Demetech Absorbable Polydioxanone Surgical Suture is indicated for use in all types of soft tissue approximation including pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable suture with extended wound support (up to six weeks) is desirable.
Device Story
Demetech Polydioxanone (PDO) Suture is a sterile, synthetic, absorbable monofilament surgical suture composed of poly(p-dioxanone) polyester. It is supplied in various lengths and diameters, with or without attached needles, and is available un-dyed or dyed with D&C Violet No. 2. Used by surgeons for soft tissue approximation, the suture provides extended wound support for up to six weeks. Absorption occurs via hydrolysis, characterized by a gradual loss of tensile strength without significant mass loss; strength retention is approximately 85% at 120 days and 25% at 180 days, with complete absorption by 220 days. The device is intended for clinical use in surgical settings.
Clinical Evidence
Bench testing only. The device meets or exceeds USP 31 performance requirements for absorbable surgical sutures, including diameter, tensile strength, and needle attachment. Implantation studies in animals were referenced to characterize the absorption profile and tensile strength retention over time.
Technological Characteristics
Synthetic monofilament absorbable suture composed of poly(p-dioxanone). Meets USP 31 standards for diameter, tensile strength, and needle attachment. Supplied sterile, un-coated, in various lengths/diameters. Colorant: D&C Violet No. 2 (21 CFR 74.3602).
Indications for Use
Indicated for soft tissue approximation in pediatric cardiovascular tissue (where growth is expected) and ophthalmic surgery. Contraindicated for adult cardiovascular, microsurgery, and neural tissue.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
CP Medical Mono-Dox Synthetic Polydioxanone Absorbable Suture (K013274)
Ethicon PDS II Synthetic Absorbable Monofilament Suture (N 18331)
Submission Summary (Full Text)
{0}------------------------------------------------
510(K) Summary
Polydioxanone Synthetic Absorbable Monofilament Suture
JAN - 8 2010
This summary is submitted in accordance with the Safe Medical Device Act (SMDA) of 1990 and Title 21 CFR § 807.92. This summary demonstrates the equivalence of Demetech Sutures to those of the legally marked devices listed.
| Applicant: | Demetech Corporation, |
|----------------------|--------------------------------------------------------------------------|
| | 8935 NW 27th Street, Miami FL. 33172 |
| Contact Person: | A. J. Dimercurio<br>tony@demetech.us |
| Date Revised: | September 27, 2009 |
| Device Names: | |
| Trade Name: | DemeTECH Polydioxanone Synthetic Monofilament<br>(PDO) Absorbable Suture |
| Common Name: | DemeTECH Synthetic PDO Monofilament Absorbable<br>Suture |
| Classification Name: | Absorbable Polydioxanone Surgical Suture |
Predicate Devices:
Demetech Polydioxanone (PDO) Synthetic Absorbable Suture is substantially equivalent to these predicate devices:
- . CP Medical Mono-Dox Synthetic Polydioxanone Absorbable Suture 510K Number K013274, CP Medical Portland Oregon
- . Ethicon PDS II Synthetic Absorbable Monofilament Suture Polydioxanone Suture PMA Number N 18331, Ethicon Inc
#### Device Description:
Demetech's Polydioxanone is a synthetic monofilament absorbable surgical suture composed of polyester polymers poly (p-dioxanone) and is supplied un-dyed violet with D & C violet #2. Demetech's Polydioxanone surgical sutures meet the requirements established by the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures.
> 8935 NW 27 Street, Miami, Florida, 33172- USA Tel (305) 597-5277 · Toll Free (888) 324-2447 · Fax: (305) 437-7607 Web: www.demetech.us · Email tony@demetech.us
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for DemeTECH. The logo consists of a circular icon on the left and the word "DemeTECH." on the right. The icon contains a triangular shape inside the circle. The text is in a bold, sans-serif font.
510(K) Summary Polydioxanone Synthetic Absorbable Monofilament Suture
- A. Intended Use:
Demetech Absorbable Polydioxanone Surgical Suture is indicated for use in general soft tissue approximation including pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable suture with extended wound support (up to six weeks) is desirable.
