K081994 · Orthomerica Products, Inc. · OAN · Sep 12, 2008 · Neurology
Device Facts
Record ID
K081994
Device Name
STARLIGHT
Applicant
Orthomerica Products, Inc.
Product Code
OAN · Neurology
Decision Date
Sep 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The STARlight is intended for medical purposes for use in infants from three to 18 months of age, with moderate to severe non-synostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Device Story
Custom-made cranial remolding orthosis; redirects infant head growth to improve symmetry/proportion. Practitioner acquires head shape via plaster impression or 3D scan; creates positive model; modifies model to provide space in flattened areas and total contact over prominent areas. Device worn ~23 hours/day. Designs include full form, side opening, and bivalve (two shells with sliding mechanism). Used in clinical settings; managed by practitioners. Benefits patient by providing pathway for growth into areas of least resistance, correcting deformities.
Clinical Evidence
Bench testing only. Accuracy, reproducibility, and repeatability of the STARscanner data acquisition system were validated using cylindrical shapes of varying diameters (100mm, 125mm, 150mm) scanned at multiple positions. Statistical process control confirmed accuracy improvements over the previous generation scanner.
Technological Characteristics
Custom-molded cranial orthosis. Materials: Surlyn, Co-Polyester, or copolymer plastic shell; optional pelite polyethylene foam liner or Aliplast padding. Weight: 7-10 oz. Mechanical principle: total contact over prominent areas, voids over flattened areas to redirect growth. Fabrication: positive model from negative cast or 3D scan. No electronic components in the orthosis itself.
Indications for Use
Indicated for infants 3-18 months with moderate-to-severe non-synostotic positional plagiocephaly (plagiocephalic, brachycephalic, scaphocephalic) to improve cranial symmetry/shape via mild pressure. Also indicated as adjunctive therapy for infants 3-18 months post-surgical correction of synostosis with residual moderate-to-severe cranial deformities.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
# SEP 1 2 2008
## 510(k) Summary
- I. Applicant Information
- 트 Applicant's Name and Address: Orthomerica Products Inc, 505 318 Street, P.O. Box 2927, Newport Beach, CA 92659, Telephone: (949) 723-4500, Facsimile: (949) 723-4501
FDA Establishment Registration Number 1058152
- I Contact: David C. Kerr, Chief Executive Officer, Telephone: (949) 723-4500, Facsimile: (949) 723-4501
- Submission Correspondent: Alan T. Sandifer, Director of Research and Development, 6333 North Orange Blossom Trail, Orlando, FL 32810, Telephone: (407) 290-6592, Facsimile: (407) 290-1303, asandifer@orthomerica.com
- 이 Summary Date - July 11, 2008
- II. Submission Information
- 트 Type: Traditional 510(k) Submission
- 트 Proprietary Name: STARlight
- 트 Common Name: Cranial Orthosis
- 트 Classification: Class II (special controls); OAN; 21 CFR 882.5970
- 트 Classification Name: Cranial Orthosis
- 트 Predicate Device: STARlight, Cranial Orthosis, K021207
- III. Manufacturing Site: 6333 North Orange Blossom Trail, Orlando, FL 32810, Telephone: (407) 290-6592, Facsimile: (407) 290-2419, FDA Establishment Registration Number 1058152
{1}------------------------------------------------
#### IV. Description of the Device
The STARlight redirects the head growth to improve proportion and symmetry. The practitioner takes a plaster impression or scan of the baby's head to acquire the existing shape. The mold is sealed and filled with plaster or the scanned shape is carved from a rigid polyurethane foam blank to create a positive model of the head shape. The positive model is modified to obtain greater symmetry and space in the areas of flattening. The STARlight provides total contact over the prominent or bossed areas of the baby's head to discourage growth there. Over the course of treatment, the inside of the band is modified further by the practitioner to provide space for growth to occur in the flat or depressed areas. The shape of the STARlight directs growth into the areas of least resistance and creates a precise pathway for the head shape to improve in symmetry and proportion.
