← Product Code [JIT](/productcode/JIT) · K082141

# ARCHITECT C-PEPTIDE CALIBRATORS, AND ARCHITECT C-PEPTIDE CONTROLS, MODELS 3L53-01, 3L53-10 (K082141)

_Biokit, S.A. · JIT · Sep 3, 2008 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082141

## Device Facts

- **Applicant:** Biokit, S.A.
- **Product Code:** [JIT](/productcode/JIT.md)
- **Decision Date:** Sep 3, 2008
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ARCHITECT C-Peptide Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine. The ARCHITECT C-Peptide Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine.

## Device Story

ARCHITECT C-Peptide Calibrators and Controls are liquid reagents used with the ARCHITECT i System to calibrate assays and monitor test precision/accuracy. Calibrators consist of synthetic human C-peptide in PBS buffer with horse serum and stabilizers; concentrations range from 0.00 to 30.00 ng/mL. Controls consist of synthetic human C-peptide in PBS buffer with horse serum and stabilizers; concentrations are 1, 4, and 16 ng/mL. Used in clinical laboratory settings by trained personnel. Calibrators establish the system's calibration curve; controls verify system performance and detect analytical deviations. Results assist clinicians in assessing C-peptide levels, which are relevant for evaluating pancreatic beta-cell function.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies (freeze-thaw, open vial, transport simulation) confirmed 6-month stability at 2-8°C. Traceability established to WHO standard NIBSC 84/510 via gravimetric dilution and RLU comparison on the ARCHITECT i System.

## Technological Characteristics

Materials: PBS buffer, heat-inactivated horse serum, synthetic human C-peptide, ProClin 300/950 preservatives. Principle: Chemiluminescent Microparticle Immunoassay (CMIA). Form factor: Liquid reagents in bottles. Connectivity: Designed for use on the ARCHITECT i System. Sterilization: Not specified.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- ADVIA Centaur and ACS:180 C-peptide Calibrator (k021532)
- BAYER Ligand Plus 1,2,3 Controls (k030452)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k082141

B. Purpose for Submission:
New Device

C. Measurand:
Calibrators and controls for C-peptide

D. Type of Test:
Calibrator and Control

E. Applicant:
Biokit S.A.

F. Proprietary and Established Names:
ARCHITECT C-Peptide Calibrators (model 3L53-01)
ARCHITECT C-Peptide Controls (model 3L53-10)

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JIT (calibrator) | Class II | 21 CFR§ 862.1150 | Clinical Chemistry  |
|  JJX (control) | Class I, reserved | 21 CFR§ 862.1660 | Clinical Chemistry  |

H. Intended Use:

1. Intended use(s):
See indications for use.

2. Indication(s) for use:
Calibrators: The ARCHITECT C-Peptide Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine.

Controls: The ARCHITECT C-Peptide Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine.

3. Special conditions for use statement(s):
For in vitro diagnostic use.

4. Special instrument requirements:
ARCHITECT i System

{1}

2

I. Device Description:

The ARCHITECT C-Peptide Calibrators and Controls are designed to be used in the ARCHITECT C-Peptide assay on the ARCHITECT i System.

Calibrators:

Calibrators A-F contain PBS buffer with heat inactivated horse serum and stabilizers. Calibrators B-F also contain human C-peptide (synthetic). Preservatives used are ProClin 300 and ProClin 950. Each ARCHITECT C-Peptide Calibrator kit contains 6 bottles of Calibrators (4.0 mL fill volume per bottle) with the following concentrations:

CAL A – 0.00 ng/mL, CAL B – 0.05 ng/mL, CAL C – 0.24 ng/mL, CAL D – 1.20 ng/mL, CAL E – 6 ng/mL, CAL F – 30 ng/mL

Controls:

The controls contain human C-peptide (synthetic) in PBS buffer with heat inactivated horse serum and stabilizers. Preservatives used are ProClin 300 and ProClin 950. Each ARCHITECT C-Peptide Control kit contains 3 bottles of Controls (8.0 mL fill volume per bottle) with the following concentrations:

