WELL LIFE SELF ADHESIVE ELECTRODE / MODELS: CM, FA, PU AND SP

K082065 · Well-Life Healthcare Limited · GXY · Apr 10, 2009 · Neurology

Device Facts

Record IDK082065
Device NameWELL LIFE SELF ADHESIVE ELECTRODE / MODELS: CM, FA, PU AND SP
ApplicantWell-Life Healthcare Limited
Product CodeGXY · Neurology
Decision DateApr 10, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Well Life Self Adhesive Electrode / CM, FA, PU and SP series are intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Well Life Self Adhesive Electrode / AP, CM, FA, PU and SP series are designed and intended to be used with marketed Electrical Stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation), MENS (Microcurrent Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) and PGS (Pulsed Galvanic Stimulation).

Device Story

Well Life Self Adhesive Electrodes are disposable, flexible, multi-layered cutaneous electrodes. Construction includes an exterior layer (fabric, foam, plastic film, or silicon), a conductive plastic film, and a biocompatible conductive hydrogel. Electrodes interface between patient skin and external electrical stimulators (TENS, MENS, EMS, IF, PGS). Connection options include snap connectors or lead wire assemblies with female sockets. Designed for single-patient, multiple-application use; adhesive hydrogel eliminates need for secondary securing materials. Used in clinical or home settings by patients or clinicians. Output is the delivery of electrical current from the stimulator to the patient's skin for therapeutic stimulation. Benefits include non-invasive, convenient, and reusable interface for various neurostimulation and muscle stimulation modalities.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1 and Tripartite Biocompatibility Guidance. Electrical performance testing conducted per ANSI/AAMI EC12.

Technological Characteristics

Non-sterile, disposable, laminated flexible structure. Layers: exterior (fabric/foam/plastic/silicon), conductive plastic film, biocompatible conductive hydrogel. Connection: snap or lead wire. Standards: ISO 10993-1 (biocompatibility), ANSI/AAMI EC12 (electrical performance).

Indications for Use

Indicated for use as a conductive interface between patient skin and electrical stimulators (TENS, MENS, EMS, IF, PGS) for patients requiring electrical stimulation therapy. Suitable for both prescription and OTC applications.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K082065 #### 510(K) SUMMARY ## APR 1 0 2009 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: #### 1. Submitter's Identifications: | Company Name: | Well Life Healthcare Limited | |-----------------|---------------------------------------------------------------------------------| | Contact person: | Jenny Hsieh | | Address: | 1FL., No.16, Lane 454, Jungjeng Rd., Yunghe City, Taipei County, Taiwan, R.O.C. | | TEL No.: | 886-2-2928-2112 | | Fax No.: | 886-2-2928-1880 | | E-mail address: | jenny@welllifehealthcare.com.tw | Date of Summary Preparation: July 14, 2008. - 2. Device Identification: Classification Name: Electrode, Cutaneous Trade/Proprietary Name: Well Life Self Adhesive Electrode / CM, FA, PU, and SP series Predicate devices: K062675 (Gemore Technology Co., Ltd) #### 3. Device Description Well Life Self Adhesive Electrode / CM, FA, PU, and SP series are non-sterile, disposable laminated, flexible structures composed of the following three main construction layers: First Layer – the exterior layer which may be made of the following different type of material coated with adhesive : white fabric(CM), white foam(FA), printed plastic film(PU) or silicon pad(SP). Second Layer - Conductive plastic film. Third Layer – Biocompatible conductive hydrogel coupling media The electrodes are designed for single-patient/multiple application use. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device the patient's skin. The electrode has one type of connection point that can be used to connect the stimulation device to the electrodes. This connection point is compatible with all standard, marketed Neurostimulation and muscle-stimulation devices. For the electrical connection, Well Life provides two different types: - 1 > Snap Series Snap connection 1.65" standard size of mail. Snap is provided to connect to the wire female snap. - 2> Wire Series Lead wire assembly 4.5" ~ 6" wire with .080 in. diameter female socket connected to one side of the wire. {1}------------------------------------------------ #### Predicate Technological Characteristics Comparison: ধ Well Life Self Adhesive Electrode / CM, FA, PU, and SP series are technologically equivalent to the predicate devices. They are physically and technically similar to the those currently being marketed for "Neurostimulation" i.e., TENS (Transcutaneous Electrical Nerve Stimulation), MENS (Microcurrent Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) and PGS (Pulsed Galvanic Stimulation). Since currently the above mentioned devices (TENS & EMS) are capable of being sold as both prescription and OTC medical device, we claimed that our electrodes are capable of being sold at both prescription and OTC TENS and EMS device application so as to fit that marketing requirement. #### 5. Safety and Effectiveness: Well Life Self Adhesive Electrode / CM, FA, PU, and SP series are as safe and effective as the K062675 (Gemore Technology Co., Ltd) which was previously found to be substantially equivalent to the 510(K) cleared models via 510(k) Premarket Notifications. The first safety issue consideration is whether or not the skin contact gel, which is used to adhere the electrode to the skin to transfer the electric current to the use patient, would cause any skin irritation. To ensure the conformity of this biocompatibility issue, we use the Axelgaard gel which has been proved suitable for the use in the skin contact part for the Axelgaard 510(K) cleared electrode with 510(K) number K983741 and K000947. or the other alternative biocompatibility material with ISO-10993 conformity certificate. All of them have passed the required skin sensitivity testing criteria as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. For the electricity performance, we have conducted and completed the electrical safety testing according to the chosen performance standard, ANSI/AAMI EC12. The testing report was included in this submission. ### 6. Conclusion: Based upon the above mention. Well Life considers its Self Adhesive electrodes to be as safe and effective as the predicate devices being marketed and manufactured by: K062675 (Gemore Technology Co., Ltd). {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Well-Life Healthcare Ltd. % Ms. Jenny Hsieh 1 FL, No. 16, Lane 454 Jungjeng Rd. Yunghe City, Taipei County Taiwan, R.O.C APR 1 0 2009 #### Re: K082065 Trade/Device Name: Well-Life Self Adhesive Electrodes/CM, FA, PU, and SP models Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 7, 2009 Received: March 9, 2009 #### Dear Ms. Hsieh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Jenny Hsieh If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/ You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sincerely yours, Mark A. Millhiser . Mark N. Melkerson Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications For Use 510(k) Number (if known): Device Name: Well Life Self Adhesive Electrode / CM, FA, PU and SP series Indications For Use: The Well Life Self Adhesive Electrode / CM, FA, PU and SP series are intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Well Life Self Adhesive Electrode / AP, CM, FA, PU and SP series are designed and intended to be used with marketed Electrical Stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation), MENS (Microcurrent Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) and PGS (Pulsed Galvanic Stimulation). Prescription Use v (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of DRH, Office of Device Evaluation (ODE) OR (Divis n Sign-Off) Division of Ceneral, Restorative, and Neurological Devices 0(k) Number K082063 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...