOR" is on top of the word "DENTAL".

# Indications for Use

Ko8 2153 510(k) Number (if known):

Device Name: major.repair

Indications for Use: Major.Repair is a cold-curing polymer for dental prosthesis. Poly-methylmethacrylate based. For reparing and rebasing dentures. Powder and liquid.

It is used to:

- repair and rebase prosthesis -
- temporary prosthesis -

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Robert Betz DDS for Dr. Saun Lunner
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K082153

---

**Source:** [https://fda-staging.innolitics.com/device/K082153](https://fda-staging.innolitics.com/device/K082153)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
