The sponsor utilized published clinical performance data to demonstrate that the body-jet device is as safe and effective as its predicate devices for aesthetic body contouring.
clinical literature; aesthetic body contouring; performance data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Published clinical performance data
Patients undergoing aesthetic body contouring
Not applicable for this study
Safety and effectiveness for aesthetic body contouring
Indications for Use
Aesthetic body contouring
Device Story
Suction lipoplasty system for aesthetic body contouring; removes adipose tissue via negative pressure. System includes powered vacuum pump (double piston, 1/3 HP), collection bags with overflow/bacterial filters, and proprietary WAL Applicator (single-use tubing set with infiltration pump). Infiltration pump delivers saline to tissue area via irrigation cannulae; vacuum pump aspirates loosened fat. Used in clinical settings by physicians. Benefits include tissue-sparing fat removal; fan-shaped saline spray loosens adipose tissue while preserving connective tissue, muscle, blood vessels, and nerves.
Clinical Evidence
No prospective clinical trials were conducted. Evidence consists of in-vitro laboratory investigations on human cadaveric and excised abdominal fat tissue to assess the impact of the fan-shaped saline spray on adipose, connective, muscle, vascular, and nerve tissues. Additionally, the submission relies on a review of published clinical performance data for the device and predicate lipoplasty techniques to support safety and effectiveness claims.
Technological Characteristics
Powered vacuum pump with two WOB-L pistons; variable vacuum range 0-26.5 inHg. Includes overflow/bacterial filters, single-use WAL Applicator tubing, and re-sterilizable proprietary cannulae. System uses dynamic compression via fixed pistons. Infiltration pump delivers saline for tissue preparation. Materials and design are equivalent to predicate suction/infiltration systems.
Indications for Use
Indicated for aesthetic body contouring.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
Liposat® Power Infiltration Pump (Model 00002274) (K053451)
Byron Medical Infiltration and Aspiration Cannulae and Needles (K981172)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Human Med Ag % Tuv Sud America Inc. Stefan Preiss 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
June 8, 2021
Re: K082025
Trade/Device Name: Body-Jet Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Stefan Preiss:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 13, 2008. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a staff with a snake entwined around it, which is a symbol of medicine and healing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Human Med AG % TUV SUD America, Inc. Stefan Preiss 1775 Old Highway 8 NW New Brighton, Minnesota 55112
AUG 1 3 2008
Re: K082025
Trade/Device Name: body-jet® Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MVV Dated: July 28, 2008 Received: July 30, 2008
Dear Stefan Preiss:
We have res iewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
#### Page 2 - Stefan Preiss
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliarice at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse ventually (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use Statement
KOSTOTE 510(k) Number (if known): body-jet® Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
Aesthetic body contouring
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
**510(k) Number** k082018
Prescription Use Over-the-Counter Use yes no AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> 00041 0-0-0-0-0-0
{4}------------------------------------------------
Section 5 -- 510(k) Summary
K082025 page 1/3
# AUG 1 3 2008
# This 510(k) Summary for body-jet® meets the requirements of 21 CFR § 807.92
### 1. Submitter Information
Human Med AG Wilhelm-Hennemann-Strasse 9 D-19061 Schwerin Germany
Contact Person:
Inge Matthiesen c/o Human Med AG Wilhelm-Hennemann-Strasse 9 D-19061 Schwerin Germany
| Phone: | +49(0)385 395 70-0 |
|--------|---------------------|
| Fax: | +49(0)385 395 70-29 |
- 2. Name of Device:
| Common name: | Suction Lipoplasty System |
|----------------------|--------------------------------------------------------|
| Proprietary name: | body-jet® |
| Classification: | Suction Lipoplasty System, Class II, 21 CFR § 878.5040 |
| Product code: | MUU |
| Indications for use: | Aesthetic body contouring |
## 3. Name of the predicate device(s)
- Liposat® Power Infiltration Pump (Model 00002274) . Moeller Medical GmbH & Co. KG (K053451)
- Dominant 50 Lipo Powered Suction Pump, . Medela AG (K063336)
- Byron Medical Infiltration and Aspiration Cannulae and Needles . Byron Medical, Inc. (K981172)
{5}------------------------------------------------
K082025 Page 2/3
#### 4. Device Description
The body-jet®, a suction lipoplasty system, is a device intended for aesthetic body contouring. The device consists of a powered vacuum pump (containing an overflow/bacterial filter in the tube between the vacuum pump and suction bag), a collection bag with integrated overflow/bacterial filter, cannulae, and a connecting tube. The suction/collection bags including bacterial filters, the tubing, and the cannulae are to be changed between patients. The powered vacuum pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 26.5 inches of mercury, vacuum control valve to regulate the vacuum with accompanying vacuum gauge, a double piston, and a safety trap. The powered vacuum pump generates a negative pressure for the removal of fat or adipose tissue from distinct body-fat deposits for aesthetic body-contouring.
