K082060 · Brainlab AG · HAW · May 8, 2009 · Neurology
Device Facts
Record ID
K082060
Device Name
CRANIAL IMAGE GUIDED SURGERY SYSTEM
Applicant
Brainlab AG
Product Code
HAW · Neurology
Decision Date
May 8, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a magnetic sensor system or a passive marker sensor system to a virtual computer image space on patient image data being processed by the IGS workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, CTA, X-Ray, MR, MRA and ultrasound based model of the anatomy.
Device Story
Intra-operative image-guided localization system; used in OR by surgeons. Inputs: pre-operative/intra-operative patient image data (CT, MR, etc.) and real-time 3D tracking data from magnetic or optical sensors attached to surgical instruments (probes, needles, catheters). Workstation processes data to register virtual image space to surgical field via landmark/fiducial matching or surface scanning. Outputs: real-time visualization of instrument position relative to patient anatomy on touch screen monitor; supports integration of microscope/ultrasound feeds. Assists surgeon in navigating minimally invasive procedures; improves accuracy of instrument placement; reduces surgical trauma.
Clinical Evidence
Bench testing only; no clinical data provided. Safety and effectiveness established through verification and validation of system design and software performance.
Technological Characteristics
Workstation-based IGS system; electro-magnetic or optical 3D tracking; touch screen interface. Connectivity via network, data carrier, or bus. Supports integration of external imaging sources (microscope, ultrasound, endoscope). Software-based registration of diagnostic images to surgical field. Components prepared for MRI environments.
Indications for Use
Indicated for patients requiring stereotactic surgery where rigid anatomical structures (skull, long bone, vertebra) can be referenced to CT, CTA, X-Ray, MR, MRA, or ultrasound imaging. Applicable for cranial procedures (tumor resection, biopsy, craniotomy, shunt placement, thalamotomy) and ENT procedures (transphenoidal, sinusotomies, ethmoidectomies).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
## KO82060
# 510 (k) Summary of Safety and Effectiveness for Cranial Image Guided Surgery System
#### Manufacturer:
| Address: | BrainLAB AG<br>Kapellenstrasse 12<br>85622 Feldkirchen<br>Germany<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Per Persson |
Summary Date: July 17, 2008
#### Device Name:
Trade names:
VectorVision cranial VectorVision ENT Kolibri cranial Kolibri ENT Cranial Essential Cranial Unlimited ENT Essential ENT Unlimited
Common/Classification Name:
Image Guided Surgery System / Instrument, Stereotaxic
#### Predicate Devices:
Kolibri Image Guided Surgery System (K042391) VectorVision Cranial/ENT (K023651)
Device Classification Name: Instrument, Stereotaxic Regulatory Class: Class II
#### Intended Use:
This submission does not change the indications for use for the predicate devices.
The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a magnetic sensor system or a passive marker sensor system to a virtual computer image space on patient image data being processed by the IGS workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, CTA, X-Ray, MR, MRA and ultrasound based model of the anatomy.
{1}------------------------------------------------
Example procedures include but are not limited to:
## Cranial Procedures:
. Tumor resections Skull base surgery Cranial biopsies Craniotomies/ Craniectomies Pediatric Catheter Shunt Placement General Catheter Shunt Placement Thalamotomies/ Palliodotomies
#### ENT Procedures:
Transphenoidal procedures Maximillary antrostomies Ethmoidectomies Spheno-idotomies/ sphenoid explorations Turbinate resections Frontal sinusotomies Intranasal procedures
:
{2}------------------------------------------------
### Device Description:
The Cranial IGS System consists of the IGS workstation, the touch screen monitor and the 3D tracking system. A set of hardware accessories provides for comfortable and accurate use of the system.
The IGS workstation holds the patient data during the surgery and runs the cranial software application.
The patient data needed for the image-guided surgery is acquired pre-operatively or intraoperatively and is transferred to the IGS workstation via network, data carrier or data bus, The cranial software application offers the display of the patient data in various reconstructions, segmentations and overlays on the touch screen in addition to position information of tracked instruments - optionally combined with outlined information. The touch screen enables the control of the cranial software application and can be draped for sterile use by the surgeon.
The electro-magnetic or optical 3D tracking system performs the localization of patient and surgical tools within the operating field.
The virtual diagnostic image spaces are correlated ("registered") to the surgical environment by collecting the 3D position of anatomical landmarks or fiducial markers with a tracked pointer probe and relating them with the corresponding features extracted from the diagnostic image data sets. Alternatively, the patient's skin surface can be scanned with a laser device or touched with a pointer device and matched to the 3D reconstruction of the patient data set. If several diagnostic image spaces have been acquired from the same patient, only one of them has to be registered whereas the remaining ones can be fused to the registered data set.
Intra-operatively acquired patient data can furthermore be correlated ("registered") to the surgical environment by determining its spatial position to the patient during its acquisition.
Structures in the patient's body are localized using trackable pre-calibrated or intraoperatively calibrated surgical instruments. Examples of surgical instruments are the pointer tool, biopsy needles, catheter stylets or suction tubes.
Surgical microscopes, ultrasound devices and endoscopes are additional intra-operative image sources, which are connected with the Cranial IGS System via signal transmission cables. They can be calibrated and tracked similar as any other surgical instrument. Their images can be displayed on the touch screen or external monitors and combined with the available patient data in correct spatial relation. The settings of microscope and ultrasound devices offering a communication interface can be controlled from the Cranial IGS System. Navigation information can be displayed in the microscope's image injection module.
Defined components of the Cranial IGS System are prepared for the use in magnetresonance environments.
The Cranial IGS System contains hardware accessories and software features to improve the support and guidance of surgical instruments.
The Cranial IGS System contains a network based software interface that allows downloading medical data (such as image sets, objects, trajectories or points) and, tracking data from the system as well as to upload and display an image stream to the system. This interface can be used to implement custom visualization of medical data (e.g. included modalities which are otherwise unknown to the cranial software application)
{3}------------------------------------------------
as well as to control other devices. These view data is strictly under the responsibility of the user and clearly marked as such.
#### Substantial equivalence:
The Cranial IGS System has been verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate devices Kolibri Image Guided Surgery System (K042391) and VectorVision Cranial/ENT (K023651).
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, superimposed over a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
BrainLAB AG % Mr. Per Person Kapellenstrasse 12 85622 Feldkirchen Germany
MAY - 8 2009
Re: K082060
Trade/Device Name: Cranial Image Guided Surgery System Regulation Number: 21 CFR 882. 4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: May 1, 2009 Received: May 1, 2009
Dear Mr. Person:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Per Person
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at (240) 276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milhusan
Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
#### K082060 510(k) Number (if known):
Device Name: Cranial Image Guided Surgery System
Indications For Use:
The BrainLAB Cranial IGS System is intended to be an intra-operative image guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a magnetic sensor system or a passive marker sensor system to a virtual computer image space on patient image data being processed by the IGS workstation. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, CTA, X-Ray, MR, MRA and ultrasound based model of the anatomy.
Example procedures include but are not limited to:
Cranial Procedures: Turnor resections Skull base surgery Cranial biopsies Craniotomies/ Craniectomies Pediatric Catheter Shunt Placement General Catheter Shunt Placement Thalamotomies/ Palliodotomies
#### ENT Procedures:
Transphenoidal procedures Maximillary antrostomies Ethmoidectomies Spheno-idotomies/ sphenoid explorations Turbinate resections Frontal sinusotomies Intranasal procedures
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
*Mat*
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Oft)*
Division of Surgical, Orthopedic,
and Restorative Devices
| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-------------------------|
| 510(k) Number | K082060 |
| (k) Number | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.