K082039 · El.En Electronic Engineering Spa · GEX · Mar 20, 2009 · General, Plastic Surgery
Device Facts
Record ID
K082039
Device Name
SYNCHRO HP PLATFORM
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nd:YAG Laser (1064 nm): removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions. Er:YAG Laser attachment (2940 nm): skin resurfacing and incision, excision, ablation or vaporization of soft bodily tissues. Q-switched Nd:YAG Laser attachment (1064 nm): removal of dark tattoos and treatment of benign pigmented lesions. Pulsed Light attachments (400, 500, 520, 550, 650 to 950 nm): permanent hair reduction, treatment of benign cutaneous vascular lesions including facial and leg veins, benign pigmented epidermal lesions, moderate inflammatory acne vulgaris. Different wavelength ranges of Pulsed Light attachments are indicated for the various treatments and skin types, as indicated in the following table: [Table data provided in source].
Device Story
Synchro HP Platform is a multi-application medical laser and pulsed light system. Inputs include user-selected laser/light sources: long-pulse Nd:YAG (1064 nm), Er:YAG (2940 nm), Q-switched Nd:YAG (1064 nm), and various pulsed light attachments (400-950 nm). Device transforms electrical energy into specific optical wavelengths for dermatological and surgical procedures. Operated by clinicians in a professional setting; emission activated via footswitch or finger switch. Output is directed optical energy used for tissue ablation, coagulation, or hair reduction. Clinical benefit includes removal of unwanted hair, treatment of vascular/pigmented lesions, and skin resurfacing. System is a floor-standing unit (170 kg).
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Multi-application platform featuring long-pulse Nd:YAG (1064 nm), Er:YAG (2940 nm), Q-switched Nd:YAG (1064 nm), and pulsed light sources (400-950 nm). Electrical requirements: 230VAC, 32A, 50-60 Hz. System dimensions: 115x53x106 cm. Activation via footswitch or finger switch.
Indications for Use
Indicated for permanent hair reduction (Fitzpatrick skin types I-VI for Nd:YAG; I-IV for Pulsed Light), treatment of benign pigmented/vascular lesions (including warts, telangiectasia, leg/spider veins), skin resurfacing, soft tissue incision/excision/ablation/vaporization, dark tattoo removal, and moderate inflammatory acne vulgaris (skin types I-IV).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Cynosure Photosilk Plus Pulsed Light & Laser Attachments (K051442)
Alma Lasers Harmony XL Multi-Application Platform (K072564)
Submission Summary (Full Text)
{0}------------------------------------------------
K082039
1 of 2
### 510(K) Summary
MAR 2 0 2009 El.En. S.p.A. Submitter: via Baldanzese, 17 50041 Calenzano (FI), Italy Phone: +39-055-882-6807 Fax: +39-055-883-2884 Email: info@elengroup.com Contact: Andrea Tozzi Quality System Manager & Official Correspondent via Baldanzese, 17 50041 Calenzano (FI), Italy Phone: +39-055-882-6807 Fax: +39-055-883-2884 Email: a.tozzi@elen.it Date Summary Prepared: March 19th, 2009 Device Trade Name: Synchro HP Platform Common Name: Medical Laser and Pulsed Light System Classification Name: Instrument, surgical, powered, laser 79-GEX 21 CFR 878.4810 K033172 - Candela Gentle YAG Family of Lasers Equivalent Devices: K051442 - Cynosure Photosilk Plus Pulsed Light & Laser Attachments K072564 - Alma Lasers Harmony XL Multi-Application Platform The Synchro HP Platform laser and pulsed light system is equipped Device Description: with a long pulse Nd: YAG laser (1064 nm), several hand held pulsed light sources (400-950 nm), a hand-held Er: Y AG laser source (2940 nm) and a hand-held Q-switched Nd: Y AG laser source (1064 nm). Emission activation is either by footswitch or finger switch. Overall weight of the system is 170 kg. Size is 115x53x106 cm (HxWxD). Electrical requirement is: 230VAC, 32A, 50-60 Hz, single phase. Indications for Use: The Synchro HP Platform laser and pulsed light system is indicated for the following treatments: Nd: YAG Laser (1064 nm): removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of pigmented and vascular lesions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions. Er:YAG Laser attachment (2940 nm): skin resurfacing and incision, excision, ablation or vaporization of soft bodily tissues.
