DEPUY SPINE LATERAL CAGE SYSTEM

K082128 · Depuy Spine · MAX · Nov 14, 2008 · Orthopedic

Device Facts

Record IDK082128
Device NameDEPUY SPINE LATERAL CAGE SYSTEM
ApplicantDepuy Spine
Product CodeMAX · Orthopedic
Decision DateNov 14, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Lateral Cage System is indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive lateral approach. The Lateral Cage System is intended for use with DePuy Spine supplemental fixation.

Device Story

Lateral Cage System consists of intervertebral body fusion implants; available in various heights and lordotic configurations; open architecture design allows packing with autogenous bone graft. Implants are inserted via open or minimally invasive lateral surgical approach; intended for use with DePuy Spine supplemental fixation. Device provides structural support to facilitate spinal fusion in patients with degenerative disc disease. System includes manual surgical instruments and cases.

Clinical Evidence

Bench testing only.

Technological Characteristics

Manufactured from Carbon Fiber Reinforced Polymer. Device features an open architecture for bone graft containment. System includes manual surgical instruments. No electronic components or software.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KCS 2128 Page 1 of 1 ## 510(k) Summary | 510(k) Summary | | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | SUBMITTER: | DePuy Spine, Inc.<br>325 Paramount Drive<br>Raynham, MA 02780 | | | NOV 14 2008 | | CONTACT PERSON: | Hande Tufan | | DATE PREPARED: | June 1, 2008 | | CLASSIFICATION NAME: | Intervertebral Fusion Device with Bone Graft -<br>87MAX - 888.3080 - Class II | | PROPRIETARY NAME: | Lateral Cage System | | PREDICATE DEVICES: | LUMBAR I/F CAGE System (P960025)<br>NuVasive CoRoent System (K071795) | | DEVICE DESCRIPTION: | The Lateral Cage System consists of implants<br>available in various heights and lordotic configurations<br>with an open architecture to accept packing of bone<br>graft materials. | | | The Lateral Cage System also contains Class 1<br>manual surgical instruments and cases that are<br>considered exempt from premarket notification. | | INTENDED USE: | The Lateral Cage System is indicated for intervertebral body<br>fusion with autogenous bone graft in patients with<br>degenerative disc disease (DDD) at one or two contiguous<br>levels from L2 to S1. These DDD patients may also have up<br>to Grade 1 Spondylolisthesis or retrolisthesis at the involved<br>levels. DDD is defined as discogenic back pain with<br>degeneration of the disc confirmed by history and<br>radiographic studies. These patients should be skeletally<br>mature and have had six months of non-operative treatment.<br>These implants may be implanted via an open or a minimally<br>invasive lateral approach. The Lateral Cage System is<br>intended for use with DePuy Spine supplemental fixation. | | MATERIALS: | Manufactured from Carbon Fiber Reinforced Polymer. | | PERFORMANCE<br>DATA: | Performance data were submitted to characterize the Lateral<br>Cage System. | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DePuv Spine % Ms. Handc Tufanyazici Senior Regulatory Affairs Associate 325 Paramount Drive Raynham Massachusetts 02767 NOV 1 4 2008 Re: K082128 Trade/Device Name: Lateral Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: MAX Dated: November 07, 2008 Received: November 12, 2008 Dear Ms. Tufanyazici: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or other to be of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I edual states and see and see and see and limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in all qualis) by over is on the end of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ ## Pagc 2 - Ms. Hande Tufanyazici This letter will allow you to begin marketing your device as described in your Section 510(k) rms reter witification. The FDA finding of substantial equivalence of your device to a legally marketed predicated device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you acon Copter for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to (210) 216 0167. alion's (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Milliman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use IV. 510(k) Number (if known): K082128 Device Name: Lateral Cage System Indications For Use: The Lateral Cage System is indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive lateral approach. The Lateral Cage System is intended for use with DePuy Spine supplemental fixation. X Prescription Use: (Per 21 CFR 801.109) (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) ision Sign-Division of General, Restorative, and Neurological Devices **510(k) Number** K082464
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...