# Technological Comparison to Predicate Devices:
# COMPARISON TABLE DEMETECH POLYDIOXANONE TO PREDICATE DEVICES
| Suture Material is a synthetic monofilament absorbable<br>surgical suture composed of polyester polymers poly (p-<br>dioxanone) | Same | Same | Same |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------|--------------------|
| Suture material is offered un-dyed and dyed with the FDA listed<br>colorant, D&C Violet No. 2 (21 CFR 74.3602) | Same | Same | Same |
| Suture Material is supplied un-coated | Same | Same | Same |
| Suture Material is designed being a sterile, flexible, monofilament<br>thread offered in a variety of lengths and a range of diameters with<br>or without various needles attached. | Same | Same | Same |
| Suture material absorption begins as a loss of tensile strength<br>without appreciable loss of mass. Implantation studies in animals<br>indicate that Polydioxanone retains approximately 85% of its<br>original tensile strength at 120 days post implantation, with<br>approximately 25% remaining at 180 days. Absorption<br>polydioxanone surgical suture is essentially complete at 220 days. | Same | Same | Same to<br>Similar |
| The Suture Material is "Intended for Use" in general soft tissue<br>approximation including pediatric cardiovascular tissue where<br>growth is expected to occur and ophthalmic surgery, but not<br>for use in adult cardiovascular, microsurgery and neural<br>tissue. These sutures are useful where absorbable suture with<br>extended wound support (up to six weeks) is desirable. | Same | Same | Same |
| Suture Material meets or exceeds the performance requirements for<br>"Absorbable Surgical Suture" as defined in the Official Monograph<br>of the United States Pharmacopeia 31. | Same | Same | Same |
8935 NW 27 Street, Miami, Florida, 33172- USA Tel (305) 597-5277 · Toll Free (888) 324-2447 · Fax: (305) 437-7607 Web: www.demetech.us · Email_tony@demetech.us
{2}------------------------------------------------
K082097
Image /page/2/Picture/1 description: The image shows the logo for "DemeTECH". The logo consists of a circular icon on the left and the company name on the right. The icon features a triangle inside a circle, and the company name is written in a bold, italicized font.
# 510(K) Summary Polydioxanone Synthetic Absorbable Monofilament Suture
# Technological Comparison to Predicate Devices (continued)
| COMPARISON TABLE TO PREDICATE DEVICES | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------|-------------------|
| Comparison Items | Demetech<br>Polydioxanone<br>Suture | CP Medical Mono-<br>Dox | Ethicon<br>PDS II |
| Suture Materials meet the performance<br>requirements for Diameter as defined in the<br>United States Pharmacopeia 31 | Same | Same | Same |
| Suture Material meets or exceeds the performance<br>requirements defined in the United States<br>Pharmacopeia 31 for "Tensile Strength" < 881 > | Same | Same | Same |
| Suture Material meets or exceeds the performance<br>requirements defined in the United States<br>Pharmacopeia 31 and the current edition USP 31<br>for "Needle Attachment" < 871 > | Same | Same | Same |
| Suture Material meets or exceeds the performance<br>requirements defined in the United States<br>Pharmacopeia 31 for "Suture Length<br>Requirement" (95% of stated label length) | Same | Same | Same |
| Suture Material is packaged in a same or equivalent<br>manner with sterile single or double package<br>having labeling conforming to 21 CFR and USP<br>XXXI. | Same | Same | Same |
Demetech Polydioxanone Suture is composed of the same material, as are the predicated devices and having the same design being a sterile, flexible, monofilament threads meeting all the requirements of the United States Pharmacopeia. Demetech Polydioxanone Synthetic Monofilament Absorbable Suture is manufactured in the same manner as the predicate devices, being composed of polyester polymers poly (p-dioxanone) and produced in operations considered standard in the fiber industry to form the finished suture fiber. The manufacturer supplies to Demetech the same suture materials as it does to other suture manufacturers including some of those listed above.
The information and the results of the performance testing presented demonstrate the substantial equivalence of Demetech's Polydioxanone Synthetic Monofilament Absorbable Suture to that of the predicate devices. It further demonstrates Demetech suture is safe and effective for its intended purpose.
> 8935 NW 27 Street, Miami, Florida, 33172- USA Tel (305) 597-5277 · Toll Free (888) 324-2447 · Fax: (305) 437-7607 Web: www.demetech.us · Email tony@demetech.us
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem resembling an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
#### JAN - 8 2010
DemeTECH Corporation % Mr. Anthony Dimercurio Vice President of Manufacturing 8935 NW 27" Street Miami, Florida 33172
Re: K082097
Trade/Device Name: DemeTECH Polydioxanone Synthetic Monofilament (PDO) Absorbable Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable polydioxanone surgical suture Regulatory Class: Class II Product Code: NEW Dated: December 21, 2009 Received: January 06, 2010
Dear Mr. Dimercurio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Anthony Dimercurio
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours.
urs,
fur
lkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
ﻢ ﺗ
{5}------------------------------------------------
K082097
হবে
ార్లు
Image /page/5/Picture/1 description: The image shows the logo for "DemeTECH." The logo consists of a circular graphic to the left of the company name. The graphic contains a triangular shape inside of the circle. The company name is written in a stylized font, with the "T" in "TECH" being larger than the other letters. There is a registered trademark symbol to the right of the company name.
Intended Use Statement
# Indication for use
510K Number:
Device Name: Demetech Absorbable Polydioxanone Surgical Suture. As per 21CFR 878.4840
Indication for Use:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Demetech Absorbable Polydioxanone Surgical Suture is indicated for use in all types of soft . tissue approximation including pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery, but not for use in adult cardiovascular, microsurgery and neural tissue. These sutures are useful where absorbable suture with extended wound support (up to six weeks) is desirable.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ And/Or Over the-Counter Use_ (Part 21 CFR 801; Subpart D) (21 CFR 801; Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Keane
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K082097
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.