← Product Code [KZE](/productcode/KZE) · K082138

# AIRGENT, MODEL AG7000010 (K082138)

_Perfaction, Inc. · KZE · Jan 23, 2009 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K082138

## Device Facts

- **Applicant:** Perfaction, Inc.
- **Product Code:** [KZE](/productcode/KZE.md)
- **Decision Date:** Jan 23, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5430
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Airgent™ System is a needle-free injection system designed for the administration of various medicines and vaccines to the body by means of a high velocity jet of fluid that penetrates the skin The Airgent™ includes a dısposable delivery kıt ıntended for multıple ınjections on a sıngle patient The Airgent™ System is indicated for professional use only

## Device Story

Airgent™ is an automated, pneumatically powered, needle-free injection system; delivers medications/vaccines via high-velocity fluid jet through small skin entry point. System consists of a console and single-use sterile injector kit. Used in professional clinical settings; operated by healthcare professionals. User controls dosage (150ul/200ul) and injection pressure via console graphical user interface. Injector kit is disposable and intended for multiple injections on a single patient. Device benefits include needle-free administration, reducing needle-stick risks and patient discomfort.

## Clinical Evidence

Bench testing only. Ex-vivo testing conducted to confirm performance and safety in accordance with ISO 21649 (Needle-free Injectors for Medical Use). Results demonstrated the device functioned as intended and was as safe and effective as the predicate devices.

## Technological Characteristics

Pneumatically powered fluid injector; console-based system with single-use sterile injector kit. Dosage control (150ul/200ul) and pressure regulation via graphical user interface. Tested per ISO 21649. Class II device.

## Regulatory Identification

A nonelectrically powered fluid injector is a nonelectrically powered device used by a health care provider to give a hypodermic injection by means of a narrow, high velocity jet of fluid which can penetrate the surface of the skin and deliver the fluid to the body. It may be used for mass inoculations.

## Predicate Devices

- Biojector 2000 ([K960373](/device/K960373.md))
- Jet Syringe ([K003741](/device/K003741.md))
- Injex 30 ([K022148](/device/K022148.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kortisy

..

# 510(k) Summary

# PerfAction's Airgent™

JAN 2 3 2009

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

PerfAction, Ltd 4 Pekeris Street, Rehovot 7670211 Israel Phone +972-8-9363004 Facsımıle +972-8-9471540

Contact Person. Dalia Givony, VP Regulatory & Clinical Affairs

Date Prepared July 28, 2008

### Name of Device and Name/Address of Sponsor

1 Airgent™ PerfAction, Inc 4 Pekeris Street, Rehovot 7670211 Israel

Common or Usual Name Airgent™

Model Airgent™ US

Classification Name/Product Code/CFR Reference Nonelectrically powered fluid injector

Product Code KZE, 21 CFR § 880 5430

#### Device Class

Class II

### Classification Panel

General Hospital

# Predicate Devices

Brojector 2000 (Bioject Inc , K960373) Jet Syringe (Equidyne Systems, Inc , K003741) 

{1}------------------------------------------------

Injex 30 (Equidyne Systems, Inc , K022148)

#### Intended Use/Indications for Use

The Airgent™ System is a needle-free injection system designed for the administration of various medicines and vaccines to the body by means of a high velocity jet of fluid that penetrates the skin The Airgent™ includes a dısposable delivery kıt ıntended for multıple ınjections on a sıngle patient The Airgent™ System is indicated for professional use only"

### Device Description

The Airgent" System is an automated, multi-use, needless injector system, intended to deliver medications and vaccines to the body by a highly accelerated pneumatically powered jet of fluid via a very small entry point in the surface of the skin The system is comprised of a console and single-use sterile injector kit

The user may control the dosage (150ul/200ul) and system pressure for the injection via the graphical user interface on the front panel of the console The entire injector kit is replaced between patients and may be used for multiple injections per patient

### Performance Data

The Argent performance and safety was tested in accordance with ISO 21649 -Needle-free Injectors for Medical Use Requirements and Test Methods

#### Comparison Testing

The equivalency test was performed to demonstrate that Airgent™ is as safe and effective as its predicates Specifically, ex-vivo testing was conducted to confirm the performance of Argent The testing results demonstrated that the Airgent is as safe and effective as its predicate devices and in all instances, the Airgent functioned as intended

# Substantial Equivalence

The Argent System is as safe and effective as the B10jector, Jet Syringe and Injex 30 It has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices The minor technological differences between the Airgent and its predicate devices rase no new issues of safety or effectiveness Performance data demonstrate that the Airgent is as safe and effective as the Injex 30 Thus, the Airgent System is substantially equivalent to the identified predicate devices

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem that resembles an abstract eagle or bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 3 2009

PerfAction, Incorporated C/O Mr Jonathan S. Kahan Hogan & Hartson L L P 555 Thirteen Street, North West Washington, DC 20004

K082138 Re

> Trade/Device Name AurgentTM Regulation Number: 21 CFR 880 5430 Regulation Name. Nonelectrically Powered Fluid Injector Regulatory Class II Product Code KZE Dated. January 16, 2009 Received January 16, 2009

Dear Mr Kahan

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a prematket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register

{3}------------------------------------------------

Page 2 - Mr Kahan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requrements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115 Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the DIVISION of Surveillance Systems at 240-276-3464. You may obtain other general unformation on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Antuan D. Matin Jr.

Ginette Y. Michaud, M. D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| Indications for Use Statement |          |
|-------------------------------|----------|
| 510(k) Number (if known)      | K082138  |
| Device Name                   | Airgent™ |

Indications for Use:

The Airgent™ System is a needle-free injection system designed for the administration of various medicines and vaccines to the body by means of a high velocity jet of fluid that penetrates the skin. The Airgent™ includes a disposable delivery kit intended for multiple injections on a single patient. The Airgent™ System is indicated for professional use only.

| Prescription Use _X<br>(Part 21 C F R 801 Subpart D) | AND/OR | Over-The-Counter Use_<br>(21 C.F R 807 Subpart C) |
|------------------------------------------------------|--------|---------------------------------------------------|
|------------------------------------------------------|--------|---------------------------------------------------|

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                                                               |         |
|-----------------------------------------------------------------------------------|---------|
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices |         |
| 510(k) Number:                                                                    | K082138 |

| Page | 1 | of | 1 |
|------|---|----|---|
|------|---|----|---|

\\\DC 081511/000001 2848126 v1

---

**Source:** [https://fda-staging.innolitics.com/device/K082138](https://fda-staging.innolitics.com/device/K082138)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