{1}------------------------------------------------
| | K082039 | | p. | 2 of 2 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----|--------|
| Indications for Use (continued): | Q-switched Nd:YAG Laser attachment (1064 nm): removal of dark tattoos and treatment of benign pigmented lesions. | | | |
| | Pulsed Light attachments (400, 500, 520, 550, 650 to 950 nm): permanent hair reduction, treatment of benign cutaneous vascular lesions including facial and leg veins, benign pigmented epidermal lesions, moderate inflammatory acne vulgaris. | | | |
| | 400-950 nm Pulsed Light attachment:<br>✓ Moderate Inflammatory Acne Vulgaris (skin types I, II, III, IV) | | | |
| | 500-950 nm Pulsed Light attachment:<br>✓ Benign Cutaneous Vascular Lesions (skin types I, II) | | | |
| | 520-950 nm Pulsed Light attachment:<br>✓ Benign Cutaneous Vascular Lesions (skin type III)<br>✓ Benign Pigmented Epidermal Lesions (skin types I, II)<br>✓ Permanent Hair Reduction (skin types I, II) | | | |
| | 550-950 nm Pulsed Light attachment:<br>✓ Benign Pigmented Epidermal Lesions (skin types III, IV)<br>✓ Permanent Hair Reduction (skin types I, II, III) | | | |
| | 650-950 nm Pulsed Light attachment:<br>✓ Permanent Hair Reduction (skin type IV) | | | |
| Comparison: | The Synchro HP Platform system has the same indications for use, the same principle of operation, mechanism of action, and very similar performance specifications as the predicate devices. | | | |
| Nonclinical Performance Data: | None | | | |
| Clinical Performance Data: | None | | | |
| Conclusion: | The Synchro HP Platform system is another safe and effective device for the indications specified. | | | |
| Additional Information: | None | | | |
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAR 2 0 2009
El.En. Electronic Engineering SPA % Andrea Tozzi Quality System Manager Via Baldanzese, 17 50041 Calenzano (FI) Italy
Re: K082039
Trade/Device Name: Synchro HP Platform Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in in dermatology Regulatory Class: II Product Code: GEX Dated: March 12, 2009 Received: March 16, 2009
Dear Andrea Tozzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Andrea Tozzi
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
A. U. I. M. for
ark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use Statement
c f
# 510(K) Number (if known): ____K082039
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
Nd:YAG Laser (1064 nm): removal of unwanted hair, for stable long term or permanent hair reduction (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of pigmented and vascular (esions, such as but not limited to warts, teleangiectasia, leg veins and spider veins, treatment of benign pigmented lesions.
Er.Y AG Laser attachment (2940 nm): skin resurfacing and incision, excision, ablation or vaporization of soft bodily tissues.
Q-switched Nd: Y A G Laser attachment (1064 nm): removal of dark tattoos and treatment of benign pigmented lesions.
Pulsed Light attachments (400, 500, 520, 550, 650 to 950 mm): permanent hair reduction, treatment of benign cutaneous vascular lesions including facial and leg veins, benign pigmented epidermal lesions, moderate inflammatory acne vulgaris.
Different wavelength ranges of Pulsed Light attachments are indicated for the various treatments and skin types, as indicated in the following table:
| Pulsed Light<br>wavelength range | Moderate<br>Inflammatory Acne<br>Vulgaris | Benign Cutaneous<br>Vascular Lesions | Benign Pigmented<br>Epidermal Lesions | Permanent Hair<br>Reduction |
|----------------------------------|-------------------------------------------|--------------------------------------|---------------------------------------|-----------------------------|
| 400 - 950 nm | Skin Types I, II, III, IV | --- | --- | --- |
| 500 - 950 nm | --- | Skin Types I, II | --- | --- |
| 520 – 950 nm | --- | Skin Type III | Skin Types I, II | Skin Types I, II |
| 550 - 950 nm | --- | --- | Skin Types III, IV | Skin Types I, II, III |
| 650 - 950 nm | --- | --- | --- | Skin Type IV |
Prescriptive Use (Part 21 CFR 801 Subpart D) Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
oncurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Dode Form xm
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K082039
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.