The original three devices from 510(k) K021207 are made of a plastic shell of 5/32" -1/4" clear Surlyn or 1/8" - 7/32" Clear Co-Polyester. Since the original clearance a variation of the STARlight Bivalve has been released. It is identical in design to the STARlight Bivalve, but is made with an outer shell of 5/32" polyethylene-polypropylene copolymer plastic and an inner liner of 1/2" pelite polyethylene foam. Optional Aliplast (closed cell polyethylene) padding is available for the clear plastic bands.
The STARlight Cap design is made of the above plastic and contains no straps. The STARlight Side Opening design has a top opening and a side opening. The band is held in place by a Velcro strap across the side opening. The STARlight Bivalve design and its variation consist of two plastic shells that overlap with a superior sliding mechanism. The right and left overlap tabs are connected via a Velcro strap with chafe and loop.
#### Statement of Indications for Use and Intended Use V.
### Indications for Use:
The STARlight is intended for medical purposes for use in infants from three to 18 months of age, with moderate to severe non-synostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
{2}------------------------------------------------
### Intended Use:
The STARlight is designed to treat infants with abnormal head shapes from age 3 months to 18 months. Since growth is the driving factor in head shape correction, the infants wear the STARlight for approximately 23 hours per day. The most common head deformities are positional plagioccphaly, brachycephaly, and scaphocephaly. However, due to new minimally invasive surgical techniques for infants with craniosynostosis, post-surgical plagiocephaly, brachycephaly, and scaphocephaly are emerging as a growing patient group.
Craniosynostosis is caused by the premature fusion of one or more cranial sutures of the skull, causing the head to grow into an unusual shape. Some types of craniosynostosis have a clinical presentation similar to deformational (positional) plagiocephaly. Therefore in cases where the physician cannot make a definitive diagnosis, patients are referred to specialists such as neurosurgeons or cranio-facial surgeons. These specialists will order a test like a CT scan or MRI to confirm the diagnosis of craniosynostosis. If a baby has craniosynostosis, surgery is indicated to realign the plates of the skull and allow normal brain and skull growth to occur.
In general, the first year of life is the optimum time frame for surgical correction since infants are growing at such an accelerated rate during that time. No matter which surgical technique is used, the end result is a patient with no fused sutures. At this point the same principles that guide cranial remolding of deformational head shapes are applicable. In both deformational head shapes and post-surgical head shapes the STARlight is designed to maintain total contact over areas where growth is not desired, and allow for space over areas where growth is desired. The STARlight provides a pathway for the baby's head growth, directing it toward a more normal shape.
#### VI. Summary of Technological Characteristics
As mentioned previously, minor changes that were determined not to require 510K submission have been implemented via Orthomerica's Engineering Change Order process. These changes are illustrated in the following table. Furthermore, there are no technological characteristics changes required for the expansion in indications requested within this 510(k) submission. Therefore this table may also serve as a comparison of the proposed device as compared to the predicate device.