The body-jet® includes a sterile, single-use tubing set with a small infiltration pump (the WAL Applicator), which transports the saline for the infiltration of the tissue area to be treated, and infiltration and irrigation cannulae. The single-use tubing set (WAL Applicator) and the reusable cannulae have been designed as proprietary products.
#### 5. Indications for Use
The body-jet® is intended for aesthetic body contouring.
#### 6. Summary of Technological Characteristics
The body-jet® vacuum pump system consists of a powered vacuum pump with two WOB-L pistons (dynamic compression by fixed piston inside cylinder) working bit-parallel in the punp. The powered vacuum pump generates a negative pressure for the removal of adipose tissue. Two suction containers of identical design and mode of operation, one on the left side and one on the right side of the body-jet® device, are provided for holding the suction bags.
The suction bags are equipped with a dual-function filter acts simultaneously as an overflow protection and a bacterial barrier and closes automatically when it contact with fluids. The connecting tube between the collection bag and the vacuum pump contains an overflow protection/bacterial filter. It is used when the dual filter in the suction bag is damaged. If the fluid reaches a maximum level in the bag, the floats in the overflow filter will block the inlet port, automatically stopping the suction. This prevents the back-up of fluid to either the pump or the patient. The tubing is able to withstand the amount of negative pressure created by the vacuum pump without collapsing.
A small infiltration pump (included in the single-use WAL Applicator) transports the saline for the infiltration of the treated tissue area via the infiltration cannulae, similar to the procedure of traditional liposuction and of its predicate devices. The WAL Applicator has been designed as a single-use proprietary product. The cannulae are re-sterilizable proprietary products.
The vacuum pump system, the infiltration pump system and the cannulae of the body-jet® are substantially equivalent to the vacuum and infiltration pump systems, and the cannulae of its predicate devices in terms of intended use, design, operating principles, and materials.
{6}------------------------------------------------
K082025 page 3/3
#### 7. Summary of Histopathological and Immunohistological Performance Data
In order to support a determination of safety and effectiveness, the effect of the fan-shaped saline spray of the body-jet® has been studied on different human tissue structures. For this purpose two in-vitro laboratory investigations have been carried out. on fat tissue samples from human corpses (Investigation 1), and from "dead" abdominal fat excisates (Investigation 2). No humans have been treated for the purpose of these in-vitro laboratory investigations.
The following tissue structures were examined: :
- . adipose tissue,
- connective tissue, �
- muscle tissue, .
- . blood vessels.
- nerves. .
Based on the results of these investigations (please refer to Section 18) it can be summarized that water assisted lipoplasty with the fan-shaped water spray of the body-jet" produces no damage to vital structures like connective tissue, muscle tissue, blood vessels and nerves, thus supporting a determination of safety and substantial equivalence of the body-jet® water-assisted technique WAL to its predicate devices and the other referenced lipoplasty techniques.
#### 8. Summary of Clinical Performance Data
The evaluation of published clinical performance data demonstrate that the body-jet" device is as safe, as effective, and performs as well as its predicate devices and the other referenced lipoplasty techniques that are currently cleared for aesthetic body contouring.
Based on the results of the published clinical performance data (as presented in Section 20) and the pathohistological performance data (as presented in Section 18) it can be summarized that water assisted lipoplasty with the fan-shaped water spray of the body-jet® loosens the adipose tissue, and separates and aspirates the adipocytes without producing any damage to vital structures like connective tissue, muscle tissue, blood vessels and nerves, thus supporting a determination of safety and effectiveness, and substantial equivalence of the body-jef waterassisted technique (WAL) to its predicate devices and the other referenced lipoplasty techniques.
#### 9. Conclusion
Based upon the information presented above and in this 510(k) submission , it is concluded that the proposed body-jet® is substantially equivalent to its predicate devices, and safe and effective for the intended use of aesthetic body contouring.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.