Control Low – 1 ng/mL, Control Medium – 4 ng/mL, Control High – 16 ng/mL

J. Substantial Equivalence Information:

1. Predicate device name(s):

Calibrators: ADVIA Centaur and ACS:180 C-peptide Calibrator

Controls: BAYER Ligand Plus 1,2,3 Controls

2. Predicate 510(k) number(s):

Calibrators: k021532

Controls: k030452

3. Comparison with predicate:

|  Calibrators (Similarities)  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | The ARCHITECT C-Peptide Calibrators are for the calibration of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine. | For in vitro diagnostic use in calibrating the ADVIA Centaur or ACS:180 C-peptide assays.  |
|  System Methodology | Chemiluminescent Microparticle Immunoassay (CMIA). | Two-site sandwich immunoassay using direct chemiluminescent technology.  |
|  Assay Protocols | Two-step immunoassay | Two-site sandwich immunoassay  |

{2}

|  Calibrators (Differences)  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Platform | ARCHITECT i System | Siemens Advia Centaur  |
|  Calibration Range/Levels | 0.00 ng/mL (0 pmol/L)
0.05 ng/mL (17 pmol/L)
0.24 ng/mL (80 pmol/L)
1.20 ng/mL (400 pmol/L)
6.00 ng/mL (2000 pmol/L)
30.00 ng/mL (10000 pmol/L) | Low Calibrator
High Calibrator  |
|  Matrix | PBS buffer with heat inactivated horse serum and stabilizers. | Citric acid buffer with casein and preservatives.  |
|  Controls (Similarities)  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | The ARCHITECT C-Peptide Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System when used for the quantitative determination of C-peptide in human serum, plasma and urine. | For in vitro diagnostic use to monitor the precision and the accuracy of immunochemistry test procedures for ADVIA Centaur®, ACS:180®, ADVIA IMS and ADVIA® Chemistry Systems.  |
|  System Methodology | Chemiluminescent Microparticle Immunoassay (CMIA). | Two-site sandwich immunoassay using direct chemiluminescent technology.  |
|  Assay Protocols | Two-step immunoassay | Two-site sandwich immunoassay  |
|  Controls (Differences)  |   |   |
|  Item | Device | Predicate  |
|  Platform | ARCHITECT i System | Siemens ADVIA Centaur  |

{3}

|  Control
Range/Levels | 1.00 ng/mL (333 pmol/L)
4.00 ng/mL (1333 pmol/L)
16.00 ng/mL (5333 pmol/L) | Control 1
Control 2
Control 3  |
| --- | --- | --- |
|  Matrix | PBS buffer with heat inactivated horse serum and stabilizers. | Prepared from Human serum. No preservatives or stabilizers added.  |

K. Standard/Guidance Document Referenced (if applicable):
- OIVD: Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final

L. Test Principle:
Not Applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Calibrators and controls are traceable to the WHO standard NIBSC code 84/510. The WHO standards are gravimetrically diluted to the target concentrations defined for the calibrators and controls to create reference materials. The reference materials are then used to standardize the consumer product.

Stability:
Freeze-thaw, open vial, and transport simulation studies were performed on the products. The stability protocols and acceptance criteria for these studies were reviewed and found to be acceptable. The ARCHITECT C-Peptide Calibrators and Controls are stable for 6 months when stored at 2-8°C.

Value Assignment:
Values for the calibrators and controls are listed in section I. Calibrators and controls are verified using Relative Light Unit (RLU) comparison between target material and test material using the ARCHITECT i System. The consumer products are referenced to a secondary standard which is in turn referenced to the WHO standard mentioned above. Adjustments are performed if necessary so that the values of new calibrators are within the sponsor's manufacturing criteria when compared to the reference.

{4}

d. Detection limit:
Not Applicable

e. Analytical specificity:
Not Applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not Applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K082141](https://fda-staging.innolitics.com/device/K082141)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