{3}------------------------------------------------
### Table 1
Comparison of Predicate Device cleared in 510(k) K021207 to the currently marketed device: Note: No changes will be made to the currently marketed device as a result of the proposed indications change in this submission
| Feature | 510K Cleared Device | Currently Marketed Device |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Maintains total contact over areas of<br>bossing or protrusion and creates voids<br>over areas of depression or flattening to<br>redirect cranial growth toward greater<br>symmetry. | Maintains total contact over areas of<br>bossing or protrusion and creates voids over<br>areas of depression or flattening to redirect<br>cranial growth toward greater symmetry. |
| Materials | Material for all 3 designs:<br>- 5/32" - 1/4" clear Surlyn or 1/8"<br>- 7/32" Clear Co-Polyester<br>plastic shell | Material for original 3 designs:<br>- 5/32" - 1/4" clear Surlyn or 1/8" -<br>7/32" Clear Co-Polyester plastic<br>shell<br>Material for Bivalve variation<br>Outer shell of 5/32" copolymer<br>plastic<br>An inner liner of 1/2" pelite<br>polyethylene foam |
| | Closure for Bivalve design<br>Sliding/Overlap closure system<br>Chicago screw (or similar) for<br>tope sliding mechanism<br>1" velcro strap<br>1" chafe buckle<br>91X speedy rivets | Closure for Bivalve design<br>Sliding/Overlap closure system<br>Chicago screw (or similar) for tope<br>sliding mechanism<br>1" velcro strap<br>1" chafe buckle<br>91X speedy rivets |
| | Closure for Side Opening design:<br>1 ½" Velcro Strap<br>1 ½" chafe buckle<br>91X speedy rivet<br>Nylon washer | Closure for Side Opening design:<br>1" Velcro Strap |
| | (no closure for full form design - just<br>plastic shell) | (no closure for full form design - just<br>plastic shell) |
| | Optional Cheek Pads<br>1/8" Aliplast foam | Optional Cheek Pads<br>1/8" Aliplast foam |
| Product Design | Custom made cranial remolding orthosis,<br>approximately 7 to 10 ounces in weight,<br>respective to the full form, side opening<br>and bi-valved design. | Custom made cranial remolding orthosis,<br>approximately 7 to 10 ounces in weight,<br>respective to the full form, side opening and<br>bi-valved design. |
| Production | Form orthosis from a positive mold of the<br>infants head; positive mold is formed<br>from a negative cast mold or three-<br>dimensional scan of the infants head. | Form orthosis from a positive mold of the<br>infants head; positive mold is formed from a<br>negative cast mold or three-dimensional<br>scan of the infants head. |
The two changes indicated on the table are the introduction of the bivalve design variation and the strap change on the STARlight Side Opening Design. The bivalve design is a simple material change. The STARlight Side Opening strap design was improved to eliminate the possibility for the metal rivet to contact the patient's skin.
{4}------------------------------------------------
#### VII. Summary and Conclusions of Non-Clinical Performance Data
The STARlight cranial orthosis has been used successfully in clinical practice since its original clearance in 2002. The minor changes mentioned in the technical characteristics section above involved standard orthotic fabrication and materials that have undergone biocompatibility testing. However, the STARscanner™ Data Acquisition System used to capture the infant's head shape has had minor technical changes and with due diligence performance testing was conducted. The changes made to the STARscanner were to improve ease of use for the practitioner and to update components to the state of the art.
The accuracy, reproducibility, and repeatability of the STARscanner was evaluated by scanning three different cylindrical shapes( 100mm, 125mm, 150mm diameters) five times at five different positions within the scan volume. Standard measurement systems statistical process control procedures were utilized to evaluate STARscanner errors, error standard deviations, repeatability of multiple scans, and reproducibility of multiple scans at multiple locations within the scan volume. The accuracy of the STARscanner II was confirmed to be better than the STARscanner I.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle.
SEP 1 2 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Orthomerica Products, Inc. % Mr. Alan T. Sandifer Director of Research and Development 6333 North Orange Blossom Trail Orlando, Florida 32810
Re: K081994
Trade/Device Name: STARlight® Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial orthosis Regulatory Class: II Product Code: OAN, MVA Dated: July 11, 2008 Received: July 14, 2008
Dear Mr. Sandifer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Alan T. Sandifer
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to procced to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Pretmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Statement of Indications for Use
510K Number (if known): unknown KOSS 1994
Device Name: STARlight®
Indications for Use:
The STARlight is intended for medical purposes for use in infants from three to 18 months of age, with moderate to severe non-synostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic- and scaphocephalic-shaped heads by applying mild pressure to prominent regions of the infant's cranium in order to improve cranial symmetry and/or shape. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
D
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** k011954